Safa Özdemir1, Efe Ilicak1, Lothar R. Schad1, and Frank G. Zöllner1,2
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 2Mannheim Institute for Intelligent Systems in Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
Synopsis
Magnetic
Resonance Electrical Properties Tomography (MREPT) technique is used to obtain
conductivity (σ) and permittivity (ε), using phase of the B1+
information. While spin echo and bSSFP based pulse sequences were shown to successfully
obtain phase images, they are either very slow or suffer from artifacts and
require multi-acquisitions. To overcome these limitations, in this work we
investigate the use of spiral trajectory-based sequences in MREPT. Our results
indicate that even for sub-second acquisition times, conductivity maps can be successfully
acquired via spiral trajectory-based sequence, and therefore can improve the utility
of MREPT techniques by shortening the acquisition time drastically.
Introduction
Magnetic
Resonance Electrical Properties Tomography (MREPT) aims to obtain electrical
properties, namely conductivity ($$$\sigma$$$) and permittivity ($$$\varepsilon$$$), which can be useful for diagnosing and
monitoring various diseases. However, in order to utilize this technique,
transceive phase ($$$\phi_{tr}$$$) should be known. In literature, spin-echo and
bSSFP based sequences are commonly used for obtaining $$$\phi_{tr}$$$1. However,
spin-echo based sequences suffer from lengthy acquisition time, while bSSFP
based sequences are susceptible to B0 inhomogeneity (banding artifacts) and
require multiple acquisitions to overcome that issue. Therefore, especially for
the challenging tissue types such as heart or lung, a fast and reliable pulse
sequence is required.
To this end, we investigate the utilization of
spiral trajectory-based sequences for the use in MREPT, since they were shown to provide high acquisition
speed, high SNR efficiency, low minimum TE, and robustness against motion
artifacts; and were successfully demonstrated in different rapid imaging
applications2,3. For
this purpose, we have developed a gradient-echo based spiral sequence and implemented
a reconstruction framework for obtaining conductivity images from the non-Cartesian
trajectory. We present phantom measurements to demonstrate the performance of
the spiral trajectory in conductivity imaging. Methods
For
sampling k-space uniformly, widely used Archimedean spiral trajectories4 were implemented as shown in Figure 1. To stay within the safety limits,
maximum gradient strength and slew rate were chosen 28 mT/m and 120 mT/m/ms,
respectively.
For the MRI
measurements, a gradient-echo based pulse sequence with spiral trajectory is
used (Figure 2) with the following parameters: TR/TE = 10/1.6 ms, FOV = 240mmx240
mm, in-plane resolution = 2mmx2mm, 2.5 mm slice thickness, 16 spiral
interleaves, 20 slices, 15° flip angle, 18 averages. It should be
noted that with these parameters, a single image can be obtained in 0.16 seconds
(6.25 frames per second).
For the
phantom experiment, a cylindrical z-independent phantom with a diameter of 16
cm and height of 20 cm was constructed. For background, agar-saline gel (20 g
agar-agar, 2 g/L NaCl, 1.5 g/L NiCl2), and for structures, saline solution (6
g/L NaCl, 1.5 g/L NiCl2) was prepared. Structures in the sample have diameter
of 3.5 cm. Experiments were conducted on 3T scanner (Magnetom Skyra, Siemens
Healthineers, Germany).
Due to the
nature of the spiral trajectory, the outer part of the k-space is less-densely sampled
than the central part. To recover the missing data points, iterative
self-consistent parallel imaging reconstruction (SPIRiT)5 framework was
utilized. This method uses Nyquist‐sampled calibration data to train a
calibration kernel, and then applies the kernel to synthesize unacquired data
samples. To obtain reconstructions efficiently from the non-Cartesian acquisitions,
image domain SPIRiT reconstructions were utilized and were solved via iterative
least squares (LSQR) method.
For obtaining conductivity images from the
reconstructions, 2D generalized phase based convection-reaction-diffusion equation MREPT (cr-MREPT) technique with artificial
diffusion term is used6. Conductivity images were generated for individual
receiver coil elements and then combined by simple averaging to obtain final
conductivity images. Results
In Figure 3, a magnitude image obtained via the spiral acquisition and
SPIRiT reconstruction is displayed in grayscale. In addition, conductivity
images obtained with different number of averages are shown. Though affected by
noise, even in the conductivity image obtained with one average structures
could be visualized. With increasing the number of averages, the SNR is
improved and structures can be seen more clearly.
In Figure 4, region masks are applied via visual inspection on the spiral magnitude image. Mean and standard deviation of conductivity
values for both structures and background region are reported for several
number of averages. Both structures have similar mean conductivity: With 18
averages, mean±std is 0.784±0.075 S/m for first structure and 0.785±0.106
S/m
for second structure. They provide distinct contrast with respect to the
background region (0.539±0.080 S/m, 18 averages).Discussion and Conclusion
As seen in Figure 4, for the first couple of averages standard deviation reduces, afterwards it stays almost constant. One of the possible explanations is that there are many different sources of error in MREPT technique and rather than noise induced, the standard deviation is affected heavily by these errors. In phase-based MREPT techniques, strong assumptions must be used, especially the transceive phase assumption7 or constant transmit B1 (B1+) magnitude1 assumption can affect final conductivity images significantly.
We have developed a spiral trajectory-based acquisition and
reconstruction framework to attain conductivity images rapidly and displayed promising
results with a phantom experiment. Our initial results indicate that robust
conductivity images can be obtained successfully even for sub-second acquisition times. While further
studies are warranted, this framework can enable conductivity imaging in
challenging tissue types, and can further improve its practical utility.Acknowledgements
No acknowledgement found.References
1. U. Katscher, D.-H. Kim, and J. K. Seo, “Recent progress and future
challenges in MR Electric Properties Tomography,” Computational and
Mathematical Methods in Medicine, vol. 2013, pp. 1–11, 2013.
2. J. G. Pipe and D. D. Borup, “Generating spiral gradient waveforms with a
compact frequency spectrum,” Magnetic Resonance in Medicine, 2021.
3. Z. Li, H. H. Hu, J. H. Miller, J. P. Karis, P. Cornejo, D. Wang, and J.
G. Pipe, “A spiral spin-echo MR imaging technique for improved flow artifact
suppression in T1-weighted postcontrast brain imaging: A comparison with
Cartesian turbo spin-echo,” American Journal of Neuroradiology, vol. 37, no. 4,
pp. 642–647, 2015.
4. J. G. Pipe and N. R. Zwart, “Spiral trajectory design: A flexible
numerical algorithm and base analytical equations,” Magnetic Resonance in
Medicine, vol. 71, no. 1, pp. 278–285, 2013.
5. M. Lustig and J. M. Pauly, “SPIRiT: Iterative Self‐consistent Parallel
Imaging Reconstruction from arbitrary k‐space,” Magnetic Resonance in Medicine,
vol. 64, no. 2, pp. 457–471, 2010.
6. N. Gurler and Y. Z. Ider, “Gradient‐based electrical conductivity
imaging using MR phase,” Magnetic Resonance in Medicine, vol. 77, no. 1, pp.
137–150, 2016.
7. A. L. H. M. W. van Lier, D. O. Brunner, K. P. Pruessmann, D. W. J.
Klomp, P. R. Luijten, J. J. W. Lagendijk, and C. A. T. van den Berg, “B1+
phase mapping at 7T and its application for in vivo electrical conductivity
mapping,” Magnetic Resonance in Medicine, vol. 67, no. 2, pp. 552–561, 2011.