Robert Stormont1,2, James Ross1, Gareth Davies1, Vasiliki Mallikourti1, Amnah Alamri1, Lionel Broche1, and David Lurie1
1Medical Physics, University of Aberdeen, Aberdeen, United Kingdom, 2GE Healthcare, Waukesha, WI, United States
Synopsis
A 4-Channel Head Coil RX Array has been created supporting
Field-Cycling Imaging studies at an 8.5 MHz readout frequency. Head imaging to date, has been accomplished
with a quadrature TX/RX birdcage coil.
The array goals include improved SNR, enhanced patient
comfort, peripheral compatibility, with fast application to various sized
heads.
Evaluation with healthy volunteers is now underway. The array appears well tolerated by individuals
when compared to the 8-rung birdcage, accommodating a communication headset, and
leaving the face quite open.
SNR is comparable to the birdcage at the image center
and improved as you move towards the array elements.
Introduction
Volunteer experiments with a Field-Cycling
Imaging (FCI-MRI) scanner are underway, studying relaxation behaviors of
various healthy and diseased tissues1 . This technology derives from MRI and, generally,
reception is done at or near 0.2 T (protons resonating near 8.5MHz). Head studies have utilized a rigid quadrature
TX/RX birdcage with careful tuning per volunteer to optimize the performance,
but this is impractical and required careful procedures that were lengthy and
prone to error.
A 4-channel, close-fitting
flexible head array has now been developed targeting improved SNR, enhanced
patient comfort, peripheral compatibility, with fast application to various
sized heads. The coil consists of 4
square 17.5 cm elements packaged in closed cell neoprene foam and interfaced to
low impedance preamplifiers. The
elements overlaps adjust for smaller or larger head compatibility. Methods
Anthropometric data2 indicates a coil
accommodating a 60 cm head would cover the majority of large volunteers, and a
coil accommodating 55 cm would work for the majority of smaller volunteers.
Four 17.5 cm sided square loops, formed onto a 62 cm cylinder (2 cm to account
for about 1 cm thick packaging), resulted in 2cm overlaps. The overlaps are increased to 3.25 cm to form
onto a 57 cm (2 cm to account for about 1 cm thick padding) cylinder. The element length was chosen for good brain
to brain stem coverage and to try and keep the face element open below the
mouth.
Based on earlier positive work with multi-turn litz wire3, the
loops were implemented at two turns utilizing HF -Litz 1600 strand x 0.020mm
Elektrisola Co., Reichshof-Eckenhagen, Germany.
Because of the flexible, adaptable
nature of the coil and its’ variable overlaps, preamplifier decoupling was essential for the
design4. Enhanced
preamplifier decoupling was achieved by utilizing a high matching impedance
transformation with a low input impedance amplifier optimized for a >150-ohm
source (WMA-08HB WanTcom Inc.,
Chanhassen, MN, USA). The preamplifier,
matching, and protection circuitry were packaged and located 8 cm from the loops,
superior to the head.
Results
The two-turn, 17.5 cm square loops, spaced from the head by approximately
1 cm of neoprene packaging, exhibited an unloaded to loaded Q ratio of 2.5.
The WMA-08HB preamplifier utilized high Q inductors and achieved an input
impedance of 2.4 ohms. This yielded an 82-ohm
blocking impedance when interfaced with a 14-ohm impedance inverter (Figure
1).
The packaging resulted in a 15 cm square opening for anterior and lateral
elements. The elements snap together to set overlaps, and the anterior and a
side element are generally left unsnapped until the volunteer is positioned (Figure
2).
Loop to loop isolation was measured in the 60 cm and 55 cm accommodating
over-lap positions (Figure 3).
The array SNR at the center of a phantom is improved by approximately 7% versus
the quadrature-birdcage coil (Figure 4).
4-channel, healthy volunteer images are now been gathered and evaluated
(Figure 5).Discussion
The loaded to unloaded Q-ratio
shows that the coils are in the desired body dominated regime, where SNR losses
due to finite loop conductance is minimised.
Isolation measurements indicate further improvement in blocking
impedance and/or better overlap controls may benefit the array. As expected, the overlaps at the 55 cm coil
circumference were closer to the critical overlap and isolation was improved.
In spite of the improved SNR versus the birdcage coil, volunteer
image non-uniformity and ringing artifacts need to be investigated and improved
to see the array as an overall IQ winner.
Set-up time is very quick, with the array quickly snapping over the
volunteer after positioning into the opened array.
Volunteer feedback with the array has been positive. Comfort, openness, and peripheral (headset)
compatibility is viewed as generally superior to the birdcage.Conclusions
A four channel, flexible and open head array been developed for use
in in an experimental Field-Cycling Imaging system.
Volunteer reaction has been positive, which largely improves
experimental conditions for human scans.
With coil and reconstruction improvements, the array has the
potential to be the preferred coil for head imaging in the Field-Cycling
Imaging system.Acknowledgements
No acknowledgement found.References
1 Broche LM, Ross
PJ, Davies GR, MacLeod MJ, Lurie DJ. A whole-body Fast Field-Cycling scanner
for
clinical molecular
imaging studies. Scientific Reports. 2019;9(1):10402.
doi:10.1038/s41598-019-46648-0.
2 Bushby KM, Cole
T, Matthews JN, Goodship JA. Centiles for adult head circumference. Arch
Dis Child. 1992;67(10):1286-1287. doi:10.1136/adc.67.10.1286.
3 Grafendorfer T, Conolly S, Matter N, Pauly J,
Scott G. Optimized Litz Coil Design for Prepolarized Extremity MRI. in 14th
ISMRM (Seattle, Washington).
4 Roemer PB, Edelstein WA, Hayes
CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med (1990)
16:192–225. doi:10.1002/mrm.1910160203