Vahid Malekian1, Nadine N. Graedel1, Alice L. Hickling1, Ali Aghaeifar1,2, Nadège Corbin1,3, Oliver Josephs1, Eleanor A. Maguire1, and Martina F. Callaghan1
1Wellcome Centre for Human Neuroimaging, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom, 3Centre de Résonance Magnétique des Systèmes Biologiques, CNRS‐University Bordeaux, Bordeaux, France
Synopsis
Geometric
distortion is a major concern for cortical depth-dependent fMRI using EPI
readouts that suffer from low bandwidth in the phase-encoded direction. Distortions
are exacerbated at higher field strengths due to increased B0 field
inhomogeneity. In this study, we quantitatively compared two distortion
correction methods (B0 field-mapping and reversed-PE) applied to submillimeter
(0.8mm isotropic) 3DEPI data at 7T. Cortical alignment was evaluated through
comparison with an anatomically-faithful MP2RAGE reference by computing the
dice coefficient and normalised mutual information. Both distortion correction
methods usefully improve alignment. The reversed-PE approach
performed better.
Introduction
Echo-planar
imaging (EPI) is the most common fMRI readout but is prone to geometric
distortions along the low bandwidth phase-encoded (PE) direction [1]. This is exacerbated at 7T by
increased B0 field inhomogeneity and by the push to longer readouts necessary
for higher resolution in cortical-depth-dependent analyses. Multiple methods
have been suggested to reduce distortions, most notably based on estimating and
correcting voxel displacements via B0 field-mapping or reversed phase-encoding (reversed-PE)
[2-5]. However, comparisons of these techniques in the
context of fMRI have previously only been performed either at lower field or lower resolution.
Here we mitigated
geometric distortions at the point of acquisition by using in-plane
segmentation (factor 2) and parallel imaging (factor 4). 3DEPI was utilised
instead of 2DEPI to gain SNR [6] and avoid slice profile effects. Magnetisation
transfer (MT) weighting was added to boost contrast between tissue types. Data
were post-processed with both B0 field-mapping and reversed-PE correction
techniques, and the resulting cortical alignment evaluated against an
anatomically-faithful MP2RAGE reference.Methods
Data
acquisition: 21
participants were scanned at 7T (Siemens MAGNETOM Terra) using an 8 transmit
and 32 receive channel head coil (Nova Medical). Following
fMRI data acquisition (a memory recall paradigm), whole brain MT-weighted 3DEPI (MT-3DEPI) volumes
were acquired with reversed PE polarity (PA then AP) and an EPI readout that
matched the fMRI time-series. B0 field mapping and T1-weighted anatomical scans
were also acquired using dual-echo gradient-echo and MP2RAGE sequences
respectively (see Table 1).
Data
processing: All
quantitative analyses were performed in the anatomical space of the individual,
defined by their MP2RAGE. The common processing steps (Fig.1) included co-registration
of the MT-3DEPI to the MP2RAGE using boundary-based registration [7], segmentation in SPM12 [8], and parcellation into 48 grey
matter (GM) regions of interest (ROIs) based on the Harvard-Oxford cortical
atlas [9]. B0 field-mapping incorporated brain masking to exclude
edge voxels in the phase images and generation of the B0 field-map for unwarping
using FEAT as implemented in FSL [10] (Fig.1b). For the reversed-PE
correction, FSL’s topup and applytopup were used (Fig.1c).
Analysis:
The effect of distortion
correction was qualitatively assessed by boundary inspection and quantitatively by
computing the dice coefficient (DC), indicating the degree of cortical overlap,
and the normalised mutual information (NMI), indicating the degree of shared
information, between the MT-3DEPI and MP2RAGE. Results
Fig.2
shows the result of distortion correction for a representative individual. Both
B0 field-mapping and reversed-PE techniques showed improvement in GM boundary alignment (orange
arrows in prefrontal cortex). Reversed-PE outperformed B0 field-mapping (green
arrows in zoomed view). However, some distortions remained regardless of
technique (blue arrows).
Fig.3a
shows the DC for the 3 pipelines as a function of GM probability. Prior to correction, the DC was 0.771±0.021 (mean±standard
deviation over thresholds and participants). This increased to 0.786±0.021 and 0.805±0.019 following B0
field-mapping and reversed-PE corrections respectively. This indicates more GM
overlap between the MT-3DEPI and MP2RAGE data after correction, and more so for
the reversed-PE method.
Fig.3b
shows the relative improvement in NMI (corrected compared to non-corrected) for
the B0 field-mapping and reversed-PE techniques. The reversed-PE method led to widespread and stronger improvements, particularly in prefrontal regions.
In some regions, such as the central opercular, middle frontal and cingulate
gyri, B0 field-mapping marginally (<1%) reduced NMI.
The average field maps
(across participants, in Hz) estimated by each correction method are
qualitatively similar (Fig.4a). A notable discrepancy was visible in the
inferior frontal cortex, which suffered from signal dropout in the MT-3DEPI
data (Fig.4b) leading to substantial under-estimation of the off-resonance relative to
the B0 field-mapping approach. The mean (across participants) root-mean-squared
frequency distribution values were 39.45Hz and 31.17Hz for the B0 field-mapping
and reversed-PE methods, respectively (Fig.4c).Discussion
In
this study, we quantitatively compared B0 field-mapping and reversed-PE based distortion
correction in the context of sub-millimeter 3DEPI. DC and NMI revealed that
both methods improve alignment between MT-3DEPI and an anatomically-faithful MP2RAGE
reference. Furthermore, reversed-PE outperformed B0 field-mapping, consistent with
a recent study examining lower resolution 2DEPI at 7T [4].
B0
field-mapping performance may have been hindered near brain edges where phase changes are rapid and poor signal-to-noise ratio (SNR) affects phase-unwrapping. Furthermore, the field-mapping
data were acquired at lower resolution (2mm) and with limited contrast
(sacrificed for SNR) whereas the MT-3DEPI data had matched 0.8mm resolution and
enhanced contrast via an MT-prepulse. Nonetheless, B0 field-mapping
still reduced distortions in most cortical regions. However, our analyses indicate that the
reversed-PE approach is to be favoured.
MT-3DEPI is useful for post-processing of fMRI
data, most notably for co-registration and segmentation. The additional
acquisition time for the reversed-PE volume was approximately 1 minute 53s,
which is less than that required for the B0 field-mapping sequence. In
addition, the finding of improved distortion correction using reversed-PE
generalises to the case of typical fMRI data by integrating a reversed-PE
volume (~4s acquisition time) into the acquisition of restricted field-of-view
3DEPI time series data used for fMRI (data not shown).Acknowledgements
The Wellcome Centre for Human Neuroimaging is supported by core
funding from the Wellcome [203147/Z/16/Z]. E.A.M. is supported by a
Wellcome Principal Research Fellowship [210567/Z/18/Z].References
1. Jezzard, P. and S. Clare, Sources of distortion in functional MRI
data. Human brain mapping, 1999. 8(2‐3):
p. 80-85.
2. Fritz, L., et al. Comparison of EPI distortion correction methods at 3T and 7T. in Poster Presented at the Annual Meeting of
the Organization for Human Brain Mapping. 2014.
3. Yamamoto, T., et al., Quantitative Evaluations of Geometrical
Distortion Corrections in Cortical Surface-Based Analysis of High-Resolution
Functional MRI Data at 7T. Journal of Magnetic Resonance Imaging, 2021. 53(4): p. 1220-1234.
4. Schallmo, M.P., et al., Assessing methods for geometric distortion
compensation in 7 T gradient echo functional MRI data. Human Brain Mapping,
2021. 42(13): p. 4205-4223.
5. Holland, D., J.M. Kuperman, and A.M.
Dale, Efficient correction of
inhomogeneous static magnetic field-induced distortion in Echo Planar Imaging.
Neuroimage, 2010. 50(1): p. 175-183.
6. Huber, L., et al., Techniques for blood volume fMRI with VASO: From low-resolution mapping
towards sub-millimeter layer-dependent applications. Neuroimage, 2018. 164: p. 131-143.
7. Greve, D.N. and B. Fischl, Accurate and robust brain image alignment using
boundary-based registration. Neuroimage, 2009. 48(1): p. 63-72.
8. Acton, P.D. and K.J. Friston, Statistical parametric mapping in functional
neuroimaging: beyond PET and fMRI activation studies. European Journal of
Nuclear Medicine, 1998. 25(7): p. 663-667.
9. Desikan, R.S., et al., An automated labeling system for subdividing
the human cerebral cortex on MRI scans into gyral based regions of interest.
Neuroimage, 2006. 31(3): p. 968-980.
10. Jenkinson, M., et al., Fsl. Neuroimage, 2012. 62(2): p. 782-790.