Alicia Cronin1,2, Patrick Liebig3, Sarah Detombe4, Neil Duggal1,4, and Robert Bartha1,2
1Medical Biophysics, Western University, London, ON, Canada, 2Centre for Functional and Metabolic Mapping, Robarts Research Institute, London, ON, Canada, 3Siemens Healthcare GmbH, Erlangen, Germany, 4Clinical Neurological Sciences, University Hospital, London Health Sciences Centre, London, ON, Canada
Synopsis
Degenerative
cervical myelopathy (DCM) is a degenerative disease of the spine that leads to
compression and neurological dysfunction. Recovery after surgery can be impacted
by hypoxia in the cord, however the magnitude of this effect is currently
unknown. Chemical Exchange Saturation Transfer (CEST) can produce contrast related
to tissue pH, an indicator of hypoxia, but the method works best at ultra-high
fields. Performing CEST in the spinal cord is also complicated by respiratory
and cardiac motion and cerebrospinal fluid flow. The purpose of this work was
to optimize pH-weighted CEST imaging in the human spinal cord at 3 Tesla.
Introduction
Degenerative cervical myelopathy (DCM) is a degenerative disease of the
spine that causes compression of the spinal cord, leading to neurological
dysfunction.1 Ischemia and hypoxia in the cord, caused by this
compression, could impact recovery after decompression surgery. Unfortunately,
direct in-vivo evidence of hypoxia and ischemia has been limited in humans. Chemical
Exchange Saturation Transfer (CEST) is an MRI contrast derived from the
transfer of magnetic saturation from selectively excited endogenous
exchangeable protons, like amide and amine protons, to bulk water protons,2
causing a signal reduction in the observed water signal.3 The rate of
exchange is pH dependent, and pH is altered by hypoxia. We have previously
shown that pH-weighted CEST contrast can be generated in the brain using a
ratiometric method called amine/amide concentration-independent detection
(AACID).4 The use of low power radiofrequency (RF) pulses to
saturate protons up field from water also produces contrast attributed to
nuclear Overhauser enhancement (NOE),5,6 which originates from
mobile macromolecules. Previous studies found that changes in tissue pH may
cause the NOE effect to be lower in tumours, but this relationship is unclear
and requires further study.5,7,8 Both AACID and NOE CEST contrast in
the spinal cord could provide a means to examine pH change caused by hypoxia. Therefore,
the objective of this study was to optimize AACID and NOE CEST contrast in the
human spinal cord, creating sufficient sensitivity and resolution to detect pH
heterogeneity in DCM patients. Methods
On a 3T Siemens MAGNETOM Prisma Fit MRI scanner, a prototype CEST sequence
was used that incorporates a single slice 2D gradient echo (GRE) readout. Scan parameters
that maximized AACID and NOE CEST effects were evaluated in egg white phantoms
with differing pH. Pearson’s correlation coefficient (r) was used to assess the
linear dependence of AACID and NOE effects on pH. Scan parameters were also
varied in human brain (N = 3) CEST images to optimize in-vivo CEST contrast. Saturation
was performed using Gaussian shaped radiofrequency (RF) pulses (pulse train
length = 30, pulse length = 100 ms, interpulse delay = 1 ms) applied at 132
offsets from -6.5 to 6.5 ppm. The RF pulse amplitude was varied to find the
optimal B1 value for AACID and NOE contrasts. Other relevant
parameters include: TR/TE = 10.0/4.4 ms, voxel size = 2 mm x 2 mm, slice
thickness = 5 mm and 1 average. To correct for B0 inhomogeneities,
the CEST spectrum was shifted on a pixel-by-pixel basis using offsets measured
from a Gaussian fitted water saturation shift referencing (WASSR) spectrum,9 acquired using five Gaussian shaped saturation pulses (same sequence parameters
as above except saturation pulse B1 = 0. 5 µT, 25 frequency offsets
from -2.0 to 2.0 ppm). Using MATLAB, bulk water, macromolecule, NOE, amide, and
amine peaks were fitted to accurately calculate the AACID and NOE effect. In the
spinal cord, additional non-saturated scans were interleaved throughout the
acquisition and the variation of intensity was used to account for the global
effect of respiration.5 The non-saturated scans were
spline-interpolated and used to normalize the spectrum. The respiratory cycle
was measured with respiratory bellows and used to calculate respiration volume
per unit time (RVT). The RVT was scaled by the range of the signal drift of the
non-saturated images to regress signal variation due to breathing out of the
data.10 One healthy subject was recruited to evaluate CEST effect
and the respiratory correction algorithm in the spinal cord.Results
In the egg white phantoms, both the AACID (p = 0.01, r = -0.94) and NOE
(p = 0.04, r = 0.90) CEST effects were found to be linearly dependent on pH
(Figure 1). Parametric maps of the AACID value and NOE effect in human brain
show a homogeneous spatial distribution (Figures 2B and 2D). In the brain, the optimal amide CEST effect
was achieved at a B1 value of 1.5 µT and the greatest NOE effect was
observed at 0.5 µT (Figure 2). In the spinal cord, the average magnitude of the
measured amide CEST effect was 1.34% and the average magnitude of the NOE CEST
effect was 2.55% (Figure 3). Discussion
The optimized CEST sequence produced high quality images in the phantom,
brain, and spine. The pH dependence of both the AACID value and NOE effects
suggested that both could be utilized to investigate pH changes due to hypoxia at
this field strength. In the spinal cord, the detected NOE effect was more prominent
and could provide an advantage for creation of pH-weighted contrast, also reducing
SAR due to the lower B1 amplitude. Conclusion
Although CEST in the spinal cord has been demonstrated previously,10,11
here we examine a prototype CEST sequence at 3T, combined with respiratory
correction, to evaluate pH sensitive contrast using both the AACID value and NOE
effect. Further optimization in the spinal cord is required to improve the
fidelity of the measurements. In the future, the two contrasts will be compared
in the spinal cord of healthy subjects to determine the reproducibility of the
pH-weighted measurements and in DCM patients to examine the pH heterogeneity at
the site of spinal cord compression. Acknowledgements
We thank Scott Charlton and Oksana Opalevych (CFMM, Robarts
Research Institute, The University of Western Ontario) for facilitating MRI
acquisitions. References
[1] Toledano M, Bartleson JD. Cervical spondylotic myelopathy. Neurol
Clin. 2013; 31: 287-305.
[2] Sherry A, Woods M. Chemical
exchange saturation transfer contrast agents for magnetic resonance imaging. Annu
Rev Biomed Eng. 2008; 10(3/4): 391-411.
[3] van Zijl PCM, Yadav NN. Chemical
Exchange Saturation Transfer (CEST): what is in a name and what isn’t? Magn
Reson Med. 2011; 65(4): 927-948.
[4] McVicar N, Li AX, Goncalves DF,
Bellyou M, Meak SO, Prado MAM, Bartha R. Quantitative tissue pH measurement
during cerebral ischemia using amine and amide concentration-independent
detection (AACID) with MRI. J Cereb Blood Flow Metab. 2014; 34: 690-698.
[5] Jones CK, Huang A, Xu J, et al.
Nuclear Overhauser enhancement (NOE) imaging in the human brain at 7T. NeuroImage.
2013; 77: 114-124.
[6] Zhou J, Payen J, Wilson DA,
Traystman RJ, van Zijl PCM. Using the amide proton signals of intracellular
proteins and peptides to detect pH effects in MRI. Nat Med. 2003; 9(8):
1085-1090.
[7] Shen
Y, Xiao G, Shen Z et al. Imaging of nuclear Overhauser enhancement at 7 and 3
T. NMR Biomed. 2017; 30(9): 1-9.
[8] Paech
D, Zaiss M, Meissner JE et al. Nuclear Overhauser Enhancment Mediated Chemical
Exchange Saturation Transfer Imaging at 7 Tesla in Glioblastoma Patients. PloS
One. 2014; 9(8): 1-7.
[9]
Kim M,
Gillen J, Landman BA, Zhou J, van Zijl PCM. Water Saturation Shift Referencing
(WASSR) for Chemical Exchange Saturation Transfer (CEST) Experiments. Magn
Reson Med. 2009; 61: 1441-1450.
[10] By S, Barry RL, Smith AK, et al.
Amide Proton Transfer CEST of the Cervical Spinal Cord in Multiple Sclerosis
Patients at 3T. Magn Reson Med. 2018; 79: 806-814.
[11] Dula AN, Pawate S, Dethrage
LM, Conrad BN, Dewey BE, Barry RL, Smith SA. Chemical exchange saturation
transfer of the cervical spinal cord at 7T. NMR Biomed. 2016; 29:
1249-1257.