Benson Yang1, Maryam Arianpouya2, Aaron Loh3, Gavin Elias3, Artur Vetkas3, Can Sarica3, Brendan Santyr3, and Simon J Graham1,2
1Physical Sciences, Sunnybrook Research Institute, Toronto, ON, Canada, 2Medical Biophysics, University of Toronto, Toronto, ON, Canada, 3Surgery, University of Toronto, Toronto, ON, Canada
Synopsis
High SAR RF pulse sequences continue to be avoided for
patients with implanted deep brain stimulation (DBS) devices, due to the increased
risk of injury. Recent studies have shown that DBS lead management can impact electromagnetic
behavior and thus, the overall RF heating effects along the implanted
device. The present work studied two patient-based DBS lead trajectories using a commercial DBS device with MR-conditional labeling for full-body 3 T MRI, implanted in a
novel anthropomorphic phantom. The preliminary results showed temperature
elevations that exceed current imaging guidelines remains possible for high SAR RF pulse sequences.
Introduction
Magnetic
resonance imaging (MRI) safety of deep brain stimulation (DBS) patients at 3 T
remains a challenge. Although progress has been made, clinical researchers
continue to avoid radiofrequency (RF)-intensive pulse sequences, such as
spin-echo imaging, due to increased risk of injury1. It is well
documented that the DBS electrode tips are associated with the highest risk of
RF heating, but recent simulation studies have suggested potential risks at
other locations along a lead trajectory, and reduced risk reduction from careful
lead trajectory management2,3. New MRI safety experiments are needed
to investigate these issues. Furthermore, there has been recent technological
advancement in commercially available DBS devices, with the Medtronic Percept PC
device becoming the first of its kind to be granted MR-conditional labeling for
full-body 3 T MRI, in 20204. The present work thus involves a series
of MRI RF heating experiments with the Percept PC DBS device implanted in a
novel anthropomorphic phantom using two patient-based DBS lead trajectories4,5.
The preliminary results demonstrate that the potential risk of elevated
temperatures (exceeding present MRI guidelines6) still remains for
high specific absorption rate (SAR) RF pulse sequences, and that additional
research will be needed to achieve safe unrestricted 3 T MRI with such devices in
the future.Methods
MRI RF heating
experiments were conducted on a 3 T Siemens MRI system (Magnetom Prisma,
Siemens, Erlangen, GER) using the standard body coil for RF transmission and the
standard 64-channel head and neck coil for RF reception. Three MRI pulse
sequences were used, ranging from low to high RF power, as listed in Figure 1. RF heating experiments were conducted using two DBS lead trajectories that reasonably
represented the surgical procedures at Toronto Western Hospital (Toronto, CAN)
and Sunnybrook Health Sciences Centre (Toronto, CAN), in the anthropomorphic
phantom as shown in Figure 2. The MRI study was conducted over two imaging
sessions that tested each DBS trajectory on separate days. The Medtronic
Percept PC device (B35200, Medtronic, Dublin, UK) with DBS leads (3387,
Medtronic, Dublin, UK), approved for MR-conditional full-body 3 T MRI, was used
bilaterally and powered off throughout. Four fibre-optic temperature sensors (T1,
Neoptix, Quebec City, CAN) were used, secured at the tips of each DBS electrode,
the DBS lead loop geometry (i.e. the ear and skull blur hole) and the deep
surface of the implanted pulse generator (IPG). These secondary locations were
selected according to our previous studies1. Temperature
measurements (±0.2⁰C accuracy) were recorded at 1 Hz for the
full duration of each pulse sequence.Results
Figures 3 to 5 shows
the temperature elevations during echo planar imaging (EPI), magnetization
prepared rapid gradient echo (MPRAGE) imaging and turbo spin-echo (TSE) imaging,
respectively. Overall, there were minimal temperature increases for EPI and
MPRAGE imaging with maximum temperature changes of 0.3 ± 0.2⁰C
and 0.5 ± 0.2⁰C, respectively; whereas TSE imaging reached a maximum temperature
elevation of 1.3 ± 0.2⁰C for the DBS lead trajectories tested. The
latter result exceeded the current imaging guidelines of +1.0⁰C for
3 T MR-conditional imaging6. In general, the RF heating effects
along the DBS device were consistent for the RF pulse sequences tested over both
imaging sessions and were concentrated at the DBS electrode tip implanted in
the right hemisphere.Discussion and Conclusion
The preliminary
RF heating results for the MR-conditional DBS device tested indicated that high
SAR pulse sequences (exemplified by spin-echo imaging) remain a concern and
adhering to the manufacturer’s safety guidelines remains advisable for 3 T MRI.
The DBS device remained off for this study, which is the typical recommendation
at our institutions. However, the manufacturer (Medtronic) recommends operating
the device in “MRI mode”6. Future work will be required to
investigate this mode, along with other configurations and product models. The
experimental results also showed that temperature elevations in a bilateral lead
setup were concentrated to a single lead. This result is consistent with
previous experiments at our institution using other DBS devices1.
Furthermore, the DBS electrode tips remained most susceptible to RF heating, but
one secondary location showed a small temperature elevation (loop trajectory,
highest applied RF power, Figure 5). The elevation was below the present safety
guidelines, but further evaluation will be necessary as the present work only tested
two of very many possible DBS lead trajectories. Lastly, the presented results
cannot be generalized to all MRI systems – such that similar surgical
procedures and experiments may result in different MRI safety outcomes.Acknowledgements
No acknowledgement found.References
[1] Davidson et
al., “Three-Tesla Magnetic Resonance Imaging of Patients with Deep Brain
Stimulators: Results from a Phantom Study and a Pilot Study in Patients.” Neurosurgery
2020.
[2] McElcheran et
al., Parallel radiofrequency transmission at 3 Tesla to improve safety in bilateral
implanted wires in a heterogeneous model. Magn. Reson. Med. 2017.
[3] Golestanirad et
al., Local
SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: A
simulation study of the effect of extracranial loops. Magn. Reson. Med. 2017.
[4] Medtronic Inc., PERCEPT™ PC NEUROSTIMULATOR. https://www.medtronic.com/ca-en/healthcare-professionals/products/neurological/deep-brain-stimulation-systems/percept-pc.html 2020.
[5] Yang et al., “Technical
note: an anthropomorphic phantom with implanted neurostimulator for
investigation of MRI safety.” Med Phys. 2020.
[6] Medtronic Inc., MRI
guidelines for Medtronic deep brain stimulation systems. 2020. https://manuals.medtronic.com/content/dam/emanuals/neuro/M929535A063A_view_d1590591026928.pdf