Difei Wang1, Rüdiger Stirnberg1, and Tony Stöcker1,2
1German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany, 2Department of Physics and Astronomy, University of Bonn, Bonn, Germany
Synopsis
We investigated the reproducibility of
multiparameter
mapping (MPM) using multi-echo
skipped-CAIPI 3D-EPI at 3T. Compared to >15 minutes of FLASH-MPM, EPI-MPM
under 3 minutes achieved considerably high repeatability. In only one fifth of the total scan time of FLASH, the
resulting CoVs of EPI were only 1.2-2.2 times larger than FLASH for R1,
PD and MTsat, while the CoV of EPI for R2* was
even smaller. Minor differences of the observed parameter estimates can be
attributed to the intrinsic difference between EPI and FLASH
sequence timing.
Introduction
A fast multiparameter mapping (MPM) acquisition protocol using
multi-echo skipped-CAIPI 3D-EPI (SC-EPI) at 3T was previously
reported and had the potential to provide high-quality R1,
PD, MTsat and R2* maps at 1mm isotropic
resolution without obvious artifacts within 3 minutes1. In this
study, we investigated the scan-rescan repeatability of the
SC-EPI-MPM protocols by comparison to FLASH.Methods
The study was conducted on a Siemens (Healthineers) Skyra 3T system
using a 32 channel head receive coil. Five subjects were scanned
twice, by exiting and re-entering the scanner. One set of SC-EPI T1w,
PDw and MTw images was acquired with 4 TEs between 5.4ms and
29.1ms at 1mm isotropic resolution.
A 16-fold segmented 2x2z1
skipped-CAIPI sampling2 was employed, resulting in EPI factor=5.
Repetition time/flip angle were chosen as follows: TR/FA = 39ms/4°
for PDw and 39ms/25° T1w images. MTw contrast was
acquired at TR/FA = 54ms/6°. The total acquisition time is 2:55
[min:sec]. For comparison, we utilized an established MPM protocol
based on a modified vendor FLASH sequence with 6 TEs up to 14.76ms3. TR/FA = 18ms/4° for PDw, 18ms/25° for
T1w and 37ms/6° for MTw. The total acquisition time is
15:32
[min:sec]. One subject underwent an approximately time-matched
SC-EPI session with 5 averages of SC-EPI (TM-EPI).
All SC-EPI- and FLASH-MTw contrasts were acquired with
Gaussian-shaped RF pulse (2kHz, 500°) applied prior to excitation pulse using a rectangular RF pulse (2.4ms duration). B1
and B0 maps were acquired simultaneously using 3DREAM4
at 4mm resolution for flip angle correction1. All parameter maps
were calculated within the hMRI toolbox5 and normalized into MNI
space
using
DARTEL tools6
in SPM. For each parameter, subject-wise
coefficient
of variation (CoV) maps were calculated. The
root-mean-squared (RMS) CoV was calculated in various
regions
of interest (ROI) per subject.Results
Fig.1 shows axial slices of quantitative
parameter maps of SC-EPI and FLASH. Both show good soft tissue
contrasts. Fig.2 demonstrates CoV maps.
The CoV map of SC-EPI appears noisier than FLASH, same as the
parameter maps and weighted images. The CoV of FLASH for MTsat in WM
are overall smaller than that of SC-EPI, so is the CoV of SC-EPI for
R2* in WM. The distribution of RMS-CoV in
various ROIs are depicted in Fig.3. The CoV
boxplots show that deep brain or cortical structures and cerebral WM
share similar RMS-CoV across subjects, except for a few outliers.
CoVs of SC-EPI are roughly 1.5 times larger than that of FLASH, except for
R2*, where CoV of SC-EPI and FLASH are
comparable. The time-matched TM-EPI achieved smaller CoVs than FLASH,
except for MTsat, where CoV of TM-EPI and FLASH are comparable. Fig.4 shows that R1 estimates of
SC-EPI are higher than those of FLASH in both WM and GM. The same
holds for MTsat. PD estimates of SC-EPI and FLASH agree with each
other. SC-EPI tends to have higher R2* in GM
and smaller R2* in WM.Discussion
Our study showed that the proposed 1mm isotropic SC-EPI-MPM protocol
acquired within 3 minutes at 3T achieved a high reproducibility for
PD, R1, R2* and MTsat, in particular
when compared to the literature7,8. Considering SC-EPI took only
one fifth of the total acquisition time as the FLASH protocol
(roughly half of the SNR), the RMS-CoV ranged from only 1.2% to 2.9%
for PD (1.4-2.1times larger than FLASH), from 1.9% to 5.6% for R1
(1.4-1.9 times larger), from 2.87% to 11.7% for R2*
(0.58-0.96 times smaller than FLASH except for in CC) and from 4.37%
to 8.68% for MTsat (1.3-2.2 times larger). The time-matched SC-EPI
achieved comparable or smaller CoV for all parameters in most of the
ROIs.
PD estimates of SC-EPI and FLASH agreed with each other due to WM
calibration. R1 estimates of EPI were ~1.03 times larger
than that of FLASH, which was also translated into slightly elevated
MTsat estimates. R2* estimates of EPI were
~1.03 times larger than that of FLASH in deep brain GM structures and
~0.9 times smaller in WM. This WM-GM difference, in particular,
likely comes from the mono-exponential R2*
model9,10 and a bias by short and long T2*
components due to the different TE ranges of SC-EPI and FLASH11.Conclusion
We have investigated the scan-rescan reproducibility of the MPM
acquisition using multi-echo skipped-CAIPI 3D-EPI at 3T, by
comparison to the MPM protocol using a conventional FLASH sequence.
SC-EPI-MPM shows lower CoV of R2* values and
slightly higher CoV of the other parameters in most of the ROIs
acquired within 3 minutes. The time-matched session proves that with
enough SNR, SC-EPI can accomplish even higher reproducibility than
FLASH. The difference of R1 parameter estimates between
SC-EPI and FLASH was translated into PD and MTsat. The difference of
R2* estimates could be the result of the
T2* model and the different TE coverage. The optimized
SC-EPI-MPM protocol can be easily utilized in clinical research
considering the short scan time.Acknowledgements
No acknowledgement found.References
1. Wang D, Stöcker T, Stirnberg R. Fast Quantitative Multiparametric
Mapping using 3D-EPI with Segmented CAIPIRINHA Sampling at 3T. In:
Proc. Int. Soc. Magn. Reson. med.; 2020
2. Stirnberg R, Stöcker T. Segmented K‐space blipped‐controlled
aliasing in parallel imaging for high spatiotemporal resolution EPI.
Magnetic Resonance in Medicine 2021.
3. Callaghan MF, Lutti A, Ashburner J, et al. Example dataset for the
hMRI toolbox. Data Br. 2019.
4. Ehses P, Brenner D, Stirnberg R, Pracht ED, Stöcker T.
Whole‐brain B1 ‐mapping using three‐dimensional DREAM. Magn.
Reson. Med. 2019.
5. Tabelow K, Balteau E, Ashburner J, et al. hMRI – A toolbox for
quantitative MRI in neuroscience and clinical research. Neuroimage
2019.
6. Ashburner J. A fast diffeomorphic image registration algorithm.
NeuroImage 2007.
7. Cooper G, Hirsch S, Scheel M, et al. Quantitative Multi-Parameter
Mapping Optimized for the Clinical Routine. Frontiers in Neuroscience
2020.
8. Leutritz T, Seif M, Helms G, et al. Multiparameter mapping of
relaxation (R1, R2*), proton density and magnetization transfer
saturation at 3T: A multicenter dual-vendor reproducibility and
repeatability study. Human Brain Mapping 2020.
9. Weiskopf N, Suckling J, Williams G, et al. Quantitative
multi-parameter mapping of R1, PD*, MT, and R2* at 3T: A multi-center
validation. Front. Neurosci.
10. Weiskopf N, Callaghan MF, Josephs O, Lutti A, Mohammadi S.
Estimating the apparent transverse relaxation time (R2*) from images
with different contrasts (ESTATICS) reduces motion artifacts. Front.
Neurosci.
11. Juras V,
Apprich S, Szomolanyi P, Bieri O, Deligianni X, Trattnig S.
Bi-exponential T2*analysis of healthy and diseased Achilles tendons:
An in vivo preliminary magnetic resonance study and correlation with
clinical score. European Radiology 2013.