Intravoxel incoherent motion (IVIM) model approximates the tissue perfusion as a form of pseudo-diffusion, extracted as fast diffusion component in diffusion weighted imaging (DWI). Despite the central role of tissue perfusion in the angiogenesis in cancer, the clinical application of IVIM is hampered due to the susceptibility to noise and tendency of overfitting bi-exponential decay function. Recent introduction of Bayesian algorithm significantly enhanced the robustness and accuracy of IVIM method, renewing its potential as a clinical tool. We therefore set out to examine current IVIM algorithms in the context of neoadjuvant chemotherapy on patients with breast cancer preceding the treatment.
1. Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis E, Laval-Jeantet M. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology. 1986;161(2):401-7.
2. Liu C, Liang C, Liu Z, Zhang S, Huang B. Intravoxel incoherent motion (IVIM) in evaluation of breast lesions: comparison with conventional DWI. Eur J Radiol. 2013;82(12):e782-9.
3. Bedair R, Priest AN, Patterson AJ, et al. Assessment of early treatment response to neoadjuvant chemotherapy in breast cancer using non-mono-exponential diffusion models: a feasibility study comparing the baseline and mid-treatment MRI examinations. Eur Radiol. 2017;27(7):2726-36.
4. Vidić I, Jerome NP, Bathen TF, Goa PE, While PT. Accuracy of breast cancer lesion classification using intravoxel incoherent motion diffusion-weighted imaging is improved by the inclusion of global or local prior knowledge with bayesian methods. J Magn Reson Imaging. 2019;50(5):1478-88.
5. Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018;68(6):394-424.
6. Barbieri S, Donati OF, Froehlich JM, Thoeny HC. Impact of the calculation algorithm on biexponential fitting of diffusion-weighted MRI in upper abdominal organs. Magn Reson Med. 2016;75(5):2175-84.
Table 1. Sensitivity – a comparison of effect gradients for IVIM-derived parameters from four fitting algorithms
Perfusion fraction (f), apparent diffusivity (D) and pseudo-diffusivity (D*) from each algorithm (n=17). The difference between Bayesian and non-Bayesian algorithms was calculated by Wilcoxon signed rank test with z statistics and p values. The Pearson’s correlation coefficient r and p values are also shown, and the gradient of the line of best fit was calculated for significant correlations. Figures in bold indicate statistical significant difference (p < 0.017).
Table 2. Precision – a comparison of coefficients of variation (CoVs) for IVIM-derived parameters from four fitting algorithms
The CoVs obtained from four fitting algorithms are shown. The left column shows the mean and standard deviation of CoVs (in percentage) (n=17). The right column shows p values of pairwise comparison of CoV between Bayesian method and non-Bayesian methods. Figures in bold indicate the lowest CoV among the algorithms and statistical significant difference (p < 0.017).
Table 3. Uniqueness – a comparison of correlations between IVIM-derived parameters in each fitting algorithm
The correlations of f & D, f & D* and D & D* in each algorithm are shown. The left column shows Spearman’s rho (ρ) of the pairwise comparison of parameters (n=17). The right column shows p values of Spearman’s rank correlation test. There were no significant correlations in f & D, f & D* and D & D* in all algorithms.
Figure 1. Effect gradient of the four fitting algorithms.
The scatter plots of IVIM-derived parameters (a) Bayesian f vs Segmented unconstrained (SU) f, (b) Bayesian f vs Segmented constrained (SC) f, (c) Bayesian D* vs SU D*, (d) Bayesian D vs nonlinear least squares (Free) D, (e) Bayesian D vs SU D and (f) Bayesian D vs SC D. Only significant Pearson’s r correlations (p < 0.017) with the line of best fit are shown. Each dot represents an IVIM-derived parameter of an individual patient.
Figure 2. Dot plots of coefficients of variation (CoVs) of the four fitting algorithms.
The differences in CoVs in (a) f, (b) D and (c) D* between the four fitting algorithms are shown in dot plots. Each dot represents the CoV within an individual patient. The error bar indicates the mean and standard deviation. The CoVs obtained from different algorithms were compared using within-subject ANOVA, followed by post hoc analysis with Bonferroni correction (p < 0.017, 3 multiple comparisons) to assess the difference between Bayesian algorithm and non-Bayesian algorithms.