Irène Brumer1, Dominik F. Bauer1, Lothar R. Schad1, and Frank G. Zöllner1
1Computer Assisted Clinicial Medicine, Mannheim Institute for Intelligent Systems in Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
Synopsis
Synthetic
kidney ASL data with respiratory motion was generated using models from the
XCAT phantom and matching recommendations for in-vivo acquisitions. Both pCASL
and PASL datasets with 1 M0 and 25 control-label pairs were created and
analysed using an in-house developed processing pipeline including
registration, manual segmentation, calculation of mean perfusion-weighted image
and perfusion map. The registration performed well on the synthetic data and
the perfusion maps yielded good cortex/medulla contrast. The presented method
allows a wide range of parameter choices for creating synthetic ASL datasets valuable
for testing processing pipelines and comparing them across research and
clinical imaging centres.
Introduction
Perfusion
imaging and quantification is important for diagnosing and monitoring numerous
pathologies in different organs1-3. Compared to other modalities and techniques, arterial
spin labelling (ASL) presents the advantage of being completely non-invasive1. However, the
acquisition and processing of the data can be challenging, especially in the
presence of motion. Because of these challenges, simulated data mimicking
in-vivo acquisitions is of great value to test processing pipelines or even
compare pipelines from different imaging centres. The aim of this project was
to generate synthetic ASL datasets of the kidneys simulating in-vivo
acquisitions for pipeline evaluation and comparison purposes.Methods
Starting
from a model of the XCAT phantom4, MR images were generated using a spin echo sequence with
TR=5000ms, TE=23ms, coronal-oblique slice positioning, and voxel dimension
3x3x5mm3. These settings were chosen to match recommendations for
in-vivo acquisitions2.
Literature values were used for tissue specific parameters (proton density, T1,
T2) of organs in the abdomen for a field strength of 3T5-7. The field of view of
the generated MR images was adapted to in-vivo ASL acquisitions and Rician
noise was added to reproduce signal-to-noise ratio similar to in-vivo
acquisitions. The respiratory motion during a free breathing acquisition was simulated
by generating 100 MR images at equally spaced time points around the exhalation
part of the breathing cycle. The generated proton-density weighted MR images
with respiratory motion were used as basis for the M0 and multiple
control-label pairs. Matching previous in-vivo acquisitions, we chose 25 control-label
pairs, resulting in 51 images randomly selected from the available 100 time
points. Background suppression used for control and label images was modelled
by reducing the signal intensity of all control and label images to 20% of the
signal of the M0 image (Figure
1). A perfusion ratio of 5 was assumed between cortex and medulla8. The general kinetic
model9 was
used to create both pseudo-continuous ASL (pCASL) and pulsed ASL (PASL)
datasets, assuming arterial transit times of 1123ms and 1141ms for the medulla and
cortex10,
respectively. Single-slice synthetic ASL datasets were then analysed using our
in-house developed processing pipeline running in MATLAB 2020a (The MathWorks, Inc., Natick,
Massachusetts, USA). Rectangular masks were manually drawn
on the M0 image to register left and right kidneys separately (Figure 2(a)). A
multi-resolution groupwise parametric registration was used to register all 51
images of the dataset using the Elastix toolbox11,12. The
quality of registration was evaluated by looking at line profiles across the
time dimension of the ASL dataset as well as calculating mean structural
similarity index measures (MSSIMs)13 for all possible image pairs of the dataset. The
mean perfusion-weighted image was calculated by pair-wise subtraction of
control and label images followed by averaging. The perfusion was quantified using
the following equations:
for pCASL data $$rbf \ [mL/100g/min] = \frac{6000 \cdot \lambda \cdot \Delta M \cdot e^{-PLD/T_{1b}}}{2 \cdot \alpha \cdot 0.93^2 \cdot M0 \cdot T_{1b} \cdot (1-e^{-\tau/T_{1b}})}$$
for PASL data $$rbf \ [mL/100g/min] = \frac{6000 \cdot \lambda \cdot \Delta M \cdot e^{-TI/T_{1b}}}{2 \cdot \alpha \cdot 0.93^2 \cdot M0 \cdot TI_1}$$
A blood
relaxation time $$$T_{1b}$$$ of 1650ms, a blood-tissue partition
coefficient $$$\lambda$$$ of 0.9mL/100g, a labelling efficiency $$$\alpha$$$ of 0.85 and 0.95 for pCASL and PASL, respectively,
and two background suppression pulses after labelling were assumed2. For the pCASL dataset, the
post-labelling delay $$$PLD$$$ was 1200ms and the labelling duration $$$\tau$$$ was 1600ms. For the PASL dataset, the
inversion time $$$TI$$$ was 1800ms and the labelling duration $$$TI_1$$$ was 1200ms. Segmentation was performed
manually on the registered M0 image (Figure 2(b)).Results
The
M0, first label and first control images of a single-slice synthetic pCASL and PASL
datasets are shown in Figure
1. The calculated perfusion-weighted image and
the perfusion map for the same datasets are shown in Figure 3. Mean
renal blood flow values were 218+/-51mL/100g/min and 209+/-48mL/100g/min for
the left and right kidney respectively for the pCASL dataset and 166+/-49mL/100g/min
and 167+/-49mL/100g/min for the left and right kidney respectively for the PASL
dataset. Horizontal and vertical line profiles for
both kidneys of the pCASL dataset are shown in Figure 4 and MSSIMs for each image pair of the
pCASL dataset are shown in Figure
5. For the pCASL dataset, MSSIMs ranged between 0.05 and 1 before and between 0.42 and 1 after registration. For the
PASL dataset, MSSIMs ranged between 0.05 and 1 before and between 0.39 and 1 after registration.Discussion
Synthetic ASL datasets of the kidneys simulating
in-vivo acquisitions were successfully generated. The analysis performed with
our in-house developed processing pipeline yielded good cortex/medulla contrast
and the registration performed well on the synthetic dataset as indicated by higher
MSSIMs calculated after registration. MSSIMs always were lowest for
comparison of controls and labels with M0 as is to be expected because of the
background suppression. Beyond the parameters chosen for this abstract,
the described method can be used to create ASL datasets with different acquisition
sequences, TR, TE, voxel dimension, noise level, number of control-label pairs,
$$$PLD$$$/$$$TI$$$, $$$\tau$$$/$$$TI_1$$$,
and level of breathing motion to further test limitations of processing
pipelines.Conclusion
Synthetic
ASL datasets of the kidneys mimicking in-vivo acquisitions have been created
and used to test our in-house developed processing pipeline.Acknowledgements
This project was supported by
the German Federal Ministry of Education and Research (BMBF) under the funding
code 01KU2102, under the frame of ERA PerMed. This research project
is part of the Research Campus M2OLIE
and funded by the German Federal Ministry of Education and Research (BMBF)
within the Framework "Forschungscampus:
public-private partnership for Innovations" under the funding code
13GW0388A.References
1.
Essig M, Nguyen TB,
Shiroishi MS, et al. Perfusion MRI: the five most frequently asked clinical
questions. Am J Roentgenol. 2013;201(3):W495-W510.
2.
Nery F, Buchanan CE, Harteveld
AA, et al. Consensus-based technical recommendations for clinical translation
of renal ASL MRI. Magn Reson Mater Phy. 2020;33(1):141-161.
3.
Andescavage N and Limperopoulos
C. Emerging placental biomarkers of health and disease through advanced
magnetic resonance imaging (MRI). Exp Neurol. 2020;347:113868.
4.
Segars WP, Sturgeon G,
Mendonca S, et al. 4D XCAT phantom for multimodality imaging research. Med
phys. 2010;37(9):4902-4915.
5.
Wissmann L, Santelli C,
Segars WP et al. MRXCAT: Realistic numerical phantoms for
cardiovascular magnetic resonance. J Cardio Magn Reons. 2014:16(1):1-11.
6.
De Bazelaire CM, Duhamel GD,
Rofsky NM et al. MR imaging relaxation times of abdominal and pelvic tissues
measured in vivo at 3.0 T: preliminary results. Radiol. 2004;230(3):652-659.
7.
Stanisz GJ, Odrobina EE, Pun J,
et al. T1, T2 relaxation and
magnetization transfer in tissue at 3T. Magn Reson Med. 2005;54(3):507-512.
8.
Roberts DA, Detre JA,
Bolinger L, et al. Renal perfusion in humans: MR imaging
with spin tagging of arterial water. Radiol. 1995;196(1):281-286.
9.
Buxton RB, Frank LR, Wong EC, et
al. A general kinetic model for quantitative perfusion imaging with arterial
spin labeling. Magn Reson Med. 1998;40(3);383-396.
10. Kim DW, Shim
WH, Yoon SK, et al. Measurement of arterial
transit time and renal blood flow using pseudocontinuous ASL MRI with multiple
post‐labeling delays: Feasibility,
reproducibility, and variation. J Magn Reson Imaging. 2017;46(3);813-819.
11. Klein S, Staring M, Murphy K, et al. Elastix: a toolbox for intensity-based
medical image registration. IEEE. 2009;29(1):196-205.
12. Shamonin DP, Bron EE, Lelieveldt BP, et al. Fast parallel image
registration on CPU and GPU for diagnostic classification of Alzheimer's
disease. Front Neuroinform. 2014;7:50.
13. Wang Z, Bovik AC, Sheikh HR, et al. Image quality assessment: from error visibility to structural similarity. IEEE.
2004;13(4):600-612.