Lucia Navarro de Lara1,2, Isil Uluc1,3, Qinglei Meng1,3, Jason Stockmann1,3, Larry Wald1,3, and Aapo Nummenmaa1,3
1Martinos Center - MGH, Charlestown, MA, United States, 2Harvard Medical School, Boton, MA, United States, 3Harvard Medical School, Boston, MA, United States
Synopsis
For TMS/fMRI experiments, to acquire functional
images without artifacts, general recommendations have been proposed. In order
to apply TMS pulses in a multichannel Transcranial Magnetic Stimulation
system integrated with the new constructed TMSMR 28- channel RF coil, we have
explored the different options and artifacts that can be generated depending on
when the TMS pulses are applied with respect to the sequence timing. As
conclusion, we propose a method using an RF pick up loop to assure that the
delivery of the TMS pulses are not close to the EPI navigators or the imaging
gradients.
Introduction
Multichannel Transcranial
Magnetic Stimulation (mTMS) is an emerging technology for non-invasive
stimulation of the human brain. The use of multiple TMS coils in an array
configuration enables shifting the location TMS ‘hot spot’ electronically
without any mechanical movement. This is achieved by computationally determining
the current amplitudes to be passed to each of the coil elements to synthesize
a desired target field pattern (1). Combining non-invasive stimulation techniques such as TMS with fMRI
offer the unique benefits of studying the causal relationships and functional
connectivity between the nodes of large-scale brain networks. In particular, the mTMS technique would be become powerful when used in conjunction
with functional MRI (fMRI), since maneuvering any TMS coil inside the scanner
environment either manually or robotically is rather cumbersome.We have
constructed a whole head 28-channel RF coil array to be integrated
with the first 3-axis TMS (2) multichannel system (see
Figure1).
The basic combination of single-channel TMS and fMRI has been demonstrated (3) and proven to be safe (4). Artifacts produced
during imaging were carefully studied somewhere else (5,6) and general
recommendations were presented. To quantitatively understand the origin of the
artifacts, we propose a simple method to measure the relative timing of the MRI
and TMS pulses. We conducted an explorative experiment to analyze the effects
of the TMS timing with respect to the imaging sequence to be applied for our
novel multichannel 3-axis TMS system integrated with a whole head RF coil
array.Methods
Using our recently developed whole head TMSMR
28-channel RF coil array (see Figure1B), we applied TMS pulses during the
acquisition of functional images using simultaneous multi-slice EPI (7) of an spherical phantom (32 slices, grappa
2,TE=35ms, TR 1130ms, 2.5mm in-plane resolution, SL=2mm, MA88x88, FA90°) using
a z-coil element of the 3-axis system that was potted in epoxy (2)(see Figure2B).
The z-coil element was placed in 2 different positions in our TMSMR 28-channel
RF coil (see Figure2A). The two different positions were selected to
investigate the effects of torque and vibrations on the TMS coil (positions 1
and 2 present low and high torque conditions, respectively). The TMS pulses were
applied in 5 blocks (1 pulse per volume in a pre-defined slice acquisition
window) following a 10-volume period with no TMS pulses. The intensity of the
stimulator was chosen to be 15%. An additional z-coil element was placed close
to RF coil in the center of the bore to act as a magnetic field pick up loop to
acquire detailed sequence timing data.
The TMS pulse
triggering was synchronized with the slice timing triggers produced by the
scanner and was implemented using Presentation software (Neurobehavioural
Systems, Berkeley, CA,US). The TMS pulse were applied with various delays (0, 20ms,
30ms, 35ms, 40ms and 60ms) for both positions.
To record the
trigger signals and the signal produced in the z-coil used as a pickup loop, we
used a digital oscilloscope (RTB2004, Rhode&Schwarz).Results
Figure3 shows the oscilloscope
traces of the captured signals. The green channel is the TTL pulse produced by
the sequence for every slice acquisition, The acquisition below in yellow shows
the signal captured by our external z-probe (not the one placed inside the RF
coil for delivering magnetic pulses). This provides specific information of the
MRI pulse sequence any given time.
The TMS pulses were given
in the 6 conditions defined above, which are represented in Figure3 with a
lighting symbol on the exact time when they occurred. Pulses given with the
delays of 0ms, 20ms, and 30ms did not have any visible consequence on the
image quality. However, different artifacts were observed when the TMS pulses
were delivered exactly at 35ms (at the same time as the RF excitation pulse),
or at 40ms (affecting the EPI navigators) and at 60ms (affecting the imaging
gradients).
Figure4 and Figure5 presents
the observed artifacts for each of the studied positions (Position 1 and 2,
respectively). The artifacts produced when the delay was 40ms or 60ms are very
similar, cutting out a part of the phantom. The artifact that occurs when the
delay is exactly 35ms is similar for both positions and shows ringing effects
on the phantom. The artifacts affect always the first and the 17th
slice due to the multiband sequence (shown in red in the figures). Discussion
The results obtained in
this study show that general recommendations to avoid artifacts on TMS-fMRI
experiments should be avoided unless a relatively long ‘pause’ between EPI
volumes can be introduced (6). A pick-up loop or probe should be used to measure
of the sequence timing to determine appropriate times to deliver the TMS pluses
in continuous fast fMRI acquisition scenarios. Even though the TMS pulses are
very short (~300micros), they have a strong effect especially when placed closed
to the EPI navigators and Imaging gradients. We did not observe any influence
of the higher torque on Position 2, indicating that the explanation of
artifacts being produced by eddy
currents induced to the coil due to vibration was not supported by the data.
However, further investigations with various stimulator current intensities should
be done to elucidate the various TMS-induced artifacts and their origins.Acknowledgements
This work was funded by NIH
R00EB015445, R01MH111829, NIH R00EB021349 and the Rappaport Foundation. References
(1) Ruohonen J and Ilmoniemi R. Medical and Biological Engineering and
Computing, Vol. 36 p297-301,1998
(2) Navarro de Lara et al., NI, 224 117355, 2021
(3) Bohning et al.,Invest Radiol,33(6):336-340,1998
(4) Navarro de Lara et al., MRM, 84(2), p:1061-1075, 2020
(5) Bestmann et al. JMRI, 17:309-316 (2003)
(6) Navarro de Lara et al., MRM 74:1492-1501 (2015)
(7) Setsompot et al., MRM 67 (5) 1210-1224,
2012