Jaeyong Yu1,2, Sugil Kim3, and Jang-Yeon Park1,2
1Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Korea, Republic of, 2Department of Intelligent Precision Healthcare Convergence, Sungkyunkwan University, Suwon, Korea, Republic of, 3Siemens Healthineers Ltd., Seoul, Korea, Republic of
Synopsis
To reduce acquisition time, acceleration techniques
such as simultaneous multi-slice and parallel imaging techniques have been
applied to various sequences. In this work, we proposed a reconstruction method
for simultaneous multi-slab spatiotemporal-encoding (SMS SPEN) using split
slice-GRAPPA combined with CAIPIRINHA. The proposed method was applied to the
recently developed ultrafast 3D SPEN imaging sequence called ERASE (Equal-TE
Rapid Acquisition with Sequential Excitation) and a case of multiband factor of
3 was demonstrated with human
brain imaging at 3T.
Introduction
To reduce
the scan time without compromising spatial resolution, advanced acceleration
techniques such as parallel imaging1 and simultaneous multi-slice2
imaging have been developed in combination with numerous conventional imaging sequences,
including ultrafast echo-planar imaging (EPI).3 Recently, they have
also been applied to novel ultrafast imaging techniques using spatiotemporal
encoding (SPEN), which is non-Fourier encoding using frequency-swept pulses.
For example, Schmidt et al.4 proposed parallel SPEN imaging that
combines multiband frequency-swept pulses with a super-resolved SPEN
reconstruction algorithm and SENSE technique5. In this study, we
propose a reconstruction method for simultaneous multi-slab (SMS) spatiotemporal
encoding (SMS SPEN) using split slice-GRAPPA6 combined with
CAIPIRINHA7 imaging. Its performance was demonstrated in the case of
a multiband factor of 3 by human brain imaging at 3T.Method
Pulse sequence: The proposed method was applied to the
recently developed 3D ultrafast gradient echo-based SPEN imaging sequence
called ERASE (Equal-TE Rapid Acquisition with Sequential Excitation), which
provides constant TE in the SPEN direction with high tolerance B0 inhomogeneity
(Fig.1a).8 A linearly frequency-swept WURST-20 pulse9
was used for slab selection because of more desirable excitation profile than a
chirp pulse. The superposition of each single-band frequency-swept excitation
was implemented for multiband excitation (Fig.1b). The slab-selection
gradient was calculated based on the full-width-at-half-maximum (FWHM) of the
excitation profile obtained from Bloch simulation using a triple-band WURST-20
pulse (Fig.1c). The CAIPIRINHA method (CAIPI) was employed to reduce
aliasing especially when a multiband factor greater than two is used. To implement
CAIPI, an additional phase of π was added to the WURST-20 pulse for every even k-space line in the
phase-encoding direction.
Image reconstruction: To demonstrate the performance of the
proposed method, the multiband acceleration factor (or MB factor) was set to 3
(Fig.1b&c). Image reconstruction was performed by applying split
slice-GRAPPA (SP-SG) for SMS SPEN reconstruction to unfold the superimposed
data. To prevent each
band's quadratic phase profile from interfering each other during the super-resolved
(SR) reconstruction, SP-SG reconstruction was performed earlier than SR reconstruction. We used a SR reconstruction algorithm
proposed by Chen et al10. Each step of the reconstruction
process is shown in Fig.2 with each resulting image.
Experiments: All experiments were performed at 3T (Prisma,
Siemens) using a 64-channel head/neck coil. Scan parameters for ERASE imaging
were: FOV = 216×216 mm2,
matrix size = 108×108, slab
thickness = 32 mm, TR = 60 ms, TE = 29 ms, pulse time-bandwidth product (= Te×BW) = 146,
pulse length (Te) = 24.68 ms, and MB factor = 3.Results
Figure
2 shows an example of sagittal slices
containing the SPEN direction of the SMS 3D ERASE images at each step of the
reconstruction process. The sagittal
image reconstructed via SP-SG combined with CAIPI (a) showed better image
quality than the one obtained without CAIPI, demonstrating successful SMS
reconstruction in the SPEN (or slab-selective) direction.
Figure
3 shows three axial (in-plane) slices of 3D ERASE
images reconstructed in different conditions: (a) Original 3D ERASE
images acquired without SMS
excitation for reference, (b) SMS 3D ERASE images acquired
without CAIPI and reconstructed using SP-SG, (c) SMS 3D ERASE images acquired with CAIPI and
reconstructed using SP-SG.
When only SP-SG was used for SMS SPEN reconstruction without CAIPI, some
artifacts appeared in these axial images due to incomplete SMS reconstruction
in the SPEN direction (Fig.3b). When combined with CAIPI in data
acquisition, however, these artifacts were completely gone (Fig.3c). Discussion and Conclusion
In
this study, we proposed a reconstruction method for simultaneous multi-slab spatiotemporal
encoding (SMS SPEN) using split slice-GRAPPA (SP-SG) combined with CAIPIRINHA
(CAIPI). SMS 3D ERASE images were well reconstructed by applying SP-SG combined
with CAIPI-implemented data acquisition, showing good image quality without image
artifacts due to incomplete reconstruction of SMS SPEN. Since only MB factor 3 was
used here, the proposed method needs to be further tested in combination with in-plane
acceleration for higher acceleration. Comparisons between the proposed GRAPPA-based
method and other SENSE-based methods4 also need to be investigated in
future studies.Acknowledgements
This work was supported by NRF-2020R1A2B5B02002676
and NRF-2019M3C7A1031993.References
[1] Griswold MA, Jakob PM, Heidemann
RM, et al. Generalized autocalibrating partially parallel acquisitions
(GRAPPA). Magn Reson Med. 2002;47:1202-1210.
[2] Feinberg DA, Setsompop K.
Ultra-fast MRI of the human brain with simultaneous multi-slice imaging. J
Magn Reson. 2013;229:90-100.
[3] Ogawa S, Lee TM, Kay AR, Tank
DW. Brain magnetic resonance imaging with contrast dependent on blood
oxygenation. Proc Natl Acad Sci USA. 1990;87:9868-9872.
[4]
Schmidt R, Baishya B, Ben-Eliezer N, Seginer A, Frydman L. Super-resolved
parallel MRI by spatiotemporal encoding. Magn Reson Imag. 2014;32:60–70.
[5] Pruessmann KP, Weiger M, Scheidegger B,
Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med.
1999;42:952–962.
[6]
Cauley SF, Polimeni JR, Bhat H, Wald LL, Setsompop K. Interslice Leakage
Artifact Reduction Technique for Simultaneous Multislice Acquisitions. Magn
Reson Med. 2014;72:93-102.
[7] Breuer FA, Blaimer M, Heidemann RM, Mueller MF, Griswold MA, Jakob PM.
Controlled aliasing in parallel imaging results in higher acceleration
(CAIPIRINHA) for multiāslice imaging. Magn Reson Med. 2005;53(3):684-691.
[8]
Ryu JK, Han SH, Oh SH, Lee JS, Kim SG, Park JY. An equal-TE ultrafast 3D
gradient-echo imaging method with high tolerance to magnetic susceptibility
artifacts: Application to BOLD functional MRI. Magn Reson Med. 2021;85(4):1986–2000.
[9]
O’Dell LA. The WURST kind of pulses in solid-state NMR. Solid State Nuclear
Magn Reson. 2013;55-56:28-41.
[10]
Chen Y, Li J, Qu XB, Chen L, Cai CB, Cai SH, Zhong JH, Chen Z. Partial Fourier
transform reconstruction for single-shot MRI with linear frequency-swept
excitation. Magn Reson Med. 2013;69:1326-1336.