2195

Prospective motion correction in multi-echo line-scanning fMRI at 7T: sequence implementation and strategies for functional data analysis
Luisa Raimondo1, Nikos Priovoulos1, Jurjen Heij1, Tomas Knapen1,2, Serge O Dumoulin1, Jeroen JW Siero1,3, and Wietske van der Zwaag1
1Spinoza Centre for Neuroimaging, Amsterdam, Netherlands, 2VU University, Amsterdam, Netherlands, 3Radiology, University Medical Centre, Utrecht, Netherlands

Synopsis

We implemented prospective motion-correction (PMC) for line-scanning fMRI. In line-scanning, online motion correction is needed since 1-dimensional data do not allow motion detection in every dimension and the limited coverage makes this approach sensitive to spin-history artifacts or scanning outside the area of interest, in the presence of motion. We compared three versions of PMC and three ways of managing the T1-driven return to equilibrium (T1-transient) introduced by the navigators, in the functional data analysis. We opted for a water-excitation navigator as the best alternative, and showed that the T1-transient can be successfully overcome in the functional analysis.

Introduction

Gradient-echo line-scanning (GELINE) fMRI is a recent technique1,2 able to achieve extremely high spatio-temporal resolution in one direction (250 μm and 105 ms, in humans), at the cost of volume coverage. Subject movement leads to image artifacts but this is even more problematic at high resolution MRI, where smaller movements also have an impact. In the case of line-scanning, the 1-dimensional nature of the data only allows motion detection in the line direction. Moreover, the very limited coverage renders this approach sensitive to spin-history artifacts or scanning outside the area of interest if motion occurs during a scan. These effects cannot be corrected by posthoc motion correction; prospective motion correction is required. Therefore, we investigated navigator-based prospective motion correction (PMC) based on 3 for line-scanning fMRI acquisitions and analysis. The required time gap in the acquisition of the line-scanning data, while acquiring and coregistering every navigator, introduces signal instabilities in the timecourses, characterized by a T1-driven return to steady-state (T1-transient). We tested three image-based navigators and, for each, we evaluated the performance of three ways of managing the subsequent T1-transient in the line-scanning data acquisition during functional experiments.

Methods

We acquired data from 4 participants at 7T MRI system (Philips, Netherlands) equipped with a 2 channel transmit, and 32 channel receive surface coils4. Line-scanning data acquisition used a modified 5 echoes gradient-echo sequence1 where the phase-encoding in the direction perpendicular to the line was turned off: line resolution=250μm, TR=105ms, TE1=6ms, ∆TE=8ms, readout bandwidth=131.4 Hz/pixel, fa=16°, array size=720, line thickness=2.5mm, in-plane line width=4mm, fat suppression using SPIR. Two saturation pulses (7.76 ms pulse duration) suppressed the signal outside the line of interest. The line was positioned as perpendicular to the cortex as possible. For every scan we performed PMC, achieved by an interleaved scanning architecture (MISS, Philips) at every dynamic (i.e. every 440 timepoints = ~46s) (Figure 1). During the functional scan, the last acquired navigator volume was registered to the previous one in the series (wait=1s), and translation and rotation parameters of both the navigator and target sequence were updated in real time3. Three possible navigators were investigated (Figure 2): a highly-accelerated surface-coil-receive fat-navigator only covering the back of the head, a slower whole-head, transmit-coil-receive fat-navigator and a water-excitation navigator, aimed at reducing the amplitude of the T1-transient. We acquired one run of functional data (6min 20s) with each scan, using a block design visual task consisting of a 20 Hz flickering checkerboard, presented for 10s ON/OFF, starting and ending with 10s baseline. We investigated 3 ways of managing the presence of gaps and T1-transient:
1. Regressing out the T1-transient during the general linear model (GLM) analysis (regressed)
2. Interpolating the points corresponding to the T1-transient by substituting them with the average of the points before and after the T1-transient (interpolated)
3. Applying a T2* fit reconstruction to avoid T1 effects (T2* fit)
The timepoints of the visual task model corresponding to the time during which the navigators were acquired, were removed from the GLM. We evaluated the beta values and temporal signal to noise ratio (tSNR) along the line, to find the best acquisition and analysis strategy.

Results & Discussion

Figure 3 shows a single voxel timecourse for the three acquisitions (a, b and c), for raw data, regressed, interpolated and T2*fit. BOLD responses were visible in all timecourses, despite the T1-transient. Notice that the shorter, more undersampled, navigators (acq 1 and 3) resulted in lower T1-transient height. The utilization of water excitation for the navigators reduced the T1-transient height further. For all acquisitions, the T1-transient was much reduced after GLM-based signal regression and completely disappeared in T2* fit data. In Figure 4, the tSNR for the interpolated data was superior to the regressed and T2*fit data. This trend was consistent over participants (Figure 4d). The tSNR of the regressed data was reduced by remaining T1-related instabilities, which we were not able to remove with a single-exponential regressor. The T2* fit from our current data proved too noisy, which further reduced tSNR. Figure 5 shows the beta values for the three acquisitions (a, b and c) of a representative participant for raw data and for the three strategies of handling the presence of gap in the acquisitions and the T1-transient. Interpolated and regressed data show similar beta values to raw data. Notice that beta values were higher in the raw data due to the presence of high signal intensities caused by the T1-transient that corresponded with task ON blocks, while the estimation of betas is less, or not, biased when interpolated or regressed data are used. The T2*fit data appear to suffer from increased noise, and instabilities (as apparent from the lower tSNR) which reduces the fit results.

Conclusion

We implemented prospective motion correction for line-scanning in order to prevent motion-related signal changes during line-scanning acquisitions. Overall, among the three navigators tested, the navigators acquired with water excitation and surface coils (acq 1) presented the lowest amplitude of the T1-transient signal. The T1-transient introduced by the navigators can be best dealt with by interpolating the affected time points. This leads to higher temporal stability (tSNR) than fitting the T1-transient or T2* fit-based timecourses.

Acknowledgements

This study was supported by the Royal Netherlands Academy of Arts and Sciences Research Fund 2018 (KNAW BDO/3489), the NWO TTW VIDI grant (VI.Vidi.198.016) and the Visiting Professors Program 2017 (KNAW WF/RB/3781) granted to the Spinoza Centre for Neuroimaging.

References

1. Raimondo L, Knapen T, Oliveira ĺcaro AF, Yu X, Dumoulin SO, van der Zwaag W, et al. A line through the brain: implementation of human line-scanning at 7T for ultra-high spatiotemporal resolution fMRI. J Cereb Blood Flow Metab, 2021.

2. Morgan AT, Nothnagel N, Petro LucyS, Goense J, Muckli L. High-resolution line-scanning reveals distinct visual response properties across human cortical layers. bioRxiv, 2020.

3. Andersen M, Björkman-Burtscher IM, Marsman A, Petersen ET, Boer VO. Improvement in diagnostic quality of structural and angiographic MRI of the brain using motion correction with interleaved, volumetric navigators. PLOS ONE, 2019.

4. Petridou N, Italiaander M, van de Bank BL, Siero JCW, Luijten PR, Klomp DWJ. Pushing the limits of high-resolution functional MRI using a simple high-density multi-element coil design. NMR Biomed, 2013.

Figures

Figure1: Sequence diagram.

Figure2: Scan parameters of the navigators scans for the prospective motion correction.

Figure3: Single voxel timecourse (first 1500 timepoints) for an example subject, for the three acquisitions (a, b and c), for raw data (blue line, left column) and T1 regression (red line, right column), data interpolation in correspondence of the T1-transient (green line, right column) and T2* fitted data (yellow line, right column).

Figure4. tSNR for an example subject, for three acquisitions (a, b and c), for raw data (blue line) and T1 regression (red line), data interpolation in correspondence of the T1-transient (green line) and T2* fitted data (yellow line). (d) Average tSNR along the regions inside the brain, across subjects. The errors represent the standard error over subjects.

Figure5. Betas for an example subject, for the three acquisitions (a, b and c), for raw data (blue line) and T1 regression (red line), data interpolation in correspondence of the T1-transient (green line) and T2* fitted data (yellow line).

Proc. Intl. Soc. Mag. Reson. Med. 30 (2022)
2195
DOI: https://doi.org/10.58530/2022/2195