Amira Trabelsi1,2,3, Megdouda Benamara4, Pierre Jomin1, Stefan Enoch1, Marc Dubois4, Martine Pithioux3, David Bendahan2, and Redha Abdeddaim1
1Aix Marseille Univ, CNRS, Centrale Marseille, Institut Fresnel, Marseille, France, 2Aix Marseille Univ, CNRS, CRMBM, Marseille, France, 3Aix Marseille Univ, CNRS, ISM, Marseille, France, 4Multiwave Imaging, Marseille, France
Synopsis
We
assessed electromagnetic behavior of metallic implant used for distal femur
fracture in typical MRI situation. |B1+| field, SAR and
temperature variations were computed at 3T using a surface coil on human
model for different positions of the implant relative to the coil. To validate the
simulation, we measured the E-field and compared it to simulated E-field. |B1+|
field maps showed an interesting augmentation near the implant. Both global SAR and local SAR levels proved
that it is possible to safely image bone repair. However, temperature elevation
near the tip of the implant was important and is to be considerate.
Introduction
Osteosynthesis is a
surgical procedure commonly used for complex fractures with the aim of repairing
broken bones. Titanium (Ti) implants are used in femur fractures to constrain
bone during the healing process1. MRI is a high-interest technique
because of its non-invasive aspect and its spatial and temporal resolutions2.
However, the presence of implants can be problematic as magnetic interactions
between the implant and radiofrequency (RF) field can cause tissue heating3,4.
Safety guidelines indicate that the global SAR (globSAR) must be under 2W/kg5.
Given that globSAR does not provide information related to potential local
hotspots6. Local SAR (locSAR) must be calculated in order to assess possible
heating near the tip of the implant7. Investigating temperature is
also an important topic of ongoing research that demonstrated that a globSAR of
2.5W/kg could lead to temperature elevation up to 9°C after 15 minutes of
exposure in phantom8. In the present study, we performed
electromagnetic simulations to assess the behavior of the femur implant at 3T.
More specifically, we quantified globSAR, locSAR averaged over 10g of tissues
(SAR10g), and computed temperature. A phantom study was performed to
compare and validate the results of simulations and experiments using E‐field
measurements.Methods
Numerical modeling based
on finite-difference-time-domain method was used to calculate the RF induced E-field
and |B1+| field distributions, compute globSAR and locSAR
at 3T with a transceiver flexible surface coil in a human model with a Ti implant
(CST Microwave Studio 2019). The female model from the human voxel family was
modified to include the femur implant and screws were added to have realistic simulation
(Fig. 1). A transient time-domain solver was used for thermal simulations9.
The core temperature was assumed to be constant, the initial body temperature
was set at 37°C and the ambient temperature was 20°C. The heat transfer coefficient
of 3.2W/m2/K was computed to estimate heat losses due to radiation
and convection on the skin. Temperature changes were computed for 15 minutes of simulation
time. Four different coil positions were investigated i.e. ventral, dorsal,
right, and left to conclude on the best setting for the safest MRI acquisition. As
part of the validation procedure, a cylindrical phantom was designed and
constructed to measure E-field distribution to compare to the simulation results.
The phantom was filled with a solution (σ=0.5 S/m; εr=78 at 128MHz) where
the implant was secured by Ti screws to the femur sawbone (Fig. 4). E-field probe
was immersed in the phantom and 2D maps were acquired. Measurements were
performed over the implant length (210mm) and width (27mm) with a 3mm step. The
E-field and |B1+| field distributions from simulations and
measurements were scaled to 1W of
input power.Results
As displayed in Fig.2, the simulated |B1+|field maps illustrated an increased field close to
the implant for both ventral (3.835±0.135µT) and right (3.505±0.126µT)
positions. The GlobSAR levels
were under 2W/kg for the different positions.
The lowest SAR averaged over 10g of tissue was identified for the left position
with a value equal to 1.127W/Kg. The |B1+|maximum Bone/√LocSAR10g ratio illustrated
in Table 1 indicated that the optimal coil performance was obtained for the
dorsal position (2.245µT/√kg/W). As plotted in Fig. 3, the largest ΔTmax
elevation near the implant after 15 minutes of continuous exposure to RF, was
significant for the right position (1.32°C). Temperature variations for the
left and dorsal positions were lower (0.42 and 0.93°C, respectively) and there
were no significant changes observed without the implant.Discussion
As MRI can be of interest in order to assess bone
repair, |B1+| field, globSAR, locSAR, and temperature changes near
metallic implants should be robustly quantified. The present globSAR and locSAR
measurements demonstrated that MRI in presence of a femur implant is possible
within the human safety recommendations. However, the temperature elevation measured
in the different conditions showed that this variation is to be considered. Overall,
these parameters helped us gain insights on the best position of the coil vs.
implant to safely monitor bone consolidation. GlobSAR and locSAR failed to
predict peak temperatures after RF exposure duration10,11. The robustness of SAR measurements is tightly linked to the accuracy of electrical and thermal
properties of tissues and their thermoregulation model. Directly measuring E-field
helps validate results of simulation because the probe is not affected by the
presence of implant making measurements accurate.Conclusion
This work showed a simulation case of RF application
in implant and metrics measurements to ensure respect of MRI safety guidelines at
3T. Both results from globSAR and locSAR showed that they are not reliable to
predict RF heating near the femur implant, this was exhibited by temperature
variation up to 1.32°C when SAR measurements respected the guidelines. Directly
measuring E-field in phantom provides fast and accurate validation of
simulation results. Our future work will include measuring temperature
variation in phantom with the implant at 3T using a fiberoptic temperature probe. Finally,
to state the safety of MRI on patients with implants, field strength, RF coil,
scan conditions, and implant specifications must be considered.Acknowledgements
This project has received funding from the European Union's Horizon 2020 research and innovation program under grant agreement No 952106 (M-ONE project); and from the Excellence Initiative of Aix-Marseille University - A*MIDEX, a french "Investissements d'Avenir" program under Multiwave chair of Medical Imaging.References
1. C Krettek, M Müller, T Miclau. Evolution of
Minimally Invasive Plate Osteosynthesis (MIPO) in the femur. Injury. 2001;32:3.
2. J Fritz, B Lurie, TT Miller, HG Potter. MR Imaging
of Hip Arthroplasty Implants. Radiographics.
2014;34:4:106 - 132.
3. Winter L, Seifert F, Zilberti L, Murbach M and
Ittermann B. MRI-Related Heating of Implants and Devices: A Review. J Magn Reson Imaging. 2021;53: 1646-1665.
4. Tokaya JP, Raaijmakers AJE, Luijten PR, Bakker JF,
van den Berg CAT. MRI-based transfer function determination for the assessment
of implant safety. Magn Reson Med.
2017;78(6):2449-2459.
5. International Electrotechnical Commission. Medical
electrical equipment. Part2: Particular requirements for the basic safety and
essential performance of magnetic resonance equipment for medical diagnosis.
Geneva, Switzerland: IEC; 2010.
6. Graesslin I, Homann H, Biederer S, et al. A
specific absorption rate prediction concept for parallel transmission MR. Magn Reson Med. 2012;68:1664–1674.
7. Powell J, Papadaki A, Hand J, Hart A, McRobbie D.
Numerical simulation of SAR induced around Co‐Cr‐Mo hip prostheses in situ exposed
to RF fields associated with 1.5 and 3 T MRI body coils. Magn Reson Med. 2012;68:960–968.
8. Muranaka H, Horiguchi T, Ueda Y, Usui S, Tanki N,
Nakamura O. Evaluation of RF heating on hip joint implant in phantom during MRI
examinations. Nihon Hoshasen Gijutsu
Gakkai Zasshi. 2010;66:725–733.
9. Pennes HH. Analysis of tissue and arterial blood
temperatures in the resting human forearm.
Appl Physiol. 1948;1:93-122.
10. Oberacker E, Paul K, Huelnhagen T, et al. Magnetic
resonance safety and compatibility of tantalum markers used in proton beam
therapy for intraocular tumors: a 7.0 Tesla study. Magn Reson Med. 2017;78:1533–1546.
11. Destruel A, O'Brien K, Jin J, Liu F, Barth M,
Crozier S. Adaptive SAR mass‐averaging framework to improve predictions of
local RF heating near a hip implant for parallel transmit at 7T. Magn Reson Med. 2019;81:615–627.