Nan Li1,2, Feng Du1,2, Xiaoliang Zhang3, Xin Liu1,2, Hairong Zheng1,2, and Ye Li1,2
1Lauterbur Imaging Research Center, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China, 2Key Laboratory for Magnetic Resonance and Multimodality Imaging of Guangdong Province, Shenzhen, China, 3Department of Biomedical Engineering, State University of New York, Buffalo, NY, United States
Synopsis
The lower
natural abundance and gyromagnetic ratios of non-hydrogen nuclei require the
use of a highly efficient RF coil to optimize the quality of the image. In this
study, a novel triple-nuclear RF coil system capable
of 1H / 19F / 23Na imaging was designed. The coil consists of a triple-tuned
four-ring birdcage transmit coil and the tight fit local X-nuclei single-loop
receive coils. A range of bench tests were performed by using the
developed triple-nuclear coil. The performance evaluations showed that the coil
system was appropriately tuned, matched and decoupled
at 1H/19F/23NA frequencies when working simultaneously.
Introduction
Affected
by the lower concentration and the lower MR sensitivity compared to that of the
proton, imaging X-nuclei is challenging. Significant SNR enhancement in
X-nuclei magnetic resonance imaging can be obtained with the optimal RF coil
sensitivity [1-2]. The coil should provide homogenous B1 fields at the
resonance frequencies of both (1H) and non-proton nucleus, and preferably
acquired MRI imaging of multiple nucleus simultaneously, for the
most accurate localization,
shimming and measurement of metabolite concentration [3-4].
A
four-frequency 1H / 19F / 23Na/ 31P RF transceiver was developed in our
previous study [5]. Moreover, the efficient method to have a
homogeneous excitation and highly sensitive reception is to use
a volumetric coil for transmission (Tx) and receiver (Rx) coils for signal reception
[6]. In this study, a novel integrated triple-nuclear RF coil system which consists
of
a triple-tuned modified four-ring birdcage TX-coil and the tightly fit local multiple
nuclear single-loop (Rx) coils was designed for 1H / 19F / 23Na MR imaging
at 3T. The feasibility and performance of the coil were evaluated on the bench.
Good tuning, matching and decoupling performance was obtained from the bench
tests. The results indicate that it is possible to obtain the 1H/19F/23NA MR
images by using the proposed integrated coil system.
Method
The novel integrated triple-nuclear
RF coil system consist of a four-ring birdcage TX-coil and the tight fit local
RX coils was designed and shown in Fig.1. The four-ring quadrature birdcage TX-coil constructed with a
triple-tuned configuration (Fig1a). The configuration was consisted of three
connected mutually coupled birdcages with 8 legs. The inner structure of the
TX-coil configuration uses a low-pass birdcage with adjusted the values of capacitors
located on the inner legs for generation of the 23Na (33.9MHz) mode
corresponding to the respective Larmor frequencies at 3T. For the outer
high-pass configuration that added at each end of the inner birdcage, the 1H19F
double-tuned circuit were added on the excited ports of the outer end-rings for
generation of the 1H19F (120.6/128.2 MHz) frequency. The
high-pass and the low pass configuration of the triple-tuned
coil can be driven in quadrature mode at 3T. There was
some space at the back end of the coil to accommodate RF circuitry chain.
The outside of the plastic cylinder on which
the adhesive copper tape is attached. The coils are constructed with size 16 cm
in diameter. The copper strips width of the inner rings and end ring width were
1 cm, and 1.5 cm, respectively. The outer ring spacing was chosen as 23 cm, the
inner rings were 13 cm apart corresponding to the optimal length. The coil was
shielded with copper foils placed 3 cm away from the surface of the coil. The
local multiple nuclei single-loop (Rx) coils were with 5.5 cm in diameter as
shown in Fig1. (b) and (c).
The RF circuit diagram of the triple-tuned
birdcage TX coil was shown in Fig. 2. To drive the triple-tuned
birdcage coil in quadrature mode, outputs of equal amplitude with relative
phase offsets of 90 were generated using a quadrature hybrid for 23Na mode, and
using a wide-band power-divider combined with phase shifter for IH19F mode. Both
1H19F and 23Na cable traps were inserted. Further, low-pass filter was applied
to the inner 23Na low-pass birdcage coil, and high-pass filter on the outer
1H19F coils, to minimize coupling noise from adjacent coils. The circuit
diagram of the local Rx coils were depicted as shown in Fig. 3. The phantom (15
cm in diameter) was filled with 2.6 g/L NaCl and 1.234 g/L NiSO4. 6H2O was used
as the load. The quality factors (Qunload, Qload) of the coils was measured on
the bench using the network analyzer. S-parameters under loaded conditions were
measured In order to evaluate the match and the isolation of the channels in
different nucleus.Results
There are three resonance frequencies that will
produce a homogeneous B1+ field of the coil and can be used for 1H19F/Na
NMR. The triple-tuned four-ring birdcage TX coil and the local RX coil were
tuned and matched well to the resonance frequencies of 1H/19F/23NA, the
detailed measurement values were shown in Fig.4, as return losses of each channel
of the TX and Rx coil were less than -12 dB. The isolation between two ports
were better than -12dB for 1H, -19dB for 19F and -18dB for 23Na. The unload and
load Q factors in the bench-tests were shown in Table.1.Discussion/conclusion
The novel triple -nuclear RF coils system
consisting of a triple-tuned modified four-ring birdcage TX-coil and the tightly
fit local 1H/19F/23Na nuclei single-loop Rx coils was built for 3T MRI in this
study. The configuration was used for the involving excitation and acquisition
of 1H/19F/23Na signals. The performance evaluations through
the bench tests showed that the coil system was sufficiently matched and
decoupled. In the future work, the imaging performance of the coil will be evaluated
on system. Furthermore, more legs or struts in the coil will be used and ring
spacing will be optimized to provide better RF field homogeneity and efficacy.Acknowledgements
This
work was supported in part NSFC under Grant No. 81627901; National Key R&D Program of China, 2021YFE0204400;
the Strategic Priority Research Program of Chinese Academy of Sciences (Grant
No. XDB25000000); city grant RCYX20200714114735123.References
[1] Ha Y.H,
Choi. C. H, Shah N.J, et al. Development and Implementation of a PIN-Diode
Controlled, Quadrature-Enhanced, Double-Tuned RF Coil for Sodium MRI.
IEEE Trans. Med. Imaging. 2018;
37(7): 1626–1631.
[2] S. M.
Hong, C.H.Choi, et al. Design of a quadrature 1H/31P coil using bent dipole
antenna and 4-channel loop at 3T MRI. IEEE
Trans. Med. Imag., 2018; 37(12): 2613–2618.
[3] Murphyboesch,
J, R. Srinivasan, et al. Two configurations of the Four-ring birdcage coil for
1H imaging and H-decoupled P spectroscopy of the human head. J. Magn. Reson.1994;103(2):103-114.
[4] Y. Duan, B.S.
Peterson, et al. Computational and Experimental Optimization of a Double-Tuned
1H/31P Four-Ring Birdcage Head Coil for MRS at 3T.
J. Magn. Reson. Imaging. 2009; 29: 13-22.
[5]
N. Li, X. Yang, F. Du, et al. Simultaneous
acquisition of 1H/ 19F/ 23Na/ 31P MR imaging in Phantom at 3T using a quadruple-nuclear
RF coil system. ISMRM. 2021; 1790.
[6] Matthias M , Ruomin H , C. C. Jorge, et
al. Reducing signal-to-noise ratio degradation
due to coil coupling in a receiver array for 35Cl MRI at 9.4 Tesla: A comparison
of matching and decoupling strategies. Concepts.
Magn. Reson. Part B. 2018; 48B: e21383.