Juliane Peter1,2, Sebastian Gantz1,2, Leonhard Karsch1,2, Jörg Pawelke1,2, and Aswin Hoffmann1,2,3
1OncoRay – National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany, 2Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiooncology – OncoRay, Dresden, Germany, 3Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Dresden, Germany
Synopsis
MRI-based
proton beam visualisation in water has proven feasible in exploratory
irradiation experiments performed on a first research prototype
in-beam MRI system. Beam-induced convection was hypothesised to be
implicated into MR signal loss observed within the beam volume. In
this study, this
hypothesis was tested in liquid water-filled phantoms by suppression
of convection-induced motion using mechanical barriers and
temperature control of
water expansivity.
In
absence of convection-induced motion, no beam-induced signal changes
occurred, supporting the hypothesis that convection triggers local MR
signal loss during proton beam irradiation. The elucidation of the
exact mechanism of convection-induced signal loss requires further
investigation.
Introduction
On-line
magnetic resonance imaging (MRI) is expected to improve the targeting
accuracy
of proton therapy (PT) for moving tumours1.
In order to study the interdependence and capabilities of a hybrid
system integrating MRI and PT, a low-field open MRI scanner has been
integrated with a horizontal fixed proton research beamline at our
clinical facility². In first experiments, MRI-based proton beam
visualisation has proven feasible in fluid-filled phantoms at high
irradiation doses³ allowing image-based localisation of the beam in
water. Consequently, this method has been identified as a potential
candidate for quality assurance of isocentre alignment of imaging and
radiation system components required for accurate treatment planning
and dose delivery in MR-integrated PT.
It
was hypothesised that beam-induced
local heating concurrent with a decrease in water density results
in upwards-directed
convection-induced motion of water molecules, which
would in turn lead to the
MR signal loss observed within the beam volume³.
Experimental
testing of this hypothesis, however,
was still pending. The aim of this study was therefore to
experimentally assess the role of beam-induced convection in
MRI-based proton beam visualisation in liquid water phantoms.Methods
A
0.22 T open MRI scanner (MrJ2200, ASG Superconductors S.p.A., Genoa,
Italy) was combined with a fixed proton research beamline².
The
beam first traversed a range shifter before being stopped within the
liquid water-filled phantom positioned centrally in
the MRI scanner
(Fig. 1). For
sequence verification experiments, a cuboid tap water-filled phantom was used (Fig.
2a) which
was subsequently supplemented by a
capillary
insert in combination with horizontal chequerboard patterned flow
restriction plates
A and B (Fig. 2b-d)
to
mechanically suppress convection in this phantom4.
The capillaries were 15 mm long and had a diameter of 10 mm.
An
air-insulated tap water-filled flask phantom (Fig. 2e) was used to
irradiate the water at a temperature close
to 3.98
°C,
exploiting
the temperature dependence of the volumentric expansion coefficient
of water to suppress convection5. Imaging under simultaneous irradiation was repeated
after the phantom had assumed the
ambient temperature of the scanner overnight. Temperature
measurements at both irradiation timepoints were performed in a similar but
unirradiated phantom. All
phantoms
were irradiated for 20 s per image
acquisition, irradiation starting 15 s prior to imaging.
Proton
beam energies between 200 and 215 MeV at beam currents of 8 to 64 nA
corresponding
to dose rates of 660 to 5100 Gy/min were used. The applied doses
ranged from 180 ±
20
to 1400 ±
130
Gy. A
gradient echo (GE)-based
time-of-flight angiography (TOF-angio)
sequence (TE = 7 ms, TR= 19.2 ms, flip angle
= 60°, slice thickness = 10 mm, in-plane resolution = 1.18 ×
1.07 mm², total scan duration = 3 s) was first tested
for its ability to visualise the proton beam assessing
its beam energy and current dependence. Subsequently
it
was used
to test the convection hypothesis. The sequence was flow compensated
and a
40
mm thick pre-saturation slab was located below the imaging slice,
separated by a 10 mm gap. The
single-slice images crossed the beam volume horizontally.Results
Using
the TOF-angio sequence, the stopping proton beam was visualised as a
hypointense signature showing partial signal loss within the beam
volume in a uniform, free-floating
water phantom at room temperature. Moreover, the observed beam
signature was found to be beam current and energy dependent (Fig. 3).
Mechanical restriction
of water mobility using the
flow restriction plates
to selectively close every second vertical capillary in the
water phantom at room temperature resulted in an absence of signal
loss within the beam volume in the motion-restricted capillaries,
whereas signal loss persisted in the
motion-unrestricted capillaries
(Fig. 4). Irradiation of free-floating
water
at 5
± 1 °C also
resulted in an absence of pronounced signal loss within the beam
volume compared to irradiation at an
ambient
temperature of 28 ± 1 °C (Fig. 5).Discussion
The
TOF-angio sequence reproduced our previous
findings in which beam signatures were visualised
as hypointense regions in liquid water through inversion recovery
(IR)-GE sequences. The signatures resembled a planar proton dose
distribution in shape and showed a clear dependence on beam energy and current3.
Consequently, the TOF-angio sequence was
found suitable for testing
the convection hypothesis.
Mechanical as well as thermal
suppression of beam-induced convection both resulted in the absence
of local upthrust of water molecules responsible for local
beam-induced MR signal loss. In
the latter case, changes in the mass density of water upon
irradiation were prevented exploiting a
volumetric thermal expansion coefficient of close to zero at 5 ± 1
°C water temperature.Conclusion
Our
experimental results
reveal that convection
triggers the
local
MR
signal
loss observed
during
proton beam irradiation of
liquid water phantoms.
Consequently,
a translation of this method to humans is unlikely due to the prevailing restriction of convective motion within
the narrowly compartmentalised body. For
the development of MRI-based geometric quality assurance for hybrid
MR-integrated
PT
systems, the
elucidation of the exact mechanism of the convection-induced MR signal loss is the next step to be investigated.Acknowledgements
We
thank Julien Smeets and Erik van der Kraaij from Ion Beam
Applications S.A. (Louvain-la-Neuve, Belgium) and Andrea Serra from
ASG Superconductors S.p.A. (Genoa, Italy) for their support and the
fruitful discussions.References
1.
Hoffmann A, Oborn B, Moteabbed M, et al. MR-guided proton therapy: a
review and a preview. Radiat Oncol. 2020;15:1-13.
2.
Schellhammer S, Hoffmann A,
Gantz S, et al. Integrating a low-field open MR scanner with a static
proton research beam line: proof of concept. Phys Med Biol.
2018;63(23):23LT01.
3.
Schellhammer S. Technical feasibility of MR-integrated proton
therapy: beam deflection and image quality. Doctoral thesis.
Technische Universität Dresden. 2019.
https://nbn-resolving.org/urn:nbn:de:bsz:14-qucosa2-341326.
Accessed October 25, 2021.
4.
Domen S, Krauss A and Roos M. The problem of convection in the water
absorbed dose calorimeter. Thermochim acta. 1991;187:225-233.
5.
Schulz R and Weinhous M.
Convection currents in a water calorimeter.
Phys Med Biol. 1985;30(10):1093-1099.