Santhosh Iyyakkunnel1,2 and Oliver Bieri1,2
1Department of Radiology, University Hospital Basel, Basel, Switzerland, 2Department of Biomedical Engineering, University of Basel, Basel, Switzerland
Synopsis
Conductivity
mapping depends sensitively on the signal-to-noise ratio (SNR) of the transmit phase estimation. It is thus questionable, whether
conductivity mapping can be performed at low field. Due to reduced off-resonances, however, a possible solution for
the reduced SNR might be offered
by balanced steady state free
precession (bSSFP). Brain conductivity mapping with bSSFP was investigated at
0.55 T and appears to be feasible but besides SNR also the reduced curvature
of the transmit field becomes challenging.
Introduction
Conductivity mapping is based on an estimation of the
transmit phase, frequently assessed using the so-called transceive phase assumption1. Generally, the Laplacian calculus in the
reconstruction process requires
high SNR and therefore electrical properties tomography (EPT) is typically performed at
high field strengths. However, lower fields possibly provide many benefits for mapping
the conductivity of tissues: on
the one hand, the transceive phase assumption is more accurate due to reduced
scatter fields and, on the other hand, phase-only conductivity reconstruction
becomes a valid simplification due to the lower curvature of the B1+ amplitude1.
At high fields,
the transceive phase ($$$\varphi^{\pm}$$$) is usually measured with a spin echo. From its refocusing properties,
in principle, also balanced
steady state free precession (bSSFP) can be used2,3 but accurate
transceive phase estimation is
limited to its pass-band region. At low field, however, the pass-band region should expand over the whole
brain. In this work, we thus explore the feasibility of whole brain
conductivity mapping using bSSFP. Methods
For all presented measurements, 3D bSSFP scans were performed on a commercial clinical 0.55
T system (Magnetom
FreeMax, Siemens, Erlangen) using a 12-channel head coil for reception. Imaging parameters were: TE/TR
= 2.5 ms / 5 ms, flip angle of 40°, bandwidth of 465 Hz/px, matrix of 128×96×88, yielding an isotropic resolution of 2 mm.
Individual receive coil signals were combined according to the L1-least-squares
method suggested by Lee et al4. In addition, 6-fold long-term averaging was
performed to increase the
overall SNR, especially for the in vivo
measurement. Thus, the total scan could be acquired in a reasonable clinical scan time of just under 5 minutes.
A saline phantom (2 g/L Nacl/H2O) with an expected
conductivity of 0.34 S/m5 was used for validation purposes. The conductivity was reconstructed using
the phase only EPT equation1, $$$\Delta\varphi^{\pm}/(2\mu\omega)$$$, where $$$\mu$$$ is the magnetic permeability and $$$\omega$$$ the Larmor frequency. To adjust for errors at tissue
boundaries, the Laplacian was estimated by local parabola fitting6 with a
window of 7×7×7 voxels. The conductivity was then filtered with a
boundary-preserving median filter with a window of 17×17×17 using
the bSSFP magnitude6.Results
All bSSFP images were
essentially free of banding artifacts. Figure 1A and 1B show an example slice
of the bSSFP magnitude
and phase of the saline phantom measurement from a single scan and averaged over
6 scans. The corresponding reconstructed conductivity images are shown in Figure 2.
Restricting conductivity values between 0 and 1, over the whole phantom a mean conductivity
of (0.33 ± 0.13) S/m was found for the
single scan while the averaged measurement had a value of (0.35 ± 0.13) S/m.
Sagittal, coronal and axial bSFFP magnitude and phase images are shown in
Figure 3A and 3B, respectively. The corresponding conductivity is shown in
Figure 4. A region-of-interest (ROI) is indicated in the axial image for which an average brain
tissue conductivity of (0.39 ± 0.19) S/m was
found. The distribution of CSF conductivities peaks around the value 1 S/m.Discussion
At 0.55 T, bSSFP
provides enough SNR for subsequent conductivity mapping which is mainly due to its increased steady state
with decreasing field strength in combination with its property to deliver the highest SNR
per unit time among all imaging sequences7. SNR is crucial for proper EPT and even if the SNR at low field
can be matched with measurements at high fields with more scan time, the curvature of the transceive
phase poses an
additional challenge at low field as the Laplacian estimation is more prone to errors.
The estimated conductivity of the saline phantom agrees well with the expected
value of 0.34 S/m regardless of the averaging. The homogeneity however improved
considerably from 1 to 6 scans. Also in the in vivo brain good results are
obtained as the obtained value for average brain tissue is close to literature (0.30
S/m)8. CSF conductivity is found to be lower than the expected
value of 2.01 S/m8. The difference is mainly attributed to partial volume
effects and errors during the magnitude image based reconstruction with a large
kernel.Conclusion
In
conclusion, despite being challenged by the SNR and low curvature of the
transceive phase at low fields, conductivity mapping with bSSFP appears to be
feasible at 0.55 T within a clinically acceptable scan time.Acknowledgements
This work
was supported by the Swiss National Science Foundation (SNF grant No.
325230_182008).References
1. Lier
et al. Electrical Properties Tomography in the Human Brain at 1.5, 3, and 7T: A
Comparison Study. Magn. Reson. Med. 2014;71(1):354-363
2. Stehning et al. Real-Time Conductivity Mapping using Balanced SSFP and
Phase-Based Reconstruction. Proc. Intl. Soc. Mag. Reson. Med. 19; 2011. Abstract
no. 128
3. Stehning et al. Electric Properties Tomography (EPT) of the Liver in a
Single Breathhold Using SSFP. Proc. Intl. Soc. Mag. Reson. Med. 20; 2012. Abstract
no. 386
4. Lee et al. MR-based conductivity imaging using multiple receiver coils. Magn.
Reson. Med. 2016;76(2):530-539
5. Stogryn A. Equations for calculating the dielectric constant of saline
water. IEEE Trans. Microwave Theory Tech. 1971;19(8):733-736
6. Katscher et al. Estimation of Breast Tumor Conductivity using Parabolic Phase
Fitting. Proc. Intl. Soc. Mag. Reson. Med. 20; 2012. Abstract no. 3482
7. Bieri and Scheffler.
Fundamentals of balanced steady state free precession MRI. Magn. Reson. Med.
2013;38(1):2-11
8. Gabriel et al. The dielectric properties of biological tissues: III.
Parametric models for the dielectric spectrum of tissues. Phys. Med. Biol.
1996;41(11):2271–2293