Ronald J Beyers1 and Thomas Denney1
1MRI Research Center, Auburn University, Auburn University, AL, United States
Synopsis
Alzheimer’s
disease (AD) and other cognitive debilitating disease drive the need to improve
neuro 7T MRI methods. We developed a multi-frequency
magnetization transfer (MFMT) method for contrast improvement in 7T 3D MRI of
the human hippocampus. MFMT applied to healthy volunteers demonstrated a 2.05
(p < 0.003) factor improvement in hippocampal contrast.
Introduction
An estimated
5 million Americans have suffered from Alzheimer’s disease (AD) since 2010 and are
expected to triple by 2050 [1]. AD and Mild
Cognitive Impairment (MCI) have been associated with hippocampal atrophy when
compared to normal controls. Neurofibrillary tangle deposition and neuronal loss
have also been associated with hippocampal atrophy [2]. Ultra-high field 7T MRI is transitioning into
clinical neurological use and may become the ideal method to detect and
quantify hippocampal atrophy with its 3D high-resolution capability. Simple
translation of lower field 3T methods to 7T are insufficient due to higher inhomogeneity
effects, SAR restrictions, and need for higher contrast. Specifically for 7T contrast enhancement, we
developed an improved Multi-Frequency Magnetization Transfer (MFMT) method to
couple with low-SAR 3D acquisition. MFMT was developed and applied to healthy
volunteers to quantify the contrast improvement with 7T hippocampal MRI.Methods
Previous literature reported improved results with standard
MT methods optimized at targeted brain locations [3, 4]. These previous methods applied mono-pulse MT
saturation at single offset frequencies and bandwidths that best-matched the
target ROI biomolecular environment. Our
MFMT expands on this to allow multiplexing up to four RF pulses with independent
offset frequencies and independent bandwidths – thereby simultaneously
saturating up to four biomolecular environments. This allows a highly tailored MT spectrum to
be combined with efficient 3D-FLASH image acquisition, as shown in Figure 1. The
saturation RF pulse are Gaussian, with RF phase stepping (spoiling) between pulses. As the MFMT induces the contrast in this sequence,
the 3D-FLASH allows maximum readout SNR.
Preliminary test runs applied MFMT offsets at ±11.8, ±6.7, ±4.0 and ±3.5
ppm to find an optimal combination for best contrast in the hippocampus.
Four healthy volunteers, with informed consent, were
scanned in a Siemens 7T scanner, using a Nova 1 Tx/32 Rx head coil. A coronal
3D slab was positioned to simultaneously image both hippocampi. The 3D slab/volume image with isometric
pixels allowed for easy post-scan reslicing to any cor/sag/tra direction. Scans
with MFMT 'on' and 'off' were run to allow comparison. On multiple acquired hippocampal slices, contrast-to-noise
(CNR) analysis was performed between the hippocampal sulcus (dark) versus the neighboring
CA4 region (bright).
Scan parameters: MFMT: Simultaneous saturation frequencies
= ±11.8 and ±6.7 ppm each with bandwidth = 0.84ppm, Gaussian pulses, flip-angle
= 120º, 30-pulse saturation train length. 3D-FLASH: Slab-selective FOV =
256x240x24 mm, matrix = 688x645x64, Pixel size = 0.4x0.4x0.4 mm, FA = 7.5º,
TE = 3.4 ms, TR = 850 ms, BW = 360 Hz/Pix, Avg = 1, Duration = 7 min, 30 sec.Results
Preliminary scans indicated the best hippocampal
anatomical contrast was obtained with simultaneous saturations at offsets
+11.8, +6.7, -11.8 and -6.7 ppm, all with bandwidth 0.84ppm. From this preliminary data, in vivo hippocampal 3D
MRI sets were acquired from all four volunteers. Figure 2 shows results from one volunteer at
four transverse hippocampal slice loacations.
The Figure 2 images present a clear increase in contrast for MFMT 'on'
versus 'off'. The red arrows indicate the
hippocampal regions and the increased contrast between the thin dark
hippocampal sulcus and surrounding tissue regions. The improved contrast facilitates the detection
and tracking of atrophy. CNR
quantification analysis for MFMT 'on' versus 'off' gave a mean CNR increase of 4.43±1.63
(mean±stdev). This 4.43 mean difference is equivalent to a CNR improvement
ratio of 2.05 (p < 0.003) -- therefore MFMT effectively “doubled” the
contrast.Discussion
The effectively doubled contrast
provided discernible anatomical contrast at 0.37 mm resolution for improved investigation
of hippocampal condition. Similar to other 7T MRI, poor B0 shimming reduced the
MFMT contrast effectiveness. Typical SAR
levels were low and easily managed by lowering MFMT RF pulse(s) flip angle or increasing
sequence repetition time (TR).Conclusions
The applied MFMT
demonstrates an effective contrast enhancement method for 7T hippocampal MRI. Continued study will investigate additional
offset frequency and bandwidth combinations that further improve hippocampal
and other brain region contrast.Acknowledgements
Special thank you for project support goes to Julie
Rodiek, Steven Nichols and Adam Davila.References
1. Goukasian N, Porat S, et al “Cognitive Correlates of Hippocampal Atrophy…”, Dement Geriatr Cogn Disord Extra
2019;9:281–293
2. Zarow C, Wang L, et al "MRI Shows More Severe Hippocampal Atrophy…", Inter
J of Alzheimer’s Disease, vol. 2011, Article ID 483972, 2011
3. Mougina O,
Clemence M, et al “High-resolution
imaging of magnetisation transfer…”, NMR Biomed. 2013;26:1508–1517
4. Priovoulos N, Jacobs H, et al “High-resolution in vivo
imaging of human locus coeruleus…”, NeuroImage 2018;168:427–436