Jinmin Xu1, Huafeng Liu1, and HuiHui Ye1
1State key Laboratory of Modern Optical Science and Engineering, Zhejiang University, Hangzhou, China
Synopsis
A novel water-fat separated MRF approach was proposed to
minimize the known biases introduced by the B1+, B0 and water and fat partial
volume. By incorporating the water-fat separation approach in the framework,
the dictionary size could be greatly reduced with the known B0 and FF map.
Multiple maps of parameters including B1+, B0, FF and T1 and T2* of water and
fat can be acquired within 14s for one slice.
Introduction
Magnetic resonance fingerprinting (MRF) is an efficient technique
to simultaneously provides quantitative maps of different tissue parameters. Considering
the inevitable systematic errors, such as RF transmit field (B1+) and B0 field inhomogeneity,
a novel MRF sequence has been proposed to enable a simultaneous T1, T2*, B1+
and B0 detection1. However, there is another factor that may cause
the inaccuracy of parametric mapping — water and fat partial volume, and the
fat characterization is also considered important especially for some fat-related
diseases, such as fatty liver. There are two main approaches to solve this
problem: (a) Including the fat signal by the multi-component model and matching the temporal signal evolution by an exhausted dictionary approach. This kind of
method usually work with the dictionary compression to accelerate the matching2.
(b) Incorporating the Dixon method in the sequence to achieve water and fat
separation at first, then implementing the independent water and fat MRF matching
to avoid the large dictionary matching3,4. Inspired by these approaches,
we adapted the existed B0&B1-correction MRF sequence to achieve the fat characteration.
The multi-echo signals from Double TR blocks could be used to calculate the Fat
Fraction (FF) map and B0 map by three-Dixon method. With the known FF map and
B0 map, the dictionary size could also be greatly decreased during the
multi-component signal matching. Our proposed framework could enable the T1 and
T2* mapping for both water and fat and the reliable B1+ and B0 estimation
simultaneously.Methods
Fig.1a shows the B0&B1-correction
MRF sequence we used for our experiments. Basically it is an IR-spoiled GRE sequence
with various TRs and TEs. The sequence can be divided into one IR block and
multiple double TR block. The IR block is mainly used to efficiently map the T1
parameter and the double TR block that composes of a short TR (12 ms) and a
long TR (48 ms) is used to encode the B1+ information. Inside the double TR block,
there are four echoes for the B0 and T2* estimation. The first three echoes (TE1/TE2/TE3 = 2.4/13.2/24 ms) will be used as three-echo GRE signal to acquire the in-phase and opposed-phase
water and fat images for the Dixon method to calculate the FF map and B0 map. Fig.1b
shows the Flip angles of each TR. Spiral trajectory is designed for the sequence
with 36 interleaves at 1mm resolution for 220 mm
FOV, and the total acquisition time is around 14s/slice.
As Fig.2b illustrates, the B0 and FF values estimated from
the three echoes will be used as the known parameters to create water-fat
dictionary by Bloch simulation. The mixed signal S at echo time TE can be written
as:
$$S(TE)=M0[(1-FF)+FF*e^{(i2\pi f_{cs} TE)} ] e^{(i2π∆B_0TE)} $$
Where M0 represents the equilibrium magnetization, $$$ f_{cs}$$$ is the water-fat chemical shift. Noted that we
ignore the B1+ and relaxation effects and assume there is only one peak in fat
spectrum for a clear illustration. Over multiple excitation pulses, perfect spoiling was assumed for simulations.
Finally, the temporal signal evolutions obtained by the
proposed sequence were matched using the created dictionary. We adopted the SVD
compression approach to further accelerate the dictionary matching5.
We did the numerical simulations to validate our idea.
Fig.2a shows a simple water-fat phantom with an overlapping area between the
water and fat circles. Reasonable M0, T1, T2* are set for water and fat,
respectively. The T1 (from 200 to 2300 ms with a step size of 10ms), the T2* (from
50 to 600 ms with a step size of 5 ms), the B1+ (from 0 to 1.4 with a step size
of 0.1) are adopted for the dictionary creation and temporal signal evolutions.Results
Fig.3 shows the simulation results
of the proposed method. The Ground Truth (GT) and Results (Res) of different
parameters (B0, B1+, FF, T1 and T2* of water and fat) are shown in the table. The
B0 map and FF map is identical with the real values, and the B1+, T1 and T2*
mapping match well with the ground truths. The results further validate the
feasibility of our method. Discussion and Conclusion
Simulations suggest
that our method could enable the simultaneous mapping of B0, B1+, FF mapping, as
well as the T1 and T2* mapping for both water and fat. This preliminary
exploration has shown promising results with further real phantom and in-vivo
validation. The potential applications of the proposed approach may include the
liver and skeletal muscles, where fat plays an important role in the related
diseases. Acknowledgements
This work was
supported in part by National Key R&D Program of China (No:
2020AAA0109502), by the National Natural Science Foundation of China (No:
U1809204, 61525106, 61427807, 61701436), the Fundamental Research Funds for the
Central Universities.References
1. H Ye, Q Li, X Cao, et al. MR fingerprinting (MRF) incorporating simultaneous detection of RF transmit field and B0 inhomogeneity. Proceedings of
the 26th Annual Meeting of ISMRM, Paris,France,2018.
2. Marty
B, Carlier P G. MR fingerprinting for
water T1 and fat fraction quantification in fat infiltrated skeletal
muscles[J]. Magnetic Resonance in Medicine, 2019, 83(2).
3. Jaubert
O, Cruz G, Aurélien Bustin, et al.
Water–fat Dixon cardiac magnetic resonance fingerprinting[J]. Magnetic
Resonance in Medicine, 2020, 83(6).
4. Koolstra
K, Webb A G, Veeger T T J, et al. Water–fat separation in
spiral magnetic resonance fingerprinting for high temporal resolution tissue
relaxation time quantification in muscle[J]. Magnetic Resonance in Medicine,
2020, 84(2):646-662.
5. D.
F. McGivney et al. SVD Compression for Magnetic Resonance Fingerprinting in the
Time Domain[J]. IEEE Transactions on Medical Imaging, vol. 33, no. 12, pp.
2311-2322.