Yuxin Wu1, John Pauly2, and Greig Scott2
1Department of Electronic Engineering, Tsinghua University, Beijing, China, 2Department of Electrical Engineering, Stanford University, Stanford, CA, United States
Synopsis
This work assesses
heartbeat detection by continuous-wave (CW) Doppler radar using low and
high carrier frequencies of 433MHz and 5GHz as options for non-contact
vitals sensing in MRI. The experimental results show that using the amplitude of received
complex signals in low frequency 433MHz-detection can perform as well as
5GHz-detection due to near field interactions within a 20cm range, with similar
heartbeat detection efficacy.
INTRODUCTION
This work assesses heartbeat detection by continuous-wave (CW)
Doppler radar at a low carrier frequency of 433MHz for future use in MRI (Figure 1). In conventional MRI, respiratory bellows gating and cardiac gating by ECG
or photoplethysmography (PPG) are commonplace to sort MRI data into cardiac
phases, but often complicate setup for the patient1. In
noncontact vital signs sensing, continuous-wave Doppler radar is widely used at
2.4GHz, 5GHz or even higher. In principle, higher carrier frequencies provide
greater phase sensitivity to motion at a distance, but this is only true for
equal carrier phase noise2. In MRI,
cardiac signal detection will be performed over distances under 20cm, where
near field proximity and low frequency field penetration can modulate the
reflected signal. Here, a software-defined radio (Nuand bladeRF 2.0 SDR) is
configured to assess in-vivo heart beat detection under free-breathing
conditions for 20cm range at a low 433MHz and typical 5GHz carrier. The radar
system adopts a single channel heterodyne receiver architecture using the bladeRF
and dual antennas. The radar cardiac signals were validated with a reference
PPG signal.METHODS
The
experimental setup is shown in Figure 2. The radar system exploits one transmit
channel and two receive channels, with Tx1 and Rx1 antennas 20cm away from the chest,
and an Rx2 antenna on the subject's chest. Rx2 maximizes near field sensitivity
for potential bistatic radar configurations, if feasible. For 433MHz detection,
the Tx1 and Rx1 channel shared one antenna via a power splitter (Figure 2). For
5GHz detection, the Tx1 and Rx1 channels used separate antennas. The PPG sensor
(PulseSensor.com) was interfaced to an Arduino UNO, and synchronized with the
SDR to sense the index finger pulse. The sampling rates of the SDR and PulseSensor
were 1MHz and 500Hz, respectively. The radar and PPG signals were down-sampled
to 100Hz by MATLAB post-processing, using a lowpass Chebyshev Type I IIR filter
of order 8. Vital sign measurements were repeated 10 times - each trial containing
a 10-second time series. The signals were then bandpass filtered to select the cardiac
band (0.6Hz to 2Hz) using cascaded IIR lowpass and highpass filters in MATLAB. Finally,
Singular Spectrum Analysis (SSA) was applied to decompose the time-domain
signals into different reconstruction components. The reconstruction components
were selected based on prior knowledge of the heartbeat waveform3. Ultimately,
the radar cardiac pulse “trigger” was extracted and compared with the reference
PPG signal. To evaluate the accuracy of the reconstruction, the average time
error between the radar and PPG pulses were calculated after removing the
influence of Pulse Transit Time (PTT).
For the free-breathing 5GHz radar signals, a compensation of DC
offset based on L2-norm minimization was also applied when processing the radar
signals4.RESULTS
Figure
3 shows a reconstruction using the phase of the complex radar signal from
receiving channel 1 after down conversion with the 433MHz detection under
free-breathing condition. In 3(a), the original filtered radar signal (blue)
and recovered heartbeat waveform after SSA (red) are shown. Figure 3(b) shows the
recovered heartbeat waveform (blue), and the reference PPG signal (orange). In 3(c),
two sequences of pulses extracted from the recovered heartbeat waveform and the
reference PPG signal are compared to evaluate the reconstruction quality.
Unpaired pulses at the beginning or end of the sequences, caused by PTT, were
removed when evaluating the reconstruction. Figure 4 displays histograms of
unmatched pulse numbers between reconstruction and reference using different
post-processing. Figure 5 shows the average time error between the
reconstruction and reference pulses after removing unmatched pulses. When
calculating the average time error of reconstruction, the constant time shift
in each pulse pair, caused by PPT and estimated by the average time shift in
pairs, was subtracted from the original time shifts in each pair, and the
absolute values of the residues were summed up and averaged into the time error.
From Figure 4 and 5, with 433MHz-detection, the reconstruction of heartbeats
using the amplitude of complex signal is better than that from phase-only in
both free-breathing and breath-hold cases. Use of the amplitude of complex
signal at 433MHz also outperformed the traditional method of using phase at
5GHz to reconstruct the heartbeat under free-breathing conditions.DISCUSSION
433MHz-detection
has about 10x larger wavelength than 5GHz-detection, which ought to reduce phase
sensitivity to motion accordingly. However, for short-range detection of 20cm,
near-field effects at 433MHz cause significant amplitude and phase modulation
from cardiac activity in the complex radar signal. This phenomenon is enhanced
even more by a one-way (bistatic) radar signal instead of the reflected signal.
This should be expected since the 433
MHz ISM band corresponds to a 10T MRI frequency. Interestingly, while CW Doppler
traditionally relies on phase-based group delay, once the full complex signal
is processed, the outcome is entirely equivalent to impedance sensing.CONCLUSION
This work demonstrates the possibility of heartbeat detection using
CW Doppler radar in the 433MHz ISM band. A new evaluation criterion for
reconstruction by calculating the average time error is proposed in the work,
which aims to generate retrospective radar-based cardiac gating for MRI.
Experimental results show that the low frequency 433MHz-detection can perform
as well as 5GHz-detection. Acknowledgements
NIH grants: U01EB029427, R01EB01924105,
R01EB012031, U01EB026412, GE HealthcareReferences
[1] Rosenzweig, S. , et al.
"Cardiac and Respiratory Self-Gating in Radial MRI using an Adapted
Singular Spectrum Analysis (SSA-FARY)." IEEE Transactions on Medical
Imaging PP.99(2020):1-1.
[2] Li, C. , et al. "A Review on
Recent Advances in Doppler Radar Sensors for Noncontact Healthcare
Monitoring." IEEE Transactions on Microwave Theory and Techniques
61.5(2013):2046-2060.
[3] Ghaderi, F. , H. R. Mohseni , and S.
Sanei . "Localizing Heart Sounds in Respiratory Signals Using Singular
Spectrum Analysis." Biomedical Engineering IEEE Transactions on
58.12(2011):p.3360-3367.
[4] Zakrzewski, M. , H. Raittinen , and J.
Vanhala . "Comparison of Center Estimation Algorithms for Heart and
Respiration Monitoring With Microwave Doppler Radar." IEEE Sensors Journal
12.3(2012):627-634.