Matthias Weigel1,2,3, Peter Dechent4, Riccardo Galbusera1,2, Erik Bahn5, Ludwig Kappos1,2, Wolfgang Brück5, Christine Stadelmann5, and Cristina Granziera1,2
1Translational Imaging in Neurology (ThINk) Basel, Department of Biomedical Engineering, Faculty of Medicine, University Hospital Basel and University of Basel, Basel, Switzerland, 2Neurological Clinic and Policlinic, MS Center and Research Center for Clinical Neuroimmunology and Neuroscience Basel (RC2NB), University Hospital Basel and University of Basel, Basel, Switzerland, 3Dept. of Radiology, Division of Radiological Physics, University Hospital Basel, Basel, Switzerland, 4Department of Cognitive Neurology, MR-Research in Neurology and Psychiatry, University Medical Center Göttingen, Göttingen, Germany, 5Institute of Neuropathology, University Medical Center Göttingen, Göttingen, Germany
Synopsis
Balanced steady state free precession (bSSFP) sequences provide the
highest signal intensity per unti time, which makes them virtually
predestined for ultra-high resolution MR imaging at 3T. Their
sensitivity to susceptibility effects and demand for high
performance, however, represent two major drawbacks. It will be shown
that a carefully chosen protocol, which also includes common phase
cycling techniques, will enable artifact-free 200-microns isotropic
bSSFP acquisitions of the entire human fixed brain within
approximately 27h.
Introduction
Ultra-high resolution imaging (URI) of
the entire human ex vivo brain has become very popular lately 1-5.
The possibility to use extended acquisition times under ex vivo
conditions leads to impressive MR images that are valuable for both
clinical and scientific research. Recent publications were able to
achieve isotropic resolutions up to 100µm (7T) and 160µm (3T),
respectively, making use of T2* weighted RF spoiled gradient echo
(FLASH) sequences with very low receiver bandwidth 4,5.
Such FLASH sequences have the advantage to provide a relatively high
SNR, impose a low demand on the MR system hardware, and also
demonstrate a very good long term stability during measurement 5.
Yet, FLASH sequences, belonging to the
class of steady state MR sequences – being the methods of choice
under ex vivo conditions 5
– are still inferior in terms of signal acquisition efficiency
compared to the balanced steady state free precession (bSSFP)
sequences. In fact, Pefferbaum et al. already presented bSSFP imaging
with a very high 200µm in-plane resolution but thick 1mm slices on
an isolated cerebellum specimen 1;
and Tendler, Miller and co-authors perform isotropic 330µm bSSFP
imaging including phase cycling as a standard for ex vivo examinations 2,3.
In this work, we present initial results
regarding our investigation of the viable boundaries of a
sustainable, very-high isotropic spatial resolution bSSFP based MRI approach, dedicated for whole-brain ex vivo acquisitions on a
standard 3T MR system.Methods
The brain of a patient with MS was fixed in 4% formalin 24h after death. For MRI
acquisition, the brain was positioned in a dome-shaped container as
depicted in 6-8
and immersed in a fluorinated oil (Fomblin).
Air bubbles were aspirated through the spout of the container through
a vacuum pump. All acquisitions were performed with a 3T whole-body
MR system using a standard 20-channel head coil.
In a pre-experiment fast T1 maps were
acquired according to 9,
T2 maps were fit based on single spin echo acquisitions repeated five
times with the different TE={8,16,24,32,40}ms. As a result, based on
published theory 10,
the optimal bSSFP flip angle was determined accordingly.
For bSSFP based URI, an in-house bSSFP sequence was programmed that
circumvents typical restrictions like maximal 3D matrix size and that
can exploit the MR system’s full hardware performance on demand.
In another pre-experiment, test
acquisitions with fast bSSFP sequences of reduced 400µm isotropic
resolution were conducted using different excitation flip angles
of {30°,45°,60°,75°,90°}. The residual MR parameters were very
similar to the following URI variant of full (200μm)3
isotropic resolution: matrix 960x784x640, transverse slices with
readout in A-P direction, TR/TE=13.0ms/6.5ms, bandwidth=290Hz/Px,
flip angle 80°, four different phase cyclings of {0°,90°,180°,270°}
were applied 11, TAbase=01:48:51h, altogether
15 repetitions were measured for later averaging (TAtotal=27.5h).
This URI protocol could be sustained continuously by our MR system
over a long period of time (cf. Discussion).Results
Figure 1 shows a contrasting
juxtaposition of the flip angle test sequences. For our experiments,
the optimal flip angle was ca. 45° to 50° 10,
however, we decided using 80° instead to introduce a stronger T2
sensitivity. The resulting signal loss of a few percent was tolerated accordingly.
Figure 2 investigates the (changing) behavior of the field
inhomogeneity induced dark band artifacts. As a result, Figure 3
depicts the mean images for the different applied phase cyclings. The
resulting combined (final) image demonstrates a considerably improved
SNR and homogeneity (Figure 3).
Figure 4 shows examples of reformations
underlining the feasibility of the used bSSFP approach.Discussion and Conclusion
Based on their acquisition efficiency,
bSSFP sequences are predestined for ex vivo URI applications. Two
drawbacks, however, are the banding artifacts and the demand for high
performance regarding the gradient and RF system. Miller et al and
Tendler et al. observed that the banding artifacts can be tackled by
using common phase cycling schemes for their 330μm
acquisitions 2,3.
Indeed, our initial results confirm that this observation is still true even for
our 200μm acquisitions, despite the “long TR” of 13ms and
despite of additional drift effects over
the long scan times.
The high performance demand posed a
challenge, however, and eventually led to a mandatory reduction of
the overall sequence performance: An initial 170μm protocol
was not sustainable and stopped the MR system after a few minutes
(entire gradient system failure - a reproducible error). It could be
argued though that a radial k-space sampling scheme would be less gradient
demanding overall, yet, such a methodology would introduce other
challenges for reconstruction and stability.
Moreover, a different 180μm protocol was
sustainable overnight but lead to a serious helium boil-off in the MR
system. Consequently, and for the moment, the isotropic resolution
was reduced to 200μm, the overall gradient performance to 90% of
maximum, and at last, RF pulses of longer 2.56ms duration were
applied to reduce the permanent strain on the RF system as well.
To conclude, based on their
acquisition efficiency, bSSFP sequences
represent an interesting and valuable alternative compared to
FLASH even in the 200μm URI regime
on a clinical 3T MR system, provided that care is taken with the
demands imposed by the bSSFP sequence.Acknowledgements
This
work was funded by the Swiss National Fund PP00P3_176984 and
supported by the German Ministry of Education (BMBF; KKNMS German
competence network for multiple sclerosis).References
1. Pfefferbaum A, Sullivan EV, Adalsteinsson E, et al. Postmortem MR
imaging of formalin-fixed human brain. Neuroimage 2004;21(4):1585-95.
2. Miller KL, Stagg CJ, Douaud G, et al. Diffusion
imaging of whole, post-mortem human brains on a clinical MRI scanner.
Neuroimage 2011;57(1):167-181.
3. Tendler BC, Hanayik T, Ansorge O, et al. The Digital Brain Bank,
an open access platform for post-mortem datasets. bioRxiv preprint
doi: https://doi.org/10.1101/2021.06.21.449154
4. Edlow B, Mareyam A, Horn A, et al. 7 Tesla MRI of the ex vivo
human brain at 100 micron resolution. Sci Data
2019 Oct 30;6(1):244.
5. Weigel M, Dechent P,
Galbusera R, et al. Imaging multiple sclerosis pathology at
160μm isotropic resolution by human
whole-brain ex vivo magnetic resonance imaging at 3 T. Sci Rep 2021
Jul 29;11(1):15491. doi: 10.1038/s41598-021-94891-1.
6. Luciano NJ, Sati P, Nair G, et al. Utilizing
3D Printing Technology to Merge MRI with Histology: A Protocol for
Brain Sectioning. J Vis Exp 2016 Dec 6;(118).
7. Absinta M, Nair G, Filippi M, et al. Postmortem
Magnetic Resonance Imaging to Guide the Pathologic Cut:
Individualized, 3-Dimensionally Printed Cutting Boxes for Fixed
Brains. J Neuropathol Exp Neurol. 2014;73(8):780-8.
8. Griffin AD, Turtzo C, Parikh G, et al. Traumatic microbleeds
suggest vascular injury and predict disability in traumatic brain
injury. Brain 2019;142(11):3550-3564.
9. Wang X, Roeloffs V, Merboldt KD, et al.
Single-shot Multi-slice T1 Mapping at High Spatial Resolution
– Inversion-Recovery FLASH with Radial Undersampling and Iterative
Reconstruction. The Open Medical Imaging Journal 9(1):1-8
10. Scheffler K. On the transient phase
of balanced SSFP sequences. Magn Reson Med 2003 Apr;49(4):781-3. doi:
10.1002/mrm.10421.
11. Bangerter NK, Hargreaves BA,
Vasanawala SS, et al. Analysis of multiple-acquisition SSFP. Magn
Reson Med. 2004 May;51(5):1038-47. doi: 10.1002/mrm.20052.