Alex Barrett1 and Marc Rea1
1Medical Physics, Clatterbridge Cancer Centre NHS Foundation Trust, Liverpool, United Kingdom
Synopsis
Distortion
in MRI can have a significant impact on diagnosis and radiotherapy treatment
planning. We have 3D printed a large, low cost phantom that we used along with
a CT ground truth to measure 6 MR scanners geometric accuracy. We found
significantly more distortion for the 3T and mobile scanners tested which we
conclude means they would require a more careful evaluation of distortion if
they are to be used for radiotherapy applications and that care should be taken
for all scanners where specialist radiotherapy applications such as
stereotactic radiosurgery are used.
Introduction
Geometric
accuracy is important in MRI if pathology is to be accurately assessed and is
critical in radiotherapy planning where distortion in MR images can cause
significant deviation from the desired treatment dose1. Geometric distortion
is normally assessed by compared the measured distances between a grid of
control points with their know distances2. Most current phantoms used for
these purposes are either very expensive, bulky, or limited in their coverage.
Here we present a simple 3D printed design similar to Jafar et al3 and Wang
et al4 that meets all of the requirements of a good distortion phantom and
can be produced easily, accurately and cheaply. We also present the results of
measurements taken of 6 MR systems from two different manufacturers.Methods
A 3D lattice was designed and 3D printed that
consisted of 15 layers of 52 targets 8mm in diameter that where evenly spaced
20mm from one another in every direction. This was then encased in a Perspex
cylinder and filled with oil (see Figure 1) to avoid standing wave artefacts at
3T. The phantom was CT scanned at high resolution (0.4x0.4x0.6mm) to act as a
ground truth. MR scanning was done with 3D distortion correction enabled.
Figure 2 shows the processing used
to assess MR
distortion. First the phantom was scanned using three sequences (T2 TSE, T1 GRE
and T2 SPACE). A 3D normalised cross correlation (NCC) was performed between
the MR images and a reference target. The centre of each volume above a set
threshold was calculated to identify target positions to sub pixel accuracy.
This followed a similar processing method used by Jafar et al3. We then
performed a 6-DOF registration to the CT data and recorded the differences
between the MR and CT data points to form the distortion grid that could be
visualised in 3D or compiled into a metric.Results
Repeatability measurements showed the error in
the mean distortion was ~0.008mm and the maximum (98th percentile) error was
~4%. The results for the 6 scanners varied from an average of 0.33-0.92mm and a
maximum (98th percentile) of 0.88mm to 3.14mm (Figure 3). 3T and mobile
scanners suffered from up to 3x more distortion than the best 1.5T scanner
tested. There was significant variation between all of the 1.5T scanners
(excluding the mobile) tested with the worse having almost twice the distortion
of the best even though they were all 60cm “wide bore” scanners.Discussion
Jafar et al3 produced a phantom of similar
design, and used similar processing to measure mean distortion on a range of
scanners. They measured a significantly larger average distortion than we did
of between 1.1-1.8mm. This is likely due to the phantoms larger size and increased
distortion further from the isocentre and the use of 3D gradient echo sequences
which are more susceptible to susceptibility induced distortion.
Recent IPEM guidance5 suggests tolerances
for distortion for scanners used in radiotherapy planning. They recommend
considering two volumes of clinical significance, a 20cm diameter sphere for
brains and prostates and a 40cm for full FOV head and neck or pelvis scans.
They suggest distortion must be kept <2mm within a volume of clinical
significance for standard RT applications and <1mm for specialized RT
applications such as SRS.
Using the least distorted T2 SPACE sequence,
all but the mobile scanner had <2mm everywhere within the phantom. All
scanners except the Skyra 3T achieved <1mm everywhere within 10cm of
isocentre but all scanners had areas >1mm beyond 10cm from isocentre. One
limitation of this analysis is due to the size and shape of the phantom we
cannot assess distortion for all points within the clinically significant volumes.Conclusion
Static 1.5T scanners are likely to meet
distortion requirements for standard RT applications but careful evaluation of
distortion is needed if mobile, 3T or specialist RT applications are used.
The 3D printed phantom exceeded our
expectations in terms of accurate construction and sub millimeter
repeatability. FDM 3D printing allowed us to cheaply build a phantom much
bigger than is available with resin style printers but is susceptible to subtle
warping making a CT Ground truth essential. The nearly spherical control
targets at the intersections aided in target detection and no manual
intervention was required for any control points. The processing workflow was
largely automated producing graphs and contour plots (Figure 3) to help
understand how distortion will affect the images produced.
Work has already begun on creating a larger, easier to
print phantom that will enable all points within a 40cm volume of clinical
significance to be assessed. We also hope in the near future to further assess
the impact of shimming, off-isocentre positioning and distortion within more
clinically representative cases.Acknowledgements
No acknowledgement found.References
1. Pappas EP, Alshanqity M, Moutsatsos A, et al. MRI-Related Geometric Distortions in Stereotactic Radiotherapy Treatment Planning: Evaluation and Dosimetric Impact. Technol Cancer Res Treat. 2017;16(6):1120-1129. doi:10.1177/1533034617735454
2. McRobbie DW, Semple S. Quality Control and Artefacts in Magnetic Resonance Imaging. Institute of Physics and Engineering in Medic; 2017.
3. Jafar M, Jafar YM, Dean C, Miquel ME. Assessment of Geometric Distortion in Six Clinical Scanners Using a 3D-Printed Grid Phantom. Journal of Imaging. 2017;3(3):28. doi:10.3390/jimaging3030028
4. Wang D, Doddrell DM, Cowin G. A novel phantom and method for comprehensive 3-dimensional measurement and correction of geometric distortion in magnetic resonance imaging. Magn Reson Imaging. 2004;22(4):529-542. doi:10.1016/j.mri.2004.01.008
5. Speight R, Dubec M, Eccles CL, et al. IPEM topical report: guidance on the use of MRI for external beam radiotherapy treatment planning. Phys Med Biol. 2021. doi:10.1088/1361-6560/abdc30