Arun Joseph1,2,3, Patrick Liebig4, Gabriele Bonanno1,2,3, Jin Jin5,6, Kurt Majewski7, Tobias Kober8,9,10, and Tom Hilbert8,9,10
1Advanced Clinical Imaging Technology, Siemens Healthcare AG, Bern, Switzerland, 2Translational Imaging Center, sitem-insel AG, Bern, Switzerland, 3Magnetic Resonance Methodology, Institute of Diagnostic and Interventional Neuroradiology, University of Bern, Bern, Switzerland, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Siemens Healthcare Pty Ltd, Brisbane, Australia, 6ARC Training Centre for Innovation in Biomedical Imaging Technology, The University of Queensland, Brisbane, Australia, 7Siemens AG, Munich, Germany, 8Advanced Clinical Imaging Technology, Siemens Healthcare AG, Lausanne, Switzerland, 9Department of Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 10LTS5, École Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland
Synopsis
Echo planar imaging is an important MR technique used
for different applications in both structural and functional imaging. It is
typically used with fat suppression to avoid phase inconsistencies induced by the
chemical shift of the fat signal. pTx pulses are used to mitigate the B1+
bias field occurring at higher fields strengths; typical pTx pulses do however
not suppress the fat signal. In this work, we designed a frequency-selective
water excitation pTx pulse to be used in 3D EPI acquisitions, enabling homogenous,
high-resolution whole-brain fat-free imaging at 7T.
Introduction
Echo planar imaging (EPI)1 has been an
important tool for accelerated imaging of the whole brain since its introduction.
Two- or three-dimensional segmented EPI acquisitions have been combined with different
contrast mechanisms and have found many applications in both structural and functional
imaging such as functional MRI (fMRI), diffusion-weighted imaging, T2*-weighted
imaging2-3. The EPI sequence typically samples k-space with a
bipolar readout which can result in phase in-consistencies due to the varying
direction of the chemical shift, e.g., from fat. Therefore, EPI is typically
used in conjunction with a fat suppression technique, e.g., spectral fat
saturation or frequency-selective water excitation pulses.
Imaging at higher magnetic field strengths is
advantageous because of a higher signal-to-noise ratio. Albeit this enables
higher resolutions4, transmit RF field (B1+)
inhomogeneities are introduced. Although there are some methods to mitigate B1+
inhomogeneities such as the use of dielectric pads, the most effective solution
is employing parallel transmit (pTx) pulses5 to obtain a homogenous excitation
within the entire field of view.
Here, we design a frequency-selective water excitation
pTx pulse to be used in 3D EPI acquisitions enabling homogenous, high-resolution,
whole-brain, and fat-free imaging at 7T.Methods
kT-points
6-type excitation pulses were designed using an interior
point optimizer
7. The optimizer was used to minimize the difference
between a target flip angle and the simulated excitation profile based on B1
+
and B0 maps acquired prior to the EPI sequence. To obtain a water excitation
pulse, the cost function of the optimizer was modified to also minimize the
flip angle at an off-resonance frequency of -1040Hz. Other constraints were specific
absorption rate (surrogate limit of 10 J/kg), hardware limitations (max RF
amplitude 200 V), eleven sub-pulses, and a maximum pulse duration of 10 ms. All
pulse optimizations were performed offline using MATLAB (The MathWorks Inc.,
Natick, MA) in conjunction with own
software calling IpOpt
8.
All measurements were performed at 7T (MAGNETOM Terra, Siemens
Healthcare, Erlangen, Germany) using an 8-channel transmit, and 32-channel
receive head coil (Nova Medical, Wilmington, USA). The first experiment was
performed on a phantom consisting of a bottle filled with baby oil within
another bottle filled with water. In addition, another bottle filled with NiO
2-doped
water was also placed in the coil to have sufficient coil loading. In the
second experiment, two healthy subjects were scanned for proof of concept after
obtaining written informed consent.
For all experiments, a prototype 3D EPI sequence in four different pulse
configurations was acquired:
- Proposed frequency-selective kT-points water excitation pulse (pTx Water)
- Non-selective kT-points
excitation pulse (pTx Water + Fat)
- Slab-selective
single-transmit sinc excitation pulse (1Tx Water + Fat)
- Slab-selective single-transmit
sinc excitation pulse in combination with a spectral fat saturation module (1Tx
Water)
Table 1 shows the scan parameters used for the acquisition. The images
obtained from the different pulse configurations were visually compared.
Results and Discussion
Figure 1 shows the frequency-selective water excitation pTx pulse
generated from the B0 and B1+ maps obtained from a volunteer. Figure
2 shows the corresponding simulated flip angle distribution on a volunteer for
frequency-selective water excitation; 0Hz representing the water component and
-1040Hz representing the fat component. The obtained flip angle distribution in
water is homogenous and consistent with the intended target (8˚) for most
regions of the brain while fat stays mostly un-excited (flip-angles up to 1° except at border). Figure 3 shows the acquired
images from the phantom experiment. No fat signal is visible in the images
obtained with the proposed water excitation pTx pulse, while small residual fat
signal can be seen in the single transmit acquisition with spectral fat
saturation. Figure 4 shows the comparison of 1Tx and pTx pulses on a volunteer.
In general, the acquisitions obtained with pTx pulses show a more homogenous intensity
distribution in comparison to the 1Tx acquisitions, notably in the cerebellum. Further,
images acquired with water excitation pTx pulses do not have fat signal as
indicated in the skull regions. Additionally, the image acquired with the
spectral fat saturation shows a change in contrast (white matter is darker)
which is potentially due to additional pulse power introducing magnetization
transfer effects. In future work, these acquisitions need to be validated in a
larger cohort including patients as well as fMRI validation experiments. Conclusion
We implemented a water excitation pTx pulse and
combined it with 3D EPI acquisitions for high-resolution, whole-brain imaging without
fat signal at 7T. Phantom and in vivo experiments showed a homogenous signal
intensity distribution and successful suppression of the fat signal in the
images. The designed pulse may improve future image quality in various applications
of structural and functional imaging. Acknowledgements
No acknowledgement found.References
- Stehling
MK, Turner R, Mansfield P. Echo-Planar Imaging: Magnetic Resonance Imaging in a
Fraction of a Second. Science. 1991; 254(5028): 43-50.
- Turner R,
Le Bihan D, Maier J, et al. Echo-planar
imaging of intravoxel incoherent motion. Radiology. 1990; 177(2): 407-414.
- Turner R,
Le Bihan D, Chesnicks AS. Echo-planar imaging of diffusion and perfusion. Magn Reson Med. 1991; 19(2): 247-253.
- Poser BA,
Koopmans PJ, Witzel T, et al. Three dimensional echo-planar imaging at 7 Tesla. NeuroImage 2010; 51(1):
261-266.
- Katscher
U, Börnert P. Parallel RF transmission in MRI. NMR in Biomed 2006; 19(3):
393-400.
- Cloos MA,
Boulant N, Luong M, et al. kT-points: short three-dimensional tailored RF
pulses for flip-angle homogenization over an extended volume. Magn Reson Med. 2012; 67(1): 72-80.
- Majewski K. Simultaneous optimization of radio
frequency and gradient waveforms with exact Hessians and slew rate constraints
applied to kT-points excitation. Journ Magn Reson, 2021; 326,106941.
- Wächter,
A., Biegler, L. On the implementation of an interior-point filter line-search
algorithm for large-scale nonlinear programming. Math. Program. 2006; 106:
25–57.