Benson Yang1,2, Chih-Hung Chen2, and Simon J Graham1,3
1Physical Sciences, Sunnybrook Research Institute, Toronto, ON, Canada, 2Electrical and Computer Engineering, McMaster University, Hamilton, ON, Canada, 3Medical Biophysics, University of Toronto, Toronto, ON, Canada
Synopsis
Parallel
radiofrequency transmission (pTx) technology continues to demonstrate its
potential at addressing MRI safety concerns relating to patients with implanted
deep brain stimulation devices. One promising technique involves
electromagnetic simulation that determine optimized pTx inputs to generate a
safe mode of imaging surrounding the implanted device. However, in practice,
instrumentation uncertainty can impact the ability of the pTx system to accurately
produce the desired signals. The present work studied the safety implications of system uncertainty and system failure in a 4-channel pTx platform. The preliminary results showed that in a
worst-case scenario, temperature elevations that exceed MRI guidelines
are possible.
Introduction
Deep
brain stimulation (DBS) is a highly effective treatment option for patients
with certain neurological disorders1. Safe magnetic resonance
imaging (MRI) of DBS patients remains a challenge at 3 T because of the
increased risks of localized radiofrequency (RF) heating. At present, DBS
patient studies involving 3 T MRI have been limited to low RF power pulse
sequences that minimize the risks2. However, 3 T MRI is highly
desirable without this limitation. Parallel RF transmission (pTx) has recently shown
promise in addressing these safety concerns. By utilizing electromagnetic (EM) simulation,
an optimized set of RF signals can generate a “safe mode” of imaging
surrounding targeted regions that contain conductive impants3. With
the limited pTx MRI options available commercially, many researchers have explored
custom additions to existing MRI systems to demonstrate the benefits of the technology4.
In practice, accurate replication of the optimized RF signals can be
challenging, as instrumentation uncertainty can cause signal deviations, and in
the worst case, the safe mode could potentially be compromised. This issue
requires careful investigation. The present work thus evaluates system
uncertainty (RF transmission error) and failure (complete channel loss) effects
in a 4-channel pTx platform for 3 T MRI proposed by Yang et al.4, using EM simulations of localized specific
absorption rate (SAR) and temperature elevation.Methods
EM simulations were
conducted in FEKO (Altair Engineering Inc., Troy, USA) on a homogeneous head
model with implanted insulated copper wire and an exposed tip configured in a
patient-derived lead trajectory5, as shown in Figure 1. The head model
material properties include: an acrylic shell (relative permittivity = 2.5,
loss tangent = 0.01), grey matter tissue (conductivity = 0.69, relative
permittivity = 67), wire insulation (conductivity = 0, relative permittivity =
2.44) and a perfect copper wire conductor. Thermal simulations were conducted using
Penne’s bioheat equation, excluding blood perfusion and metabolic effects6.
The 4-channel pTx MRI safe mode parameters used in this work were previously derived
according to McElcheran et al.3. These safe mode parameters were then
scaled to produce a whole head SAR average of 1.79 W/kg that is within the low-end
range for high SAR MRI at 3 T7. Initially, a series of local SAR
simulations for a spatial position near the exposed wire tip were executed for
pTx system errors in transmitted RF phase and amplitude involving a single
channel. The investigation was bounded by the manufacturer datasheet
specifications for the actual pTx system hardware4 (phase: ±10º,
amplitude: ±0.5 dB). Next, a worst-case scenario optimization was set to determine
the maximum local SAR for any pTx system channel error combination in phase and
amplitude within a 1 cm cubic volume encompassing the wire tip and its
resultant temperature elevation. Last, the effect of pTx system channel failure
was studied by powering off each channel independently during safe mode and evaluating
the effect on temperature.Results
Figure 2 shows
the EM simulation result (percent change in local SAR) for errors in phase and
amplitude of RF transmission from one channel, with all other channels
operating correctly in pTx safe mode. The SAR levels were most sensitive to pTx
system errors in channels 2 and 3, with a maximum change (from safe mode
baseline SAR) of approximately 300% for both channels. Figure 3 shows the
simulated temperature rise for the worst-case pTx system error combination in
phase and amplitude. A maximum temperature rise of 1.48⁰C was found, exceeding the
current +1.0⁰C safety threshold for MRI at 1.5 T8. Figure 4
shows the simulated temperature change for pTx system channel failure. RF
signal failure in pTx system channels 2 and 3 resulted in a maximum temperature
rise of approximately 7.5 and 8.5⁰C, respectively, a result
that greatly compromised the prescribed safe mode. Consistent with the data
shown in Figure 2, failure of channel 2 or 3 produced the worst local SAR
increases. Discussion and Conclusion
The simulation showed
that temperature elevations were most likely to arise from RF coil elements
nearest to the wire tip, as expected, with RF intensity decreasing with
distance. In addition, it was evident that two of the four pTx channels made
major contributions to pTx MRI safe mode and thus were most sensitive to
transmission errors and failure, as shown in Figures 2 and 4. It is well
documented that RF shimming capabilities improve with higher system channel
count. While two channels were dominant in resolving safety effects and were
thus very sensitive to error, the other two channels were important for optimizing
B1 homogeneity (data not shown). Last, the simulated temperature
results indicated that in a worst-case scenario, system uncertainty and channel
failure produced concerning temperature increases that exceed the present
guidelines for imaging8. Additional simulations should thus be
conducted to investigate the safety impact of different pTx system
configurations, including those with higher channel counts.Acknowledgements
No acknowledgement found.References
[1] Fickman, “New Discoveries of
Deep Brain Stimulation Put It on Par with Therapeutics.” https://uh.edu/news-events/stories/2021/march-2021/03252021-nuri-ince-deep-brain-stimulation-parkinsons.php.
Accessed October 20, 2021.
[2] Davidson et al., “Three-Tesla
Magnetic Resonance Imaging of Patients With Deep Brain Stimulators: Results
From a Phantom Study and a Pilot Study in Patients.” Neurosurgery 2020.
[3] McElcheran et
al., Parallel radiofrequency transmission at 3 Tesla to improve safety in
bilateral implanted wires in a heterogeneous model. Magn Reson Med. 2017.
[4] Yang et al., “A Platform
for 4-channel Parallel Transmission MRI at 3 T: Demonstration of Reduced
Radiofrequency Heating in a Test Object Containing an Implanted Wire.” JMBE
2019.
[5] Wei et al., “Reducing
Radiofrequency-induced Heating in Realistic Deep Brain Stimulation Lead
Trajectories using Parallel Transmission.” ISMRM 2018
[6] Pennes, “Analysis of
tissue and arterial blood temperatures in the resting human forearm.” J Appl
Physiol. 1948
[7] Seo et al., “MRI scanner‐independent specific absorption rate measurements
using diffusion coefficients.” JACMP 2017.
[8] Medtronic Inc. MRI
guidelines for Medtronic deep brain stimulation systems 2015. http://mriquestions.com/uploads/3/4/5/7/34572113/dbs_medtronics_contrib_228155.pdf.
Accessed October 20, 2021.