Alicia Palomar Garcia1, Valentin H. Prevost2, Alba Iruela Sanchez1, Wolter de Graaf3, and Bruno Triaire2
1Canon Medical Systems Spain and Portugal, Barcelona, Spain, 2Canon Medical Systems Corporation, Tochigi, Japan, 3Canon Medical Systems Europe, Zoetermeer, Netherlands
Synopsis
This study evaluated the feasibility
of performing virtual elastography based on diffusion-weighted MRI (vMRE) on a
1.5T system with two different acceleration methods, SPEEDER and EXSPER. Proton
density fat fraction (PDFF) imaging was also acquired. The vMRE and PDFF
measurements were consistent with healthy values that can be found in the
literature. This study demonstrated the feasibility of performing vMRE at 1.5T
for the estimation of overall liver elasticity. SPEEDER and EXSPER vMRE
implementations provided consistent stiffness estimations and both would be adequate
for conducting future studies at 1.5T systems on patients.
Introduction
Chronic
liver diseases are one of the leading causes of mortality globally. For these
patients, an early diagnosis and correct staging is crucial for improving
prognosis. Although liver biopsy is the reference technique for disease staging,
many non-invasive alternatives have been proposed to provide stiffness estimation
while avoiding the possible risk of complications such as hemorrhage or
infection. Some of these methods, such as MR elastography (MRE) and ultrasound
techniques like 2D shear wave elastography (SWE), provide good diagnostic
accuracy, but present some limitations in clinical context because they require
dedicated hardware and provide local estimates of liver stiffness,
respectively. More recently, virtual elastography based on diffusion-weighted
MRI (vMRE) has been introduced as an alternative to quantify tissue stiffness
in the whole liver in a single acquisition without the need of additional
equipment1. In all previous studies using vMRE, the MRI was
performed on 3T systems1,2,3. This study evaluated the feasibility
of performing vMRE on a 1.5T system, comparing two different parallel imaging implementations
(SPEEDER and EXSPER). In addition, the relation between vMRE measurements and
proton density fat fraction (PDFF) was evaluated. Methods
Five
healthy volunteers underwent whole-liver MRI sessions on a Vantage 1.5T Orian
XGO system (Canon Medical Systems Corporation, Tochigi, Japan) with a
16-channel body coil combined with a spine coil. Scanning sessions included
diffusion-weighted imaging (DWI) with the following parameters: 2D SE-EPI; PASTA
fat suppression; TR=4614ms; TE=67ms; in-plane resolution=2x2mm; slice
thickness=6mm, parallel imaging=SPEEDER or EXSPER, acceleration factor=3; two
b-values with different number of averages: 200 s/mm2 (NAQ=2) and
1500 s/mm2 (NAQ=4); respiratory triggering and total scanning time
of 5:15 min or 6 min for SPEEDER and EXSPER, respectively. A multi-echo
breath-hold 3D FE sequence for fat fraction quantification was also acquired with
the same resolution and slice thickness.
A
dedicated Olea Sphere plugin was used to compute the virtual elasticity maps
from the DWI data, based on the linear relationship between shifted ADC and
stiffness modulus stated in previous studies1,2,3. For
quantification analysis, five measurements were done in each subject, placing
ROIs in the same position by coregistering PDFF and elasticity maps. Spearman’s
rank correlation was used to evaluate the relation between vMRE-based stiffness
and PDFF. Measurements for a volunteer with high PDFF, but no reported liver
disease, were discarded, so a total of 20 measures were considered for the
correlation. Results
Resulting
virtual elasticity maps (with SPEEDER and EXSPER) and PDFF map are shown in
Figure 1. Mean values of PDFF and vMRE with both acceleration techniques
(SPEEDER and EXSPER) for the five healthy volunteers are reported in Figure 2. Stiffness
measurements based on vMRE with SPEEDER were in the range between 1.11 ± 0.35
kPa and 3.31 ± 0.51 kPa, and measurements using EXSPER were within 1.46 ± 0.53
kPa and 2.75 ± 0.54 kPa. No significant correlation was found between vMRE
measurements and PDFF. Discussion
The
range of values obtained with vMRE is aligned with previous studies reporting
normative stiffness in healthy subjects1,2. In addition, there is no
significant difference between the two different acceleration methods that were
tested. Both SPEEDER and EXSPER implements would be could be then used for vMRE
imaging. However, in specific scenarios such as protocols with small field of
view, EXSPER would be more robust to unfolding artifacts4.
The
PDFF percentage obtained for the sample studied is consistent with the values
reported for healthy subjects5,6. In this study, no significant
correlation was found between fat fraction and virtual elasticity, possibly due
to the limited sample size and the short variability between measurements. Further
studies with larger samples and including patients with liver diseases would be
necessary to proper study the relation between these metrics.Conclusion
This
study demonstrated the feasibility of performing vMRE at 1.5T for the
estimation of overall liver elasticity. Data acquired with SPEEDER and EXSPER
provided consistent stiffness measurements on healthy livers and both would be adequate
for vMRE studies. Future works with larger samples and the inclusion of
patients with liver disease would be needed to better estimate the relation
between liver elasticity and fat fraction. Acknowledgements
No acknowledgement found.References
- Le Bihan D,
Ichikawa S, & Motosugi U. Diffusion and intravoxel incoherent motion MR imaging–based
virtual elastography: a hypothesis-generating study in the
liver. Radiology. 2017; 285(2), 609-619.
- Kromrey M L, Le
Bihan D, Ichikawa S, et al. Diffusion-weighted MRI-based virtual elastography
for the assessment of liver fibrosis. Radiology. 2020; 295(1),
127-135.
- Ota T, Hori M, Le Bihan D, et al. Diffusion-Based Virtual MR Elastography of the
Liver: Can It Be Extended beyond Liver Fibrosis? J Clin Med. 2021; Sep 30; 10(19):4553.
- Deshmane A, Gulani V, Griswold M A, et al.
Parallel MR imaging. Journal of Magnetic Resonance Imaging. 2012; 36(1),
55-72.
- Idilman I S,
Aniktar H, Idilman R, et al. Hepatic steatosis: quantification by proton
density fat fraction with MR imaging versus liver biopsy. Radiology.
2013; 267(3), 767-775.
- Caussy C, Reeder
S B, Sirlin C B, et al. Noninvasive, quantitative assessment of liver fat by
MRI‐PDFF as an endpoint in NASH trials. Hepatology. 2018; 68(2),
763-772.