Jürgen Herrler1, Jonathan Endres1, Raphael Tomi-Tricot2,3,4, Patrick Liebig5, Rene Gumbrecht5, Christian Richard Meixner6, Andreas Maier7, Arnd Dörfler1, Shaihan Malik3,4, and Armin Michael Nagel6,8
1Department of Neuroradiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 2MR Research Collaborations, Siemens Healthcare Limited, London, United Kingdom, 3Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King’s College London, London, United Kingdom, 4Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King’s College London, London, United Kingdom, 5Siemens Healthcare GmbH, Erlangen, Germany, 6Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 7Department of Computer Science, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 8Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany
Synopsis
Direct Signal
Control with Variable Excitation and Refocusing (DiSCoVER) was combined with a dynamic, Fast Online-Customized (FOCUS)
parallel transmit excitation pulse for a 3D TSE sequence using an 8Tx/32Rx head
RF coil at 7 Tesla. DSC could improve the signal, mainly in
the cerebellum region compared to CP mode. FOCUS excitation pulses achieve better FA
and phase homogeneity than static pTx pulses (standardly used in DiSCoVER). Combining a FOCUS excitation pulse with DiSCoVER-optimized refocusing pulses, led to signal gain across the echoes, which was evaluated in
simulations for 132 subjects. Universal and individual DiSCoVER-refocusing pulses showed comparable performance.
Introduction
Turbo-spin echo (TSE) sequences at ultra-high fields suffer
from poor B1+ homogeneity and conservative local
specific absorption rate (SAR) limits1. Direct signal
control with Variable Excitation and Refocusing (DiSCoVER), using parallel transmit (pTx) pulses, has shown to improve the MR
signal homogeneity across a refocusing pulse train under strict SAR constraints2. This method
calculates individually optimized, pTx scale factors for each pulse in the train (excitation and refocusing) and has been integrated into an MR system despite requiring to solve a large problem during the
scan (one minute online-calculation3). Dynamic pTx
pulses apply time-varying B1+-fields and apply B0-field gradients to adjust the nuclear spin phases during the pulse. These
have shown greater potential to achieve homogeneous flip angle (FA) and phase
patterns than static pTx4. Universal
pulses (UPs), as exemplary dynamic pTx pulses, have been shown to be applicable
for both gradient echo (GRE)5 and TSE6 sequences with
robust performance and without the need of any online-calculation time.
To individually optimize dynamic pTx pulses in a clinically acceptable time (~1 minute), fast online-customized (FOCUS) pulses have
been presented, which consist of a universal pulse and parameter optimization
prior to the scan and a thereby faster and more robust individual online-optimization7. In this work,
we combine a dynamic, FOCUS pTx excitation pulse with refocusing pulses that
are optimized with DiSCoVER for a 3D TSE sequence using an 8Tx/32Rx head coil (Nova
Medical, Wilmington, USA) at 7 Tesla.Methods
Firstly, a nonselective, dynamic, universal
pTx excitation pulse was optimized by using a kT point trajectory with
globally optimizing 6 kT point locations and respective sub-pulse
durations in one step and, in a second step, corresponding universal pulse
shapes. We used B1+ and B0 maps of 20 training
subjects and a specifically optimized interior-point method for the optimization8. The UP served
as an initialization for further individual optimization to achieve a 90° FA
and homogeneous phase excitation, which was assumed to provide optimal
conditions for generating TSEs with high signal. Bloch simulations were performed for the FOCUS excitation pulses, applied on
every subject to obtain the corresponding FA and phase patterns.
Secondly, based on these
patterns, the following 9 static pTx refocusing pulses
were optimized universally across all training subjects by using DiSCoVER. Then, further individually DiSCoVER-optimized refocusing pulses were generated, using the universal ones as initialization.
Finally, four different cases were evaluated in simulations and compared
with respect to their achieved homogeneity under the same SAR (channel power
limits) and peak-voltage constraints: Purely circularly polarized (CP) pulses
(1), DiSCoVER (2) and a FOCUS, dynamic excitation pulse, followed
by refocusing pulses, which were optimized either universally (3) or individually (4) using DiSCoVER.Results
Figure 1 shows the FA and
phase distribution in a central sagittal brain slice of one subject for a static and a FOCUS, dynamic pTx excitation pulse. Both
distributions become clearly more homogeneous when using the dynamic pulse.
Figure 2 shows the simulated spatial signal distribution, relative to the target signal (starting from 1, relaxation with T1=1s and T2=70ms in every voxel) in a central sagittal slice of another subject, averaged over the first 10 echoes for the
four mentioned cases. DiSCoVER reaches much stronger signal compared to the
CP mode, especially in the cerebellum. Further improvement is reached
when using a dynamic excitation pulse and universal DiSCoVER-calculated shims. The highest
signal gain is achieved in the frontal lobe and cerebellum. Individually
optimized shims then only slightly change the signal, compared to universal
ones.
The relative differences of these echo-averaged signals from the target were then evaluated on 132 subjects and
averaged over all valid voxels across the brain region. The resulting scalar
values are shown in Figure 3. It demonstrates lower signal deviations in both
mean and variance among different subjects when using DiSCoVER instead of CP mode and also when using FOCUS excitation pulses instead of standard ones across all subjects.Discussion and Conclusions
DiSCoVER can
be further improved when combined with a dynamic, FOCUS pTx excitation pulse.
The FOCUS pulse achieves more homogeneous FA and phase patterns than a
static pTx pulse. Additionally, an individual DiSCoVER-optimization shows similar quality as previously universally optimized refocusing pulses. Thus, as also suggested to earlier study findings9, skipping the DiSCoVER-optimization time of approximately one
minute3 could ultimately be a valid option. In this case, after rapid B1+
and B0 mapping, only about 6 seconds online-optimization time
for the FOCUS excitation pulse would be necessary.
To further improve the
performances, the dynamic excitation pulses could be integrated into the
DiSCoVER-optimization. Additionally, the refocusing pulses could consist of several sub-pulses (B1+-fields) and gradient fields as well,
which might lead to even better homogeneity throughout the echo train. On the
other hand, this would require an even more complex problem to solve with DiSCoVER.
Furthermore, it might lead to worse SAR efficiency when using fixed per-channel
power limits. To still guarantee safety and meet the local SAR limits, virtual
observation points (VOPs)10, which estimate spatial distribution of local SAR
exposure for different B1+ fields, could be advantageous. Those have
shown a potentially great reduction of the local SAR estimations, especially
for highly dynamic pulses (many different B1+ fields)11.Acknowledgements
The authors thank Dr. Ioannis-Angelos Giapitzakis for proofreading and useful suggestions that helped to improve this abstract.References
1 Fiedler, T. M., Ladd, M. E. &
Bitz, A. K. SAR Simulations & Safety. Neuroimage
168, 33-58,
doi:10.1016/j.neuroimage.2017.03.035 (2018).
2 Malik, S. J.,
Beqiri, A., Padormo, F. & Hajnal, J. V. Direct signal control of the
steady-state response of 3D-FSE sequences. Magn
Reson Med 73, 951-963,
doi:10.1002/mrm.25192 (2015).
3 Tomi-Tricot, R. et al. Fully Integrated Scanner Implementation of Direct Signal Control for 2D T2-Weighted TSE at Ultra-High Field. 29th Annual Meeting of the Intl. Soc. Magn. Reson. Med. (2021).
4 Padormo, F.,
Beqiri, A., Hajnal, J. V. & Malik, S. J. Parallel transmission for
ultrahigh-field imaging. Nmr in
Biomedicine 29, 1145-1161,
doi:10.1002/nbm.3313 (2016).
5 Gras, V.,
Vignaud, A., Amadon, A., Le Bihan, D. & Boulant, N. Universal pulses: A new
concept for calibration-free parallel transmission. Magn Reson Med 77,
635-643, doi:10.1002/mrm.26148 (2017).
6 Gras, V. et al. Robust nonadiabatic T2
preparation using universal parallel-transmit kT -point pulses for 3D FLAIR
imaging at 7 T. Magn Reson Med 81, 3202-3208, doi:10.1002/mrm.27645
(2019).
7 Herrler, J. et al. Fast online-customized (FOCUS)
parallel transmission pulses: A combination of universal pulses and individual
optimization. Magn Reson Med 85, 3140-3153, doi:10.1002/mrm.28643
(2021).
8 Majewski, K.
Simultaneous optimization of radio frequency and gradient waveforms with exact
Hessians and slew rate constraints applied to kT-points excitation. J Magn Reson 326, 106941, doi:10.1016/j.jmr.2021.106941 (2021).
9 Beqiri, A.,
Hoogduin, H., Sbrizzi, A., Hajnal, J. V. & Malik, S. J. Whole-brain 3D
FLAIR at 7T using direct signal control. Magn
Reson Med 80, 1533-1545,
doi:10.1002/mrm.27149 (2018).
10 Eichfelder, G.
& Gebhardt, M. Local Specific Absorption Rate Control for Parallel
Transmission by Virtual Observation Points. Magnetic
Resonance in Medicine 66,
1468-1476, doi:10.1002/mrm.22927 (2011).
11 Williams, S. N. et al. SAR Management in pTx Sequence Design: The Impact of Electromagnetic-Field-Derived Virtual Observation Points. 29th Annual Meeting of the Intl. Soc. Magn. Reson. Med. (2021).