Emre Kopanoglu1
1CUBRIC / Psychology, Cardiff University, Cardiff, United Kingdom
Synopsis
Safety models on scanners are unaware of the actual patient
position while excitation pulses are inherently position dependent. This study
investigates the effect of this positional mismatch on SAR estimation for
axial, coronal and sagittal slice orientations. The positional mismatch yields
up to 5.2-fold underestimation of peak local SAR. RF shimming and 2-spoke parallel-transmit pulses for axial slice orientation have reduced SAR-sensitivity to positional
mismatch with the worst-case underestimation being <2.0-fold whereas a reduced
SAR-sensitivity was not observed for coronal and sagittal slices. For extreme head
positions not represented in safety-models, axial RF shimming / 2-spokes parallel-transmit pulses
maybe beneficial.
Introduction
Safety models 1-3 on
scanners are unaware of the actual patient position while excitation pulses are
inherently position dependent (referred to as ‘positional mismatch’ throughout).
Patient position is affected by variations in head shape/size, padding
material alternatives, coil hardware specifications and circumstantial
variations (patient settling in etc.). Recent studies showed that resulting ‘positional
mismatches’ may have a considerable effect on SAR estimation 4,5, focusing on axial slice-selective pulses. Here, a library of 26,082 realistic pulses were designed to
investigate if certain slice orientations / pulse complexities are less
susceptible to SAR underestimation due to ‘positional mismatch’. Methods
Body model Ella 6 was simulated at 161
different positions (Figure 1a) with respect to a generic 8-channel parallel-transmit
(pTx) array at 7T, using Sim4Life (Zurich MedTech, Zurich, CH). Electromagnetic
fields were exported to Matlab (Mathworks Inc., Natick, MA, USA) where
small-tip angle (30-degrees) pulses were designed using Matching
Pursuit-guided-Conjugate Gradient 7-9 with
Tikhonov regularization (β=0.5) over power in each channel. Full
Q-matrices 2,3 were
used for SAR calculations, that were 10-gram averaged over cubical volumes 10. Omitted details
followed Ref 11.
Quadrature (single-channel) excitation, 1-spoke (RF
shimming), 2-/3-/4-/5-spoke pTx pulses were designed for axial, coronal,
sagittal slice-selective excitation at each position (Figure 1b). Pulses were
designed for two different regions-of-interest (ROIs) for coronal and sagittal
excitation (Figure 1c). A total of 26,082 pulses were designed. Each pulse was evaluated
at its actual position ($$$SAR_{actual}$$$) and using the centred body position as the ‘safety-model’
($$$SAR_{safety-model}$$$). The ratio $$$SAR_{actual}/SAR_{safety-model}$$$ is reported for peak
spatial (local) SAR (psSAR) and whole-head SAR (whSAR) to characterize how much
‘positional mismatch’ affects SAR estimations. Maximum intensity projections (MIP)
of the three-dimensional local SAR distributions are given. Effect of the ROI
on pulses were tested at p=0.05 using two-sample t-tests against the null
hypothesis that ROI does not affect mean SAR sensitivity to ‘positional
mismatch’.Results
Actual peak local SAR was up to 5.2-fold higher than that
estimated by the safety model (Figure 2). For axial, sagittal and coronal pTx-pulses,
the worst-case underestimation was 4.2-fold, 5.2-fold and 3.8-fold,
respectively. Quadrature excitation was invariant on slice orientation/location
as channel coefficients are fixed relative to each other (global flip-angle
scaling cancels in $$$SAR_{actual}/SAR_{safety-model}$$$), the worst-case underestimation
being 4.2-fold.
RF shimming and 2-spokes axial pulses have less SAR-sensitivity
to ‘positional mismatch’ than quadrature and pTx pulses with more spokes
(Figure 2). For coronal and sagittal pulses, there was no considerable difference
between pulses types.
Targeting a smaller ROI yielded reductions in SAR
sensitivity for coronal and sagittal slices (Figure 2). The reductions were
statistically significant in 6/10 comparisons.
‘Positional mismatch’ yielded up to 1.4-fold, 1.8-fold and
1.6-fold underestimation of whole-head SAR for axial, sagittal and coronal
pulses, respectively (Figure 3). The size of the ROI affected SAR sensitivity,
as 8/10 comparisons across the sagittal and coronal pulses were statistically
significant.
‘Positional mismatch’ may also lead to misinterpretation of
the location of the hotspots (Figures 4-5). When channel coefficients are the largest
for the coils closest to the head, the hotspots positions are correctly
estimated by the safety-model (e.g. Figure 4 sagittal) although the peak value
is underestimated. When distal coils are more heavily used, however, the actual
hotspot positions can vary drastically compared to the estimate (e.g. Figure 4
axial/coronal).
Worst-case SAR underestimation depended considerably on the
pulse type and did not always occur at the same head position (Figures 4-5).
Figure 5 shows that for the worst-case RF shimming and 2-spoke pulses while the
spatial distribution of local SAR was correctly estimated, the peak was underestimated, whereas both
the distribution and the peak were incorrectly estimated for quadrature,
4-/5-spokes pulses.Discussion
The results here are in agreement with Refs. 4,5 for overlapping cases: axial slice-selection via RF shimming 4,5 and multi-spokes pTx pulses 4.
Local SAR depends directly on channel coefficients of and proximity
to coils (as well as interactions between coils). The ‘positional mismatch’ may
cause misrepresentation of the head-coil proximity, leading to SAR
underestimation. Furthermore, it may cause the location of local SAR hotspots to
be estimated incorrectly, when distal coils have higher coefficients. This
latter issue may cause concerning hotspot overlaps if SAR hopping 12 is used.
This study shows that slice orientation has an impact on SAR-sensitivity to ‘positional
mismatch’. Importantly, axial RF shimming and 2-spokes pTx pulses suffer less
SAR-underestimation than other axial/sagittal/coronal pulses. This is due to
their self-correcting nature: coils closer to the head will likely have smaller
coefficients (compared to a centred position) to ensure that the flip-angle does
not exceed the target in their vicinity. This isn’t true for more spokes as individual
spokes target inhomogeneous distributions that complement each other, potentially
leading to large channel coefficients in coils closer to the head. Coronal and
sagittal pulses are not self-correcting either: with the coils being distributed
azimuthally and outside the planes of interest, the effect of changing coil-head
proximity is less direct and requires less self-correcting adjustments.
Conclusion
Axial RF shimming and 2-spokes pTx pulses suffer less
SAR-underestimation than other axial/sagittal/coronal pulses. Hence, such
pulses may be preferable for extreme/non-represented head positions.Acknowledgements
This project was supported by the Wellcome Trust
[204824/Z/16/Z].References
1. Eichfelder G, Gebhardt M. Local
specific absorption rate control for parallel transmission by virtual
observation points. Magn Reson Med 2011;66(5):1468-1476.
2. Graesslin I, Homann H, Biederer S,
Börnert P, Nehrke K, Vernickel P, Mens G, Harvey P, Katscher U. A specific
absorption rate prediction concept for parallel transmission MR. Magn Reson Med
2012;68(5):1664-1674.
3. Bardati F, Borrani A, Gerardino A,
Lovisolo GA. SAR optimization in a phased array radiofrequency hyperthermia
system. Specific absorption rate. IEEE Trans Biomed Eng 1995;42(12):1201-1207.
4. Kopanoglu E. Patient specific
parallel transmit pulses are patient position dependent while safety models are
fixed: safety implications. 2021; online. p 2299.
5. Ajanovic A, Hajnal J, Tomi-Tricot R,
Malik S. Motion and Pose Variability of SAR Estimation with Parallel
Transmission at 7T. 2021; online. p 2487.
6. Andreas C, Wolfgang K, Eckhart GH,
Katharina H, Marcel Z, Esra N, Wolfgang R, Rolf J, Werner B, Ji C, Berthold K,
Peter S, Hans-Peter H, Jianxiang S, Michael O, Dominik S, Anthony K, Joshua WG,
Niels K. The Virtual Family—development of surface-based anatomical models of
two adults and two children for dosimetric simulations. Physics in Medicine
& Biology 2010;55(2):N23.
7. Grissom W, Yip CY, Zhang Z, Stenger
VA, Fessler JA, Noll DC. Spatial domain method for the design of RF pulses in
multicoil parallel excitation. Magn Reson Med 2006;56(3):620-629.
8. Kopanoglu E. Near real-time
parallel-transmit pulse design. Proc. ISMRM and ESMRMB; 2018; Paris, France. p
3392. (Proc. ISMRM and ESMRMB).
9. Kopanoglu E, Constable RT.
Radiofrequency pulse design using nonlinear gradient magnetic fields. Magn
Reson Med 2015;74(3):826-839.
10. IEC/IEEE International Standard --
Determining the peak spatial-average specific absorption rate (SAR) in the
human body from wireless communications devices, 30 MHz to 6 GHz - Part 1:
General requirements for using the finite-difference time-domain (FDTD) method
for SAR calculations. IEC/IEEE 62704-1:2017 2017:1-86.
11. Kopanoglu E, Deniz CM, Erturk MA, Wise
RG. Specific absorption rate implications of within-scan patient head motion
for ultra-high field MRI. Magn Reson Med 2020;84(5):2724-2738.
12. Guerin B,
Adalsteinsson E, Wald LL. Local SAR reduction in multi-slice pTx via “SAR
hopping” between excitations. 2012.