Yusuf Samet Yaras1, Lee W Bradley1, Dursun Korel Yildirim2, Ozgur Kocaturk3, John Oshinski4,5, and F. Levent Degertekin1
1Mechanical Engineering, Georgia Institute of Technology, ATLANTA, GA, United States, 2Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, MD, United States, 3Institute of Biomedical Engineering, Bogazici University, Istanbul, Turkey, 4Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology, Atlanta, GA, United States, 5Department of Radiology and Imaging Sciences, Emory University, Atlanta, GA, United States
Synopsis
Electric
field and temperature sensors are essential tools for RF safety assessment of
implants during magnetic resonance imaging. In this work, an acousto-optic
sensor was used for both local tangential electric field and temperature
measurements. The inherent electric field sensitivity of the piezoelectric
crystal mechanically coupled to a fiber Bragg grating provides the electric
field information while the thermal sensitivity of FBG is used for temperature measurement for
dual mode sensing. The sensor was used to measure the electric field
concentration around a reference implant with high SNR as well as the temperature
rise at the tip of the implant.
INTRODUCTION
One
of the most crucial magnetic resonance (MR) safety concerns is excessive
heating of metallic implants induced by the deposited high radio frequency (RF) energy.
Current standards for RF-heating assessment of medical implants require temperature
measurements (ASTM F2182-19e2) and electric (E) field measurements (ISO/TS
10974:2018). In this work, we present an acousto-optic modulator-based dual
mode sensor platform for both temperature and electric field measurements around
potential hotspots for RF safety assessment of implants.METHODS
The acousto-optic (AO) sensor consists of a piezoelectric
transducer mechanically coupled to a fiber Bragg grating (FBG) at the distal end of an optical fiber1. E field sensing relies on the inherent E field
sensitivity of piezoelectric crystals whereas temperature sensing relies on the
inherent sensitivity of FBGs to ambient temperature change2. A
schematic overview of the measurement set-up and a close-up image of the AO
sensor are shown in Figure 1a. Optical read out is achieved through side slope
detection method. An optical receiver and a narrow linewidth laser are used for
probing the FBG at a wavelength on a slope in the reflection spectrum. The central
notch was used for E field detection, as it provides the highest sensitivity,
whereas one of the side lobes was used for temperature measurement for
increased range (Figure 1b). AC coupled photodetector output is monitored for time-domain
E field measurement, whereas DC coupled photodetector output is monitored for
slow changing temperature measurement. In
this particular sensor, a π-phase shifted FBG sensor was used with 0.4pm
bandwidth and central wavelength of 1550nm (Teraxion Inc., Quebec, Canada)
embedded in a 125μm diameter optical fiber. A 140μm
thick 36o Y-cut LiNbO3 piezoelectric crystal with 1mmx5mm
lateral size (Boston Piezo Optics Inc., MA, USA) was used around its third
harmonic resonance (63 MHz) to match the Larmor
frequency at 1.5T. The AO sensor was tested in a gel phantom per ASTM F2182-19e2 standard,
using a 1.5 T scanner (Magnetom Espree, Siemens Healthineers, Erlangen,
Germany). Measurements were taken near the tips and middle of the reference
implant (1⁄8in.×10cm titanium rod), which was oriented parallel to the long
side of the phantom and placed 3cm away from the phantom wall. E field measurements were performed
using a gradient echo (GRE) sequence (flip angle, 45°; TR, 24.78ms; TE, 1.6ms;
slice thickness, 7mm; bandwidth, 651Hz/pixel and matrix size, 192×192),
whereas the heating experiment was conducted using a bSSFP sequence (flip
angle, 66°; TR, 31.79ms; TE, 0.96ms; slice thickness, 10mm; bandwidth,
977Hz/pixel and matrix size, 128×128). Temperature response of the AO
sensor was characterized in a temperature-controlled water bath and reference
measurements were performed using a J type thermocouple (Extech Instruments,
NH, USA). RESULTS
Time-domain
E field signal measurement at the geometric center of the phantom using the AO
sensor for the GRE sequence is given in Figure 2a. Frequency spectrum of the
same signal shows an SNR over 40dB at a flip angle of 45°
(Figure 2b). E field measurements at various locations of the phantom including
near the tips and the midpoint of the Ti rod as well as away from the Ti rod at a
reference point are shown in Figure 3. E field was elevated slightly at the
midpoint (0.24V) and significantly at the tips (0.96V and 1.04V) compared to
the reference point (0.17V), as predicted. Temperature rise at the tip of the
Ti rod was measured by probing the side lobe of the FBG. DC output voltage of
the photodetector increased from 4V to 5.7V during a 6-minute-long high SAR (bSSFP)
sequence (Figure 4a). A comparison between the AO sensor and thermocouple is
given in Figure 4b. Temperature change of 1.3oC was measured for the
same voltage range observed during the heating test. Thermal sensitivity of the
AO sensor is calculated 1.3V/oC with an accuracy of ±0.11oC
in this particular configuration. DISCUSSION
The AO
sensor was able to capture the E field elevation around a reference implant in
a phantom study with high SNR. SNR of the AO sensor is more than sufficient for
E field mapping purposes, whereas temperature measurement accuracy is adequate
for the ASTM standard with limited temperature range. Although temperature and E
field measurements were not taken simultaneously in this experiment, simultaneous
measurement can be achieved by adding wide range wavelength tracking capability
to the read-out electronics. The E field
can be still measured through the side slope, whereas temperature can be
measured by tracking the slow drift in the central wavelength with greater
accuracy and dynamic range. In this study, AO sensor size was limited by the
piezoelectric crystal, which can be decreased by using a smaller crystal or
implementing a thin piezoelectric film coating over the FBG. CONCLUSION
We performed proof-of-principle experiments on a dual mode acousto-optic based RF safety sensor to measure both electric field and temperature.Acknowledgements
Research reported in this abstract was
supported by National Institute of Biomedical Imaging and Bioengineering of the
National Institutes of Health under award number 1R01EB029331. The content is solely the responsibility of
the authors and does not necessarily represent the official views of the
National Institutes of Health. References
1. Yaras YS, Yildirim DK, Kocaturk O, Degertekin FL.
Sensitivity and phase response of FBG based acousto-optic sensors for real-time
MRI applications. OSA Contin. 2020;3:447-458.
2. Rao YJ, Webb DJ, Jackson DA , et al. In-fiber
Bragg-grating temperature sensor system for medical applications. J Light Technol.
1997;15:779-785.