3283

SENSE-like reconstruction of multi-average body DWI to remove motion-induced signal loss: application in liver DWI
Anh Tu Van1, Sean McTavish1, Johannes K. J. Raspe1, Felix Harder1, Johannes M. Peeters2, Kilian Weiss3, Marcus R Makowski1, Rickmer F. Braren1, and Dimitrios C Karampinos1
1Department of Diagnostic and Interventional Radiology, Technical University of Munich, Munich, Germany, 2Philips Healthcare, Best, Netherlands, 3Philips GmbH Market DACH, Hamburg, Germany

Synopsis

Motion-induced signal loss and phase maps were generated and used in a SENSE-like reconstruction routine to solve for a single DWI from a multi-average liver diffusion-weighted imaging experiment. The proposed reconstruction yields homogeneous liver signal and possibly improves diagnostic values.

Purpose

Diffusion-weighted imaging is nowadays used in routine body oncological imaging to detect lesions [1]. However, body DWI remains strongly challenged by motion effects that induce different types of artifacts including the motion-induced signal loss typically observed in the left liver lobe [2, 3]. With respiratory gating, the motion-induced signal loss in the left liver lobe is caused primarily by cardiac motion [2]. To mitigate the signal loss problem, multiple averages of the same b-value and diffusion-encoding direction are typically acquired and averaged. The conventional sum-of-squares (SOS) averaging method disregards the difference in the quality of the averages and treat all averages equally. As a result, significant liver signal inhomogeneity can still be observed. A previous work has tried to learn with a network the quality (weights) of different averages and perform the weighted averaging [4]. In this work, complex motion-induced signal loss and phase maps were generated and used in a SENSE-like reconstruction routine to solve for a single diffusion-weighted image (DWI) from a multi-average liver diffusion-weighted imaging experiment.

Methods

Data acquisition: One volunteer and two patients were scanned on a 3T whole-body scanner (Ingenia Elition, Philips Healthcare) using the 16-channel torso coil and the built-in-table 12-channel posterior coil. A spin echo diffusion-weighted sequence in combination with a single-shot EPI readout was employed. Patient data were acquired with the following parameters: FOV = 420 x 370 mm2, voxel size = 3 x 3 x 4 mm3, 60 slices with 0.4 mm gap, b-values = 0, 50, 300, 600 s/mm2 with the number of averages of 2, 1, 2, 6, respectively, 3 diffusion encoding directions (b-value > 0). For volunteer data, 10 slices with 6 mm gap, b-values = 0, 200, 400, 600 s/mm2 with the number of averages of 4, 5, 6, 8, respectively were acquired. All scans were respiratory triggered with a fixed trigger delay of 200 ms.
Image reconstruction: Individual averages for each b-value and diffusion direction were reconstructed offline. The complex motion-induced signal loss and phase (MISLP) maps were determined as
$$S_{(k,d,b)}(r) = \frac{I^{lr}_{(k,d,b)}(r)}{\max_{k}|I^{lr}_{(k,d,b)}(r)|}$$
where $$$I^{lr}_{(k,d,b)}$$$ is the complex low resolution DWI at average $$$k$$$, diffusion direction $$$d$$$, and b-value $$$b$$$. Further smoothing of the magnitude of the MISLP maps were achieved by employing a 2D median filter of an empirical width of 8 x 8 voxels.
Let $$$m$$$ be the signal-loss-free DWI reconstructed from the multi-average data, similar to a SENSE problem with reduction factor of 1, we have
$$\begin{pmatrix}I_{(k_1, b, d)}(r) \\
I_{(k_2, b, d)}(r) \\
\vdots \\
I_{(N_k, b, d)}(r)
\end{pmatrix} = \begin{pmatrix}S_{(k_1, b, d)}(r) \\ S_{(k_2, b, d)}(r) \\ \vdots \\S_{(N_k, b, d)} \end{pmatrix}m(r) $$
where $$$N_k$$$ is the number of averages for direction $$$d$$$ and b-value $$$b$$$, $$$I_{(k, d, b)}$$$ is the high resolution DWI. The solution for the above equation is
$$m(r) = \frac{\sum_k {S^*}_{(k,d,b)}(r)I_{(k,d,b)}(r)}{\sum_k|S_{(k,d,b)}(r)|^2}.$$

Results

Figure 1 shows examples of the MISLP maps from 6 averages in one volunteer slice. Areas of signal loss especially in the left liver lobe and their different appearance across averages were captured (Figure 1a). Rapidly varying motion-induced phases that are different from average to average can also be observed (Figure 1a). Figure 2 shows the DWIs from three diffusion encoding directions with SENSE-like reconstruction (SENSE-like) and SOS averaging. The SENSE-like results remove the observed signal loss in comparison the SOS results, yielding more homogeneous liver signal. Average DWI across three diffusion encoding directions are presented in Figure 3 for two patients. Similar to the volunteer data, SENSE-like reconstruction results significantly reduce the appearance of signal loss, giving better visualization of anatomies and pathologies (lesion on the left lobe of patient 1).

Discussion

A methodology was proposed formulating the averaging of the DWIs as reconstruction problem with SENSE at R=1 and treating the motion induced signal loss maps as sensitivity maps. The proposed SENSE-like reconstruction of the multi-averages DWIs mitigated motion-induced signal loss compared to the conventional SOS averaging method. The proposed SENSE-like reconstruction of multi-average DWI data yields more homogeneous liver signal and should be tested whether it improves diagnostic image quality in a larger clinical DWI dataset.

Acknowledgements

The present work was supported by Philips Healthcare.

References

1. Lewis S, Dyvorne H, Cui Y, Taouli B. Diffusion-weighted imaging of the liver: techniques and applications. Magn Reson Imaging Clin N Am 2014;22(3):373-395.

2. Kwee TC, Takahara T, Niwa T, Ivancevic MK, Herigault G, Van Cauteren M, Luijten PR. Influence of cardiac motion on diffusion-weighted magnetic resonance imaging of the liver. MAGMA 2009;22(5):319-325.29.

3. Taouli B, Koh DM. Diffusion-weighted MR imaging of the liver. Radiology 2010;254(1):47-66.

4. Gadjimuradov F, Benkert T, Nickel MD, Maier A. Deep Learning-based Adaptive Image Combination for Signal-Dropout Suppression in Liver DWI. Proc. Intl. Soc. Mag. Reson. Med. 29 (2021):0534

Figures

Figure 1: Motion-induced signal loss and phase maps from 6 averages. Areas of signal loss and their different appearance across averages were captured. Rapidly varying motion-induced phases can be observed.

Figure 2: DWIs from three diffusion encoding directions on a volunteer. SENSE results eliminate the motion-induced signal loss observed in the SOS results (blue arrows).

Figure 3: Averaged DWI across three diffusion encoding directions. SENSE results significantly reduce the appearance of signal loss, giving better visualization of anatomies and pathologies.

Proc. Intl. Soc. Mag. Reson. Med. 30 (2022)
3283
DOI: https://doi.org/10.58530/2022/3283