2245

Initial Study on Imaging Cyclic Tissue Motion with Motion Encoding Gradients in the Shape of Wavelet Basic Functions
Yuan Le1, Jun Chen1, Phillip J. Rossman1, Bradley D. Bolster Jr.2, Stephan Kannengiesser3, and Richard L. Ehman1
1Radiology, Mayo Clinic, Rochester, MN, United States, 2MR Collaborations, Siemens Medical Solutions USA, Inc., Salt Lake City, UT, United States, 3MR Application Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany

Synopsis

The goal of this study was to better detect cyclic motion with multiple frequencies in tissue. A novel MRE technique was developed using motion encoding gradient (MEG) pulses of multiple scales (multi-scale MRE). These MEG pulses were designed and timed to approximate a complete basis of Haar wavelet system. Our hypothesis was that a set of orthogonal MEG pulses can capture the tissue motion more efficiently and tissue motion can be reconstructed by a simple inverse transform. Initial tests were performed on a phantom and a volunteer.

Introduction

MR Elastography (MRE) uses specialized motion encoding gradients (MEG) to encode tissue motion into the phase images1, 2. Recently multi-frequency and transient MRE have been under intense study3-8. For these applications the MEG needs to be carefully selected, and complicate deconvolution is needed for motion reconstruction3, 9-11. The goal of this study was to image this broadband tissue motion with MEGs of multiple scales (multi-scale MRE) in the shape of a Haar wavelet series12, which more efficiently encode broadband motion, and simplify the motion reconstruction into an inverse transform.

Theory

In MRE, the accumulated phase due to motion for each voxel is (eq.1):
$$\theta(t)=\frac{\gamma}{2\pi}\int_{0}^{t}\overrightarrow{G(\tau)}\cdot\overrightarrow{r(\tau)}d\tau \ \ \ [1] $$

G(t) is the MEG and r(t) is the displacement. According to this equation, the phase is the cross-correlation between the tissue motion and the MEG. Therefore, if MEG is designed to approximate to a Haar series12 (eq.2, Figure 1):
$$Scaling\ function:\varphi\left(t\right)=\begin{cases}1 & 0≤t<1\\0 & otherwise\end{cases} $$$$mother wavelet function: \psi(t)=\begin{cases}1 & 0\leq t<1/2\\-1 & 1/2\leq t<1\\0 & otherwise\end{cases} \ \ [2]$$ $$wavelet function: \psi_{j,k}(t)=\begin{cases}\sqrt{2}^{j} & \frac{k-1}{2^{j}}\leq t<\frac{k-1/2}{2^{j}}\\ -\sqrt{2}^{j} & \frac{k-1/2}{2^{j}}\leq t<\frac{k}{2^{j}}\\0 & otherwise\end{cases} $$
The accumulated phase would approximate the Haar transform of the motion (eq. 3):
$$a_0=\frac{1}{2^n}\int_{0}^{1}f\left(t\right)\varphi\left(t\right)dt$$
$$a_1=\frac{1}{2^n}\int_{0}^{1}f\left(t\right)\psi\left(t\right)dt \ \ \ \ [3]$$
$$a_{j,k}=\frac{1}{2^n}\int_{0}^{1}{f\left(t\right)\psi_{j,k}\left(t\right)dt}$$

Then displacement can be decoded using the inverse Haar transform (eq 4):
$$f_n\left(t\right)=a_0\varphi\left(t\right)+a_1\psi\left(t\right)+\sum_{2≤j≤n,\ 0≤k<2j-1}a_{j,k}ψ_{j,k}(t) \ \ \ \ [4]$$

Methods

Figure 1 shows the diagram of a spin-echo-EPI-based MRE with multi-scale trapezoidal ‘Haar wavelet’ MEGs. In this figure two scales of wavelets and scaling function for two consecutive time windows is displayed (7 offsets). A single scaling function would crush the signal in MRI so a bipolar gradient that covered two consecutive time windows was used instead to encode the difference in a0 between the two windows.
Firstly, a simulation verification was performed. A simulated true motion was defined to be a frequency sweep (20 Hz-80 Hz in 245.75 ms). This 245.75 ms interval was divided into 6 windows. Four wavelet scales plus scaling function were used to approximate the ‘true’ motion. Trapezoidal approximation of the wavelet series (slope length = 1ms) was simulated in addition to the rectangular theoretical Haar wavelet series. This was compared to a simulation performed with standard trapezoidal MEG of 60 Hz with a trigger delay of 4ms.
Secondly, a PVC phantom was scanned using both standard and a prototype multi-scale MRE sequence on a 3T scanner (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany). A pneumatic driver (Resoundant Inc., Rochester MN, USA) generated the wave motion: (1) A composite continuous wave with 30Hz and 60Hz component having of equal amplitude, and (2) one cycle of 90-Hz sinusoidal wave (11 ms pulse length) per TR per slice. In multi-scale MRE a total 36 ms of motion were sampled, with 2 scales of wavelet in addition to the scaling function. MEG amplitude was 5mT/m for scaling and 2nd scale wavelet, and 7mT/m for the 1st wavelet. TR/TE=3200ms/47ms. With standard MRE 8 phase offsets were acquired, motion frequency=30 Hz, MEG frequency = 120 Hz, MEG amplitude = 40mT/m, and TR/TE=3200ms/35ms.
Lastly, a volunteer test was performed with multi-scale MRE using 1 scale of wavelets. Motion frequency=30 Hz, MEG amplitude=35mT/m and other parameters were the same were the same as for the phantom test.

Results

Figure 2 shows that motion within the simulated range can be detected by both standard and multi-scale MEG. With 4 scales of the Haar wavelets the reconstructed results followed the ‘true’ motion very well; With the trapezoidal ‘Haar wavelet’, the sensitivity decreased, especially when the frequency is higher, while the waveform was still preserved. The standard MEG is more sensitive to the frequency around 60 Hz and the phase response varied with the frequency.
In the composite wave test, both MRE techniques detected both frequency components. The standard MRE images showed more of the higher frequency component. The detected signal amplitude ratio of 60 Hz vs. 30 Hz was 1.38 with standard MRE and 0.44 in multi-scale MRE in the Y direction (the direction of the applied motion). Given that the standard 60-Hz MEG are 3.27 times as sensitive to 60 Hz as to 30 Hz, the composite wave imaging results from the two techniques were consistent. Stiffness calculated from the 30 Hz wave was 2.4 kPa with multi-scale MRE and 2.3 kPa with standard MRE, very close to each other.
Figure 4 shows the transient wave field caused by a single 90 Hz pulse in phantom. It can be observed that multiple frequency components are present in the phantom due to this pulse, and that standard MRE seemed to be more sensitive to the higher frequency component.
In the volunteer test, the 30 Hz wave images were reconstructed very well from multi-scale MRE (Figure 5).

Conclusions

The multi-scale MRE technique detected broadband motion with uniform frequency response hence better preserve the motion waveform. Phantom test results from multi-scale MRE and standard MRE were consistent. Multi-scale MRE is potentially useful to image transient or multi-frequency motion or motion with unknown frequency.

Acknowledgements

This work was supported by grant from National Institutes of Health R01 EB001981.

References

1. Litwiller, D. V., et al. Curr Med Imaging Rev. 2012;8(1):46-55.

2. Muthupillai, R., et al. Science. 1995;269(5232):1854-7. Epub 1995/09/29.

3. McCracken, P. J., et al. Magn Reson Med. 2005;53(3):628-39. Epub 2005/02/22.

4. Souchon, R., et al. Magn Reson Med. 2008;60(4):871-81. Epub 2008/09/26.

5. Hofstetter, L. W., et al. Magn Reson Med. 2019;81(5):3153-67. Epub 2019/01/22.

6. Hofstetter, L. W., et al. Phys Med Biol. 2020. Epub 2020/12/23.

7. Smith, D. R., et al. J Biomech Eng-T Asme. 2020;142(7).

8. Shahryari, M., et al. Magnetic Resonance in Medicine. 2021;85(4):1962-73.

9. Guo, J., et al. Rofo. 2014;186(3):260-6. Epub 2013/09/04.

10. Dittmann, F., et al. Magn Reson Med. 2015.

11. Solanas, P. S., et al. Nmr in Biomedicine. 2021;34(2).

12. Haar, A. Mathematische Annalen. 1910;69(3):331-71.

Figures

Figure 1. Haar basic function system and the diagram of a spin echo EPI multiscale MRE sequence with 2 scales of wavelet.

Figure 2. Simulation of (a) acquired phase information from standard MEG and (b) the calculated displacement with multi-scale Haar wavelet MEG. A 20Hz-80Hz frequency sweep signal was used as ‘true motion’.

Figure 3. Wave images acquired using multiscale MRE vs. standard MRE with a motion composed of both 30 Hz and 60 Hz wave.

Figure 4. Wave images from multiscale MRE vs. standard MRE with transient motion induced with one cycle of 90 Hz sinusoidal wave per TR per slice.

Figure 5. Wave images from a volunteer using 30 Hz motion and multiscale MRE.

Proc. Intl. Soc. Mag. Reson. Med. 30 (2022)
2245
DOI: https://doi.org/10.58530/2022/2245