Vahid Malekian1, Nadine N. Graedel1, Oliver Josephs1, and Martina F. Callaghan1
1Wellcome Centre for Human Neuroimaging, UCL Queen Square Institute of Neurology, University College London, London, United Kingdom
Synopsis
3DEPI is
widely used for cortical depth-dependent fMRI studies at 7T. However, it
suffers from poor image contrast that can be problematic for fMRI post-processing
steps. One solution to bolster contrast is to implement a magnetization
transfer (MT) module. Here, we experimentally investigated the different MT
pre-pulse characteristics to optimize image contrast, while considering power
limitations at 7T. Contrast gains eased co-registration to an MP2RAGE
reference, and improved cortical segmentation. Our analyses confirmed that the
MT contrast is spatially variable and dependent on transmit field inhomogeneity.
Introduction
3D GE-EPI is
widely used for high-resolution fMRI at 7T. Relative to 2D-EPI it offers higher
SNR and avoids slice profile effects1,2, but exhibits greatly reduced image contrast which
can be problematic for accurate
registration3 and cortical segmentation, particularly in the context
of partial coverage. EPI also suffers from
spatial distortions due to low bandwidth in the phase-encoding (PE) direction,
which is exacerbated by larger echo-spacing as resolution is increased.
One solution
is to acquire a whole-brain EPI volume with enhanced image contrast and matched
spatial distortions. Recently, magnetization
transfer (MT) contrast has been proposed for laminar fMRI4. Within power constraints, often a limiting factor at
7T, we examined the effect of pulse shape, duration, off-resonance frequency
and flip-angle of the MT pre-pulse to optimize contrast. Methods
Pulse sequence: The MT
module consisted of two components: an off-resonance pulse to selectively
saturate the bound-pool and a spoiler gradient on the second PE direction to
spoil any inadvertent on-resonance excitation (Fig.1a).
Data
acquisition: A
healthy volunteer was scanned at 7T (Siemens-Terra) using an 8 transmit-channel
and 32 receive-channel head coil (Nova-Medical). The MT-3DEPI was acquired with:
TE/TR =16.2/100ms, FA=80, voxel-size=0.8mm-isotropic, GRAPPA=4, Partial-Fourier=0.75, in-plane segmentation=2,
coverage=192×192×128mm3. Acquisition time was 224s for two volumes
with reversed-PEs to facilitate distortion correction. An MP2RAGE was acquired
as an anatomical reference. Data were acquired with and without the MT
pre-pulse. To explore MT dynamics, two different pulse shapes (Gaussian and
Fermi) and durations (4ms and 8ms) were investigated. Power was explored by
varying FA (300o and 430o) and duration (4ms and 8ms) and
data were acquired with two off-resonance frequencies (2kHz and 4kHz). A total of six different acquisitions were
obtained (see Table1).
The
highest contrast acquisition was repeated in a second volunteer, along with a
no pre-pulse reference and a transmit field (B1+) map
based on the Bloch-Siegert effect5, to quantify contrast dependence on transmit field
efficiency.
Analysis: Pre-processing was performed using
FSL and SPM12. First, distortion correction was performed using topup6 followed by co-registration to the MP2RAGE. Tissue
masks obtained by segmenting the MP2RAGE image were used to define a brain
mask. Voxels with extreme signal dropout
were excluded from the analysis. Tissue contrast was computed using Eq.1 to 3,
together with power for each experimental condition7. Eq.4 was used to compute the contrast improvement
(effectively the MT-ratio, MTR) from the second experiment and plotted against
B1+ efficiency.
ContrastGM-WM=2*(SGM-SWM)/(SGM+SWM) [1]
ContrastGM-CSF=2*(SGM-SCSF)/(SGM+SCSF) [2]
ContrastGM-WM=2*(SWM-SCSF)/(SWM+SCSF) [3]
MTR=(Sno-MT-SMT)/Sno-MT [4]
Modelling: For each of the experiments, simulations
were performed using the EPG-X framework8 to quantify the percent signal decrease expected in white-matter
(WM) due to MT effects. Tissue parameters were as described in Corbin&Callaghan9, but neglected diffusion effects.Results
Example
images without (exp#1, Table1) and with (exp#9, Table1) an MT-prepulse are
shown in Fig.1a together with signal profiles along a section passing through
CSF, GM and WM (Fig.2b.) As expected, based on relative macromolecular content,
the signal intensity change is substantial in WM but minimal in CSF.
Tissue
contrast and relative power are reported in Table1. Contrast generally
increased linearly with pulse power with the exceptions that a) the Fermi pulse
in exp#3 produced lesser contrast despite higher power and b) for matched power
contrast was higher with lower off-resonance frequency (exp#8 vs. exp#9).
These findings
were replicated by simulations (Table1, final column). Of particular note, exp#3 was predicted to
have greater contrast per unit power in WM than exp#5 and the contrast was
predicted to be largest for exp#9.
To assess
the effect of contrast on GM segmentation, the MT (exp#9) and no-MT 3DEPI images
were segmented and overlaid on the MP2RAGE image (Fig.3). Segmentation was
improved by MT-induced contrast (yellow arrows in three exemplar regions).
However, some segmentation failure remained (blue arrows).
Fig.4 confirms
that the contrast improvement (MTR) achieved varies spatially with the transmit
field efficiency.Discussion
In this
study, the contrast improvement from employing MT-weighting with a standard
pre-pulse module was investigated empirically and via simulation. Our parameter
selections were motivated by the desire to maximise contrast while keeping
within power (SAR) constraints. While power is a key determinant of contrast
(Fig.4), it can also be affected by the dynamics of the magnetisation transfer
between the bound and free pool. The good
correspondence between simulation and experiment suggests that efficient
optimisation of parameter space could be achieved numerically.
Of the
conditions tested empirically, the greatest contrast was achieved with a 4ms
long Gaussian pulse with FA=300o and off-resonance frequency=2kHz. Although a lower off-resonance frequency
bolstered contrast, this risks inadvertent on-resonance excitation if the
spoiler gradient is insufficient. However, the Gaussian pulse had a
comparatively sharp spectral response (Fig.1b) and no artefacts were visible in
the images.
Our analysis
confirmed that the contrast imparted by the MT pre-pulse is spatially variable
and largely driven by transmit field inhomogeneity. This may be particularly problematic at 7T due
to increased field inhomogeneity. A
future direction would therefore be to develop more B1+-robust
solutions that also meet SAR constraints.
MT-weighting improved
contrast, segmentation and eased co-registration of EPI data to the more
anatomically faithful MP2RAGE reference, all of which are key steps in cortical
depth-resolved fMRI applications.Acknowledgements
Acknowledgements:
The Wellcome Centre for Human Neuroimaging is
supported by core funding from the Wellcome [203147/Z/16/Z].References
1. Huber, L. et al.
Techniques for blood volume fMRI with VASO: From low-resolution mapping towards
sub-millimeter layer-dependent applications. Neuroimage 164, 131-143
(2018).
2. Poser, B. A., Koopmans, P. J., Witzel, T., Wald, L. L.
& Barth, M. Three dimensional echo-planar imaging at 7 Tesla. Neuroimage 51, 261-266, doi:10.1016/j.neuroimage.2010.01.108 (2010).
3. Greve, D. N. & Fischl, B. Accurate and robust brain
image alignment using boundary-based registration. Neuroimage 48, 63-72,
doi:10.1016/j.neuroimage.2009.06.060 (2009).
4. Chai, Y. et al.
Magnetization transfer weighted EPI facilitates cortical depth determination in
native fMRI space. Neuroimage 242, 118455,
doi:10.1016/j.neuroimage.2021.118455 (2021).
5. Corbin, N.,
Acosta‐Cabronero, J., Malik, S. J. & Callaghan, M. F. Robust 3D
Bloch‐Siegert based mapping using multi‐echo general linear modeling. Magn Reson Med 82, 2003-2015 (2019).
6. Andersson, J. L. R., Skare, S. & Ashburner, J. How to
correct susceptibility distortions in spin-echo echo-planar images: application
to diffusion tensor imaging. Neuroimage
20, 870-888,
doi:10.1016/S1053-8119(03)00336-7 (2003).
7. Bernstein, M. A., King, K. F. & Zhou, X. J. Handbook of MRI pulse sequences. (Elsevier, 2004).
8. Malik, S. J., Teixeira, R. P. A. & Hajnal, J. V.
Extended phase graph formalism for systems with magnetization transfer and
exchange. Magn Reson Med 80, 767-779 (2018).
9. Corbin, N. & Callaghan, M. F. Imperfect spoiling in
variable flip angle T1 mapping at 7T: Quantifying and minimizing impact. Magn Reson Med 86, 693-708 (2021).