Yongquan Ye1, Xiaodi Liu2, Ying Wu2, Zhongqi Zhang1, and Jian Xu1
1UIH America, Inc., Houston, TX, United States, 2United Imaging Healthcare, Shanghai, China
Synopsis
In this work, we evaluated the performance of a recently
proposed high resolution multi-parametric method, i.e. MULTIPLEX, on a
state-of-the-art 5T whole body scanner. Comparing to the 3T counterpart,
MULTIPLEX on 5T generally offers improved SNR, image quality and imaging
efficiency, while maintaining satisfactory performance on B1/T1 mapping.
Introduction
It has long been a challenge for most currently available
multi-parametric imaging methods to achieve satisfactory performance on ultra-high
fields (>3T) due to several major factors, such as the elevated specific
absorption rate (SAR) and peripheral nerve stimulation (PNS) level, longer T1
and shorter T2/T2* relaxometries, stronger B1+ inhomogeneity and chemical
shifts, to name a few. For instance, the increased SAR level prohibits faster
scans with spin echo based methods1,
the longer T1 requires longer TR and/or TI (thus longer scan times) for adequate signal
recovery1, and the shorter T2* effectively reduces the imaging resolution of EPI2 and SPIRAL3 methods through the intrinsic filtering effect during readouts.
Recently, a dual-TR, dual-FA, multi-echo GRE based
multi-parametric imaging method, namely MULTIPLEX4,
was proposed and demonstrated on 3T. In principle, GRE based methods can offer
better performance at ultra-high fields in terms of T1, PD, susceptibility, SNR and
scan times, while without limitations on imaging resolution or SAR levels. However,
the more severe B1+ field inhomogeneity may potentially hamper the accuracy of T1
mapping5.
Therefore, this work aims to evaluate the MULTIPLEX performance at a newly
introduced field strength of 5T6,
in comparison with well-controlled 3T results. Methods
High resolution MULTIPLEX brain images were collected on the
same healthy volunteer (male, 36/yro, passed safety screening and gave written consents) on a
5T whole body scanner (uMR Jupiter, United Imaging Healthcare, Shanghai, China) with a 48-channel
head coil, as well as on a 3T scanner (uMR790, United Imaging Healthcare, Shanghai, China) also with a 48-channel
head coil.
For this initial comparison tests, the same scanning protocol was used
on both fields, including following key parameters4:
7 echoes with TE = 3.05~23.24ms, TR1/TR2 = 7.54/32.46ms, FA1/FA2
= 4°/16°, matrix size = 272x231x72 and voxel size = 0.8x0.8x2mm3,
bandwidth = 260Hz/px. With 2x acceleration, the total scan time was 8:10.
Dicom images of the MULTIPLEX scans were directly reconstructed
by the integrated reconstruction pipeline on both scanners and can be readily
compared. Furthermore, raw k-space data of both scans were saved and processed
offline to evaluate the imaging signal-to-noise ratio (SNR) of individual
echo images. Results
The MULTIPLEX B1+ and T1 maps are shown in Figure 1. Other
typical MULTIPLEX images are compared in Figure 2, including augmented T1W
(aT1W) images, PD/R2*/QSM maps and SWI mIP images. ROI values from typical
tissues are shown and compared in Table 1. And the image magnitude, noise
distribution and SNR maps of the same MULTIPLEX echo (i.e. FA1-TR1-Echo1) from
both fields are compared in Figure 3.Discussion
In this work, we have presented the initial experience of MULTIPLEX scans on a state-of-the-art 5T whole body scanner, which were
compared to 3T results.
The 5T field strength has only recently become available for
whole body imaging6, thus no literatures on the relaxometry properties (e.g. T1/T2*) or other
quantitative quantities (e.g. PD/QSM) are readily available. However,
referencing the 3T and 7T literatures7,
one can safely assume that these MRI signal properties at 5T will likely fall
between 3T and 7T. By using an identical scanning protocol, reliable and similar outcomes
on both 5T and 3T should be expected.
The major high-field challenge in MULTIPLEX is expected to
be T1 mapping with B1+ correction, like any other dual-FA methods5. With the dual-TR design and actual flip angle imaging (AFI) technique8,
the B1 fields can be reliably extracted at both fields. As shown in Figure 1, the
B1 field inhomogeneity at 5T is indeed more severe than at 3T due to stronger
dielectric effects9.
Nevertheless, the 5T T1 map became homogeneous (Fig.1) and yielded reasonable
values (Table 1) after B1+ correction. However, some unevenness in 5T T1
maps can still be observed, most likely due to the higher order components in
the B1+ fields not being taken account for, and it will be our future focus to further improve the B1+ correction effect.
Table 1 also indicates that T2* values at 5T were lower than at
3T, which was expected. Proton density (or relative PD to be accurate) and QSM
values, on the other hand, remained relatively consistent between the two
fields, as in principle both quantities are field independent. And with the
same echo times, SWI at 5T showed stronger susceptibility weighted contrasts than its 3T
counterpart (Fig.2). These observations suggest the 5T MULTIPLEX results were generally better than 3T's, while well
within reasonable and expected range.
Generally, 5T results yielded better imaging quality and
higher SNR than 3T (Figs.1~3). This is clearly shown in Figure 3 that 5T images
had slightly higher signal intensity, lower noise level and subsequently higher SNR. For
future works, the MULTIPLEX protocol can be and should be optimized specifically
for 5T or even higher fields like 7T to achieve further optimized outcomes, such as
higher resolution and/or shorter scanning time. Also, more stringent tests are needed to validate the accuracy of MULTIPLEX's quantitative maps at
5T.
Conclusion
In conclusion, we have for the first time demonstrated the feasibility of MULTIPLEX imaging on a 5T whole body scanner, suggesting a strong potential of MULTIPLEX as a high field multi-parametric imaging solution. Acknowledgements
No acknowledgement found.References
1. Warntjes
JB, Leinhard OD, West J, Lundberg P. Rapid magnetic resonance quantification on
the brain: Optimization for clinical usage. Magn Reson Med 2008;60(2):320-329.
2. Wang F, Dong Z, Reese TG, et al.
Echo planar time-resolved imaging (EPTI). Magn Reson Med 2019;81(6):3599-3615.
3. Buonincontri G, Schulte RF,
Cosottini M, Tosetti M. Spiral MR fingerprinting at 7T with simultaneous B1
estimation. Magn Reson Imaging 2017;41:1-6.
4. Ye Y, Lyu J, Hu Y, Zhang Z, Xu J,
Zhang W. MULTI-parametric MR imaging with fLEXible design (MULTIPLEX). Magn
Reson Med 2021.
5. Deoni SC. High-resolution T1
mapping of the brain at 3T with driven equilibrium single pulse observation of
T1 with high-speed incorporation of RF field inhomogeneities (DESPOT1-HIFI). J
Magn Reson Imaging 2007;26(4):1106-1111.
6. Zhenhua Shen, Xuchen Zhu, Shihong
Han, et al. Initial Experience of Body Imaging at 5T. Joint Annual Meeting
ISMRM-ESMRMB & SMRT 30th Annual Meeting; 2021. p. 3120.
7. Wright PJ, Mougin OE, Totman JJ, et
al. Water proton T1 measurements in brain tissue at 7, 3, and 1.5 T using
IR-EPI, IR-TSE, and MPRAGE: results and optimization. MAGMA
2008;21(1-2):121-130.
8. Yarnykh VL. Actual flip-angle
imaging in the pulsed steady state: a method for rapid three-dimensional
mapping of the transmitted radiofrequency field. Magn Reson Med
2007;57(1):192-200.
9. Lutti A, Stadler J, Josephs O, et
al. Robust and fast whole brain mapping of the RF transmit field B1 at 7T. PLoS
One 2012;7(3):e32379.