Zainul Ihsan1, Michael Schwarz2, Mubashir Husain2, and Gregor Schaefers1,2
1MR:comp GmbH, Gelsenkirchen, Germany, 2MRI-STaR - Magnetic Resonance Institute for Safety, Technology and Research GmbH, Gelsenkirchen, Germany
Synopsis
Numerical simulation
was performed to evaluate tissue damage due to magnetically induced (MI) force
in 3T Magnetic Resonance (MR) environment for multi-configuration passive
medical devices. Static magnetic and mechanical simulations were employed to determine
the impact of MI force on the tissue surrounding the implant. The simulation
results were used to select the worst-case for passive medical devices and for
quantitative assessment of the tissue damage.
Introduction
A magnetically induced
displacement of a medical implant can cause hazard to a
patient inside an MR scanner. Furthermore, the displacement of the implant can even
lead to a fatal outcome1 especially if it consists of sharp
metallic surfaces or ferromagnetic materials. Currently, the standard test for
MR compatibility of passive implants consists in performing a qualitative assessment
based on the experimentally measured ratio of the MI force to the
gravitational force2. In general, the worst-case for multi-configuration
passive medical devices is selected based on a rational approach, namely by
considering the largest mass and the highest permeability, whereas the effect of
a sharp geometry may be omitted. The implant may be labelled as an MR-safe device
when the MI force is lower than the gravitational force. However, the entire impact
of the MI force on the biological tissues is not covered completely in the currently-available
physical test. It has been implicitly recommended in the current version of the
FDA-guidance3 that a criterion of acceptance should address the
tissue damage caused by the MI force, torque, and vibration. The experimental
measurement of tissue damage is highly complex to be realized; therefore, a numerical
simulation is proposed as a preliminary evaluation. In this contribution, a framework
for simulating the tissue damage by using a commercial numerical-software is
presented.Methods
Both static and mechanic solvers of the simulation
platform 3DS CST Studio Suite 2019 (Darmstadt, Germany) were employed to assess
the tissue damage. First, a static magnetic field was generated by modelling the
cylindrical magnet using the static solver as shown in Fig. 1. An implantable clip
with a predefined relative magnetic permeability (µr), that was located at the entrance of the cylindrical magnet, was
considered as an object under test in simulations. The force density calculated
from the interaction of the induced magnetic field and the clip was imported as
the field source of the mechanic solver. Secondly, the clip surrounded by tissue
was generated in the mechanic solver as shown in Fig. 2, where a displacement
boundary was defined at the bottom of the tissue. In this scenario, two clips
with different tip geometries were compared as illustrated in Fig. 3. It can be
seen that the clip 2 has a sharper tip than the clip 1, whereas the remaining
parameters were kept the same.Results and discussion
Simulation results shown in Fig. 4 were generated with µr = 500 to depict a significant effect of the MI force. The simulation result shows
that the clip with the sharper tips could generate a higher risk, which is
indicated by the larger displacement. Based on this simulation, the variation
of geometry should be considered in the worst-case analysis instead of just
selecting the largest mass. Since
the maximum stress on the tissue was calculated higher than the tensile stress
(the limit of irreversible damage), the tissue would have been damaged. Based on this study, the tissue damage information can
be assessed by calculating displacement and comparing the calculated maximum stress
with the tensile stress. However, further mechanical properties of the tissue, which
show the limit of hazard, pain and sensitivity of the organ should be thoroughly investigated. Furthermore, the
safety labelling for MR compatibility should be evaluated by considering tissue
damage caused by the MI force, torque and vibration. Finally, the validation of
the simulation based on an experimental test is required.Conclusion
A framework for predicting the tissue damage is
proposed as a procedure to assess the worst-case MI force for multiple-configuration
passive medical devices, which is not covered by the conventional test. The
effect of different geometry in the product matrix could be determined
numerically and the tissue damage could be estimated.Acknowledgements
No acknowledgement found.References
- R P Klucznik, DA Carrier, R Pyka, RW Haid. Radiology, Vol 187, 855-856, 1993.
- American Society for Testing and Materials. Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, 2015.
- U.S. Food and Drag Administration. Testing and Labelling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, May 2021.
- Kuthe CD, Uddanwadiker RV, Ramteke A. Experimental evaluation of fiber orientation based material properties of skeletal muscle in tension. Molecular & Celluler Biomechanic. 2014 Jun;11(2):113-28.
- Suzan C.D., Erol C, Yunus Z.A. A
review of finite element applications in oral and maxillofacial biomechanics. Journal
of Mechanics in Medicine and Biology Vol. 18, No. 2, 2018.
- http://www.matweb.com. Accessed November 10, 2021.