Matthijs H.S. de Buck1, James L. Kent1, Aaron T. Hess1, and Peter Jezzard1
1Wellcome Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom
Synopsis
DANTE-SPACE can be used for
visualizing intracranial vessel walls at 7T, where the flow-sensitive signal
attenuation of the DANTE preparation simultaneously suppresses signal from the internal
blood and external CSF. However, the efficacy of DANTE is reduced for blood
moving upstream through the neck due to limited B1+ coverage at 7T. Here,
EPG simulations are used to estimate the effect of this limited B1+ coverage
on the achieved contrasts. A pTx-based approach which uses separate RF shims
for the DANTE-preparation and the SPACE readout to improve the attenuation of
blood moving through the neck is proposed, and validated in vivo.
Introduction
Vessel wall imaging (VWI)
can be used to assess the pathology of intracranial vessel walls. For the
assessment of intravascular pathology in the vessels above and around the
Circle of Willis, such as the middle cerebral arteries (MCAs), the lumen signal
needs to be suppressed while retaining signal from the vessel wall. A previous study presented the DANTE-SPACE sequence at
7T1 (Figure 1), which achieves delineation
of the MCA walls by employing an optimized variable-flip-angle turbo spin-echo
sequence (SPACE2) with additional DANTE-preparation3 for flow-sensitive signal suppression
of both the CSF and the lumen. However, assessment of intravascular pathology
and calculation of vessel wall thickness using black-blood sequences such as
DANTE-SPACE can be hindered by remaining unsuppressed blood signal.
In this work, a previously
introduced framework for the simulation of DANTE-SPACE4 was used to identify and quantify the
cause of this unsuppressed blood signal. Based on the simulation results, a
parallel transmission (pTx)5 implementation of DANTE-SPACE is proposed
to reduce the effects of limited B1+ coverage in the neck at 7T, and
thereby obtain improved blood suppression. Methods
DANTE-SPACE was simulated using
the extended phase graph (EPG) framework6,7. For accurate representation of the
blood signal, the simulations included pulsatile flow velocity variations8, estimated flow trajectories, and B1+
variations in different parts of the vasculature based on in vivo B1+ maps. Constant B1+ signal was
assumed for the vessel walls.
To characterize the limitations
when using single-channel transmit coils, B1+ flip angle maps were
acquired using the 3D-DREAM sequence9 from four volunteers on a 1Tx/32Rx
system. To assess the potential B1+ improvements when using pTx, additional
channel-by-channel relative B1+ maps were acquired in 3 volunteers for
all 8 transmit channels using an 8Tx/32Rx coil. Those individual maps were
combined using circular polarization (CP)-mode, as well as using an optimized
radiofrequency (RF) shim aimed at maximizing the minimum B1+ values
in the neck.
DANTE-SPACE signal was
both simulated and acquired (from 3 healthy volunteers) using the acquisition
parameters used in Ref. (1), adopting
the optimized DANTE-preparation proposed in Ref. (4) for
improved contrast and reduced SAR. These parameters include 200 DANTE pulses of
9º, and a SPACE readout consisting of 75 pulses with an echo spacing of 4.6ms.
The TR for each readout was 2.64s. The scan time was 11 minutes for data acquired
at 0.5×0.5×1.0 mm3 using a GRAPPA acceleration factor
of 4. pTx DANTE-SPACE data were acquired using an 8Tx/32Rx coil on a Siemens
Magnetom 7T scanner (Erlangen, Germany) under an institutional ethical
agreement. Rigid-body image registration was used where necessary. Results
Figure 2a shows measured B1+ profiles at 7T from four
volunteers. Figure 2b shows the possible B1+ improvement in the neck
when using pTx with a neck-specific RF shim, based on masked voxels in the neck
of 3 volunteers.
Figure 3 shows the simulated blood signal and blood/vessel wall contrast
when running pTx DANTE-SPACE with both DANTE and SPACE in CP-mode (red circles),
and when running a neck-specific pTx RF shim for DANTE and CP-mode for SPACE
(blue circles). On average, the blood signal is expected to decrease by 26±15%
when a neck-shim is used during DANTE-preparation.
Figure 4 shows a single-slice segment acquired both with and without a neck-specific
RF shim for the DANTE-preparation. Figure 5 shows a comparison of the resulting
signal in the lumen, vessel wall, and CSF for multiple slices in all three
volunteers, with values relative to the vessel wall signal in CP. Across the
three subjects, the lumen signal is found to decrease (by 7±6% of the CP-mode lumen
signal), while the vessel wall and CSF signal both increase.Discussion
DANTE-SPACE simulations (Figure
3) show a reduced vessel wall/blood contrast as a result of the reduced B1+
coverage in the neck of a 1Tx system at 7T, which reduces the efficiency of the
DANTE-preparation on blood spins flowing into the MCAs from the neck. If the B1+
coverage is increased during DANTE-preparation by using a pTx shim tailored for
maximizing the minimum signal in the neck, the simulations show improved
suppression of the blood signal.
The reduction in blood signal using
a neck-specific shim is also visible in vivo (Figures 4-5), while the
signal from both the vessel walls and the CSF increases. Those signal increases
can be a result of reduced B1+ outside the neck when using the neck-specific
shim, resulting in reduced DANTE-suppression of the vessel walls and CSF. Although
this was not a target during this work, the increased vessel wall signal further
increases the vessel wall/lumen contrast, while also providing increased vessel
wall-CSF contrast despite the higher CSF signal. Future work will look into
increasing this effect using RF shims that aim to increase the B1+ in
the neck while simultaneously suppressing the B1+ around the vessels
of interest. Future work will also study the potential of using dielectric pads10 around the neck for additional
improvement of the B1+ in the neck. Conclusion
Suppression of blood signal in DANTE-SPACE at 7T is less effective due
to reduced B1+ coverage in the neck. Using a neck-specific pTx shim
during the DANTE-preparation reduces this effect and improves vessel wall
contrast.Acknowledgements
MdB receives financial support from Siemens
Healthineers and the Dunhill Medical Trust. JK is supported by an iCASE stipend
award in collaboration with Siemens Healthineers. PJ receives support from the
Dunhill Medical Trust and the NIHR Oxford Biomedical Research Centre. The
Wellcome Centre for Integrative Neuroimaging is supported by core funding from
the Wellcome Trust (203139/Z/16/Z).
References
1. Viessmann O, Li L, Benjamin P, Jezzard P. T2-Weighted
intracranial vessel wall imaging at 7 Tesla using a DANTE-prepared variable
flip angle turbo spin echo readout (DANTE-SPACE). Magnetic Resonance in
Medicine. 2017;77(2):655-663. doi:10.1002/mrm.26152
2. Mugler JP. Optimized three-dimensional fast-spin-echo MRI.
Journal of Magnetic Resonance Imaging. 2014;39(4):745-767.
doi:10.1002/jmri.24542
3. Li L, Miller KL, Jezzard P. DANTE-prepared pulse trains: A
novel approach to motion-sensitized and motion-suppressed quantitative magnetic
resonance imaging. Magnetic Resonance in Medicine. 2012;68(5):1423-1438.
doi:10.1002/mrm.24142
4. de Buck MHS, Hess AT, Jezzard P. Optimized DANTE preparation
for intracranial DANTE-SPACE vessel wall imaging at 7T. Proceedings of the 29th
Annual Meeting of ISMRM. Published online 2021:4201.
5. Padormo F, Beqiri A, Hajnal JV, Malik SJ. Parallel
transmission for ultrahigh-field imaging. NMR in Biomedicine.
2016;29(9):1145-1161. doi:10.1002/nbm.3313
6. Hennig J. Multiecho imaging sequences with low refocusing
flip angles. Journal of Magnetic Resonance (1969). 1988;78(3):397-407.
doi:10.1016/0022-2364(88)90128-X
7. Weigel M. Extended phase graphs: Dephasing, RF pulses, and
echoes - Pure and simple. Journal of Magnetic Resonance Imaging.
2015;41(2):266-295. doi:10.1002/jmri.24619
8. Augst AD, Barratt DC, Hughes AD, Thom SAM, Xu XY. Various
issues relating to computational fluid dynamics simulations of carotid
bifurcation flow based on models reconstructed from three-dimensional
ultrasound images. Journal of Engineering in Medicine. 2003;217(5):393-403. doi:10.1243/095441103770802568
9. Ehses P, Brenner D, Stirnberg R, Pracht ED, Stöcker T.
Whole-brain B1-mapping using three-dimensional DREAM. Magnetic Resonance in
Medicine. 2019;82(3):924-934. doi:10.1002/mrm.27773
10. Teeuwisse WM, Brink WM, Webb AG. Quantitative assessment of
the effects of high-permittivity pads in 7 Tesla MRI of the brain. Magnetic
Resonance in Medicine. 2012;67(5):1285-1293. doi:10.1002/mrm.23108