Ibrahim A. Elabyad1, Maxim Terekhov1, David Lohr1, Maya Bille1, and Laura M. Schreiber1
1Chair of Molecular and Cellular Imaging, Comprehensive Heart Failure Center (CHFC), University Hospital Wuerzburg, Wuerzburg, Germany
Synopsis
A dedicated flexible 16-element transceiver coil array was developed,
simulated, and tested for cardiac MRI (cMRI) in pigs at 7T. The 16-elements of
the array are printed on a flexible (Kapton polyimide) substrate to conform to the
pig thorax. The performance of the elastic array is evaluated by EM-simulations
and MR-measurements in a pig phantom and a 45-kg pig in-vivo. The flexible
array supports parallel imaging with acceleration factors up to R=4 with
reasonable g-factors (g=1.2) in an ROI at the position heart. The array enables
geometrical conformity to pig body weights within a large range of sizes and
shapes.
Introduction
A wide variety of different coil array designs have allowed significant
progress for body imaging at ultrahigh-field (UHF) (e.g., using local Tx/Rx loop1-4 and dipole
antenna arrays5,6). Significant advancement of both $$${B_1^+}$$$-shimming
and parallel imaging capabilities were demonstrated for in-vivo cMRI in pigs using an antisymmetric 8Tx/16Rx loop array7. To
obtain the best image quality for different pigs (25-90kg), three dedicated rigid
coil arrays were implemented. These coil arrays have a rigid form and are thus
optimized for MR-experiments on pigs with a certain pig weight and thorax shape.
However, building an individual coil for each weight range is both time and cost-consuming.
The purpose of this work is to design, simulate, implement, and get initial
usage experience with elastic coil array, which can be
used potentially for cMRI in pigs of a wide range of thorax dimensions and
weights (25-90kg).Methods
In this work, we introduce a dedicated transceiver array printed on a
flexible 100-µm thick Kapton
polyimide (PI) substrate (εr=3.5 and $$$tanδ$$$=0.001). The copper (Cu) trace width is 4mm with 35-µm thickness. The array
is composed of 16-elements distributed anti-symmetrically around the central
two elements7 [Figure 1]. The
sizes of the elements 1, 2, 7, 8, 9, and 10 were 4.5×9.8cm2. The
decoupling between the central two elements was accomplished using a common
conductor and shared decoupling capacitor (SDC) ($$${C_1^d}$$$). The
size of elements 3 and 4 was 5×8cm2. Elements 3 and 4 were decoupled from elements 1 and 2 using SDC ($$${C_2^d}$$$). The size of
elements 5 and 6 was 5.6×12.2cm2. The sizes of elements 7 and 9 and
the identical elements 8 and 10 are 4.5×9.8cm2. They were decoupled
using a shared decoupling capacitor ($$${C_3^d}$$$). Elements
1, 2, and 3 were decoupled from the neighboring elements 5 and 6 using
capacitive decoupling ($$${C_5^d}$$$, $$${C_6^d}$$$,
and $$${C_7^d}$$$) in addition to a decoupling gap of 2cm. The
external dimension for the array was 57.9×20.8cm2. EM-simulations
were performed using CST-Microwave-Studio for coil design and $$${B_1^+}$$$-field
optimizations. For matching, tuning, and decoupling, RF-circuit co-simulation was
employed in CST-Design-Studio to get an initial guess for the optimal lumped
elements8. Static phase-only
$$${B_1^+}$$$-shimming was performed within a
pig phantom using two different optimization cost functions ($$${FC_1}$$$ for $$${B_1^+}$$$-homogeneity and $$${FC_2}$$$ for
weighted combination of $$${B_1^+}$$$-homogeneity and transmit
efficiency7. To
characterize the Tx performance of the array, $$$CoV=\frac{std(B_1^+)}{mean(B_1^+)}$$$ and $$$Tx_{eff}=\frac{mean(B_1^+)}{\sqrt{P_{acc}}}$$$ were computed. To form
an 8Tx/16Rx array compatible with the pTx system, every two neighboring elements
were combined in one Tx-channel [Figure 1a]. All measurements
were performed on a 7T whole-body MAGNETOM Siemens Terra scanner. The array was
tested in phantom with the measurement of SNR-maps, $$${B_1}$$$-shimming, g-factor with
high parallel imaging acceleration factors. The reconstructed experimental
g-factor and SNR-maps were post-processed in MATLAB. Female pigs (German
Landrace) were used after approval by the responsible local animal welfare
committee (project 55.2 DMS 2532-1134-16, Regierung von Unterfranken). All in-vivo
measurements were done using a GRE cardiac sequence
(BEAT) provided by the
scanner vendor. To visualize RF-excitation coverage of the heart volume, 10
coronal, sagittal, and transversal slices positioned within the heart were
acquired with prospective triggering. Imaging parameters were: TR=44ms, TE=2.9ms,
matrix size=150x120, FOV=300x300mm2, slice thickness 6mm,
inter-slice distance 1.2mm, FA=20°. Then CINE-imaging with a
retrospective reconstruction of 30 heart phases was performed for standard
cardiologic views of the heart: four-chamber, two-chamber, and SA.Results
Figure
2 shows the simulated $$${B_1^+}$$$-field distribution for
the 16-element flexible array after static phase-only $$${B_1^+}$$$-shimming
within a dedicated pig phantom with PV1 and PV2 compared to the zero phases. With PV1, $$$CoV$$$ was improved by 70% in the central transversal slice compared to zero-phases.
With PV2,
$$$Tx_{eff}$$$ was improved by 3-times compared to zero-phases.
Figure 3 shows
the measured g-factor maps (using acceleration factors R=2, 3, 4, and 6), the
SNR-maps, the noise correlation matrix of the flexible array loaded with a pig
phantom, and in-vivo pig. The mean SNR within the heart region from the novel
design was 70. With the implemented hardware phases and even prior $$${B_1^+}$$$-shimming,
the coil shows good Rx properties with high SNR.
Figures
4 and 5 show in-vivo images of a
45-kg pig with normal and high resolutions acquired using the flexible
array. The developed coil demonstrates sufficient $$${B_1^+}$$$-field
homogeneity within the pig heart without significant destructive interference.Discussion
The initial experience of using the coil in the animal study has
demonstrated that the fully elastic array
provides sufficient $$$Tx_{eff}$$$ with $$${B_1^+}$$$-penetration on
the full depth of the pig heart and coverage of the whole heart volume. The reasonable imaging quality with SNR
moderately lower than delivered by the dedicated rigid array optimized for specific
weight range was shown. Further optimizations of the tuning/matching circuits should be performed to make possible application of the fully elastic array for
the wide range of the animal’s thorax shapes.Conclusion
A dedicated flexible pig array with the best bendability to conform to the
thorax shape of several pigs was introduced. All fixed capacitors, cable traps, and semi-rigid cables were fixed on the flexible PI-substrate without getting
broken from several bending. Further optimizations of the tuning/matching will
be performed to optimize the array for a wide range of pig weights.Acknowledgements
This project was funded by the Federal Ministry of
Education and Research (BMBF), Grant/Award Number: 01EO1004 & 01EO1504.References
-
Graessl A, Renz W, Hezel
F, et al. Modular 32-channel transceiver coil array for cardiac MRI at
7.0T. Magn Reson Med. 2014;72(1):276-290.
- Dieringer
MA, Renz W, Lindel T, et al. Design and application of
a four-channel transmit/receive surface coil for functional cardiac
imaging at 7T. J Magn
Reson Imaging. 2011;33(3):736-741.
- Grassl A, Winter L,
Thalhammer C, et al. Design, evaluation and application of an eight
channel transmit/receive coil array for cardiac MRI at 7.0 T. Eur J
Radiol. 2013;82(5):752-759.
- Thalhammer
C, Renz W, Winter L, et al. Two-dimensional sixteen
channel transmit/receive coil array for cardiac MRI at 7.0 T: design,
evaluation, and application. J Magn Reson Imaging. 2012;36(4):847-857.
- Raaijmakers AJ, Ipek O,
Klomp DW, et al. Design of a radiative surface coil array element at 7 T:
the single-side adapted dipole antenna. Magn Reson Med. 2011;66(5):1488-1497.
- Raaijmakers AJ,
Italiaander M, Voogt IJ, et al. The fractionated dipole antenna: A new
antenna for body imaging at 7 Tesla. Magn Reson Med. 2016;75(3):1366-1374.
- Elabyad IA, Terekhov M,
Lohr D, Stefanescu MR, Baltes S, Schreiber LM. A Novel Mono-surface
Antisymmetric 8Tx/16Rx Coil Array for Parallel Transmit Cardiac MRI in
Pigs at 7T. Scientific
Reports. 2020;10(1):3117.
- Kozlov M, Turner R. Fast
MRI coil analysis based on 3-D electromagnetic and RF circuit
co-simulation. Journal
of Magnetic Resonance. 2009;200(1):147-152.