Stefan Sommer1,2,3, Tom Hilbert3,4,5, Tobias Kober3,4,5, Natalie Hinterholzer2, Daniel Nanz2,6, and Constantin von Deuster1,2,3
1Siemens Healthcare AG, Zurich, Switzerland, 2Swiss Center for Musculoskeletal Imaging (SCMI), Balgrist Campus, Zurich, Switzerland, 3Advanced Clinical Imaging Technology (ACIT), Siemens Healthcare AG, Lausanne, Switzerland, 4Department of Radiology, Lausanne University Hospital (CHUV), Lausanne, Switzerland, 5LTS5, École Polytechnique Fédérale de Lausanne, Lausanne, Switzerland, 6University of Zurich, Zurich, Switzerland
Synopsis
T1 mapping of species with very fast
T2/T2* relaxation is impossible with conventional methods measuring echoes refocused
by gradients or radiofrequency pulses. We therefore present a new approach to estimate
T1 relaxation times of short-T2 tissue such as bone, tendon, or ligament by
sampling transversal magnetization at different inversion times with a short-T2
sensitive ultra-short-TE readout. The presented T1 mapping method is efficient and
B1-insensitive. By acquiring multiple echoes, it is possible to obtain T1 relaxation
times at short and long echo-time and therefore assess the influence of short-T2
components.
Introduction
Transverse
water-proton magnetization in densely packed tissues, such as ligament, tendon,
and cortical bone, exhibits fast T2* decay due to slow molecular re-orientation
and long correlation times. Sequences with center-out sampling trajectories and
ultra-short echo times (UTE) can still detect a signal from rapidly decaying
T2* in such tissue.
A quantitative
T1 estimation of short- and long-T2 tissue might allow for a better assessment
of tissue structure, composition or changes within pathology or during healing
processes [1].
The aim of this study was to explore the
potential of a magnetization-prepared 3D radial UTE prototype sequence at two
inversion times with a dual echo readout (MP2UTE) to map T1 relaxation times of
short-T2 components in a clinically feasible scan time; proof-of-concept is
shown on a knee and an ankle acquisition.Materials and Methods
Experiments
were performed on a 3T whole-body MR scanner (MAGNETOM Prisma, Siemens
Healthcare, Erlangen, Germany) equipped with a 1Tx/15Rx-channel knee coil (QED,
Quality Electrodynamics, Mayfield Village, OH, USA) and a 16Rx-channel
foot/ankle coil (Siemens Healthcare, Erlangen, Germany).
Magnetization-prepared
3D isotropic UTE imaging was performed with a prototype sequence that acquired
data in two distinct readout blocks at different inversion times similar to an
MP2RAGE acquisition [2]. Each readout block consisted
of several shots, each with a short non-selective excitation pulse followed by
two consecutive rewound center-out readouts and by a spoiler gradient (Figure
1).
MP2UTE
data was acquired in the knee and ankle of a healthy volunteer using an
adiabatic inversion pulse and sampling at two different inversion and echo times
(TI1: 700ms, TI2: 2500ms, TE1: 40µs, TE2: 2.46ms) with flip angles of 4° (TI1) and
5° (TI2), TR: 5s, inter-spoke TR: 5.8ms, 20’020 spokes (220 per readout block) in
TA: 7min 31s for each scan. RF-pulse duration: 40µs, dead time between RF and
ADC: 10µs, resolution 1mm isotropic for both scans. Prior to data gridding, the
k-space trajectory was corrected for gradient imperfections using a gradient
impulse response model [3]. Resulting complex
image data was combined to a unified image and T1 maps were calculated for each
echo time using numerical simulations based on sequence parameters as described
in Marques et al. [2]Results
Figure 2 shows magnitude images of the
knee for all four contrasts (two inversion times at two echo times) in the
three orthogonal orientations.
Figure 3 depicts the two T1 maps obtained from
the two inversion contrasts at short (TE1) and long (TE2) echo times and the
difference (T1 at TE1 – T1 at TE2) between the T1 maps for the knee scan.
Figure 4 shows the T1 relaxation times
from the ankle scan measured at TE1: 40µs and at TE2: 2.4ms respectively. The right column visualizes
the difference in T1 between the short (TE1) and long (TE2) echo times.
Most of the T1 differences show a lower
value in TE1 compared to the estimated value at TE2. Additionally, a lower T1
relaxation time can be found in tissue such as cortical bone, tendons,
ligaments, where short-T2 signal contribution is expected. Especially in the
knee (Fig 3), the interior of the tibia and femur also show a slight decrease
of T1. In contrast, in e.g., the calcaneus (Fig 4), the T1 estimation at TE1
(40µs) seems to be even
slightly increased.Discussion
A shorter T1 relaxation time obtained from
the short echo time in the interior of the tibia and femur might be caused by
the contribution of trabecular bone.
It is important to note that short-T2 signal
might not get completely inverted by the inversion recovery pulse but only
saturated as mentioned e.g. by Wei et al. [4] Therefore, in future work, the T1 model needs to be
adapted in order to better fit the underlying signal simulation and additional
work on optimizing the inversion pulse might improve the inversion of short T2
tissue. Nevertheless, through the additional data at two distinct echo-times,
it should be feasible to obtain quantitative data of T1 relaxation of short-
and long-T2 signal components.
Furthermore, due to the radial nature of
the readout, the k-space center is repeatedly acquired for each spoke and
therefore the T1 relaxation will vary across the read-out block. This might
lead to a T1 blurring effect. However, we did not find any indication in the resulting
T1 maps.Conclusion
The combination of
the MP2RAGE concept with a short-T2 sensitive readout allows for B1-insensitive
quantification of T1 relaxation times in tissue such as ligament, tendon, or
cortical bone. Furthermore, the extension to multiple echoes during the readout
enables the distinction between short-T2 and long T2 signal contributions to T1
relaxation times within a voxel. This new
ability to simultaneously study the microstructure of tissues in two
quantitative dimensions may lead to a deeper understanding and better
assessment of their composition as well as pathological processes.Acknowledgements
No acknowledgement found.References
[1] Bachmann,
E. et al. T1-and T2∗-Mapping
for Assessment of Tendon Tissue Biophysical Properties: A Phantom MRI Study. Invest.
Radiol. 54, 212–220 (2019).
[2] Marques, J. P. et al. MP2RAGE,
a self bias-field corrected sequence for improved segmentation and T1-mapping
at high field. Neuroimage 49, 1271–1281 (2010).
[3] Vannesjo, S. J. et al. Image
reconstruction using a gradient impulse response model for trajectory
prediction. Magn. Reson. Med. 76, 45–58 (2016).
[4] Wei, Z., Ma, Y. J., Jang, H., Yang,
W. & Du, J. To measure T1 of short T2 species using an inversion recovery
prepared three-dimensional ultrashort echo time (3D IR-UTE) method: A phantom
study. J. Magn. Reson. 314, 100–106 (2020).