Neta Stern1, Kai T Block2, Chen Solomon1, Tamar Blumenfeld-Katzir1, and Noam Ben-Eliezer1,2,3
1Biomedical Engineering, Tel Aviv University, Tel Aviv, Israel, 2New-York University Langone Medical Center, New York, NY, United States, 3Sagol School of Neuroscience, Tel Aviv University, Tel Aviv, Israel
Synopsis
To optimize the use of radial spin-echo sequences, SNR characterization for relevant T1 and T2
ranges is required. In this work, we studied the interplay
between the number of excitations (Nexc) and the turbo factor (NTF
- number of echoes within each TR) on the SNR of images acquired using the
RAISE (RAdIal Spin Echo)
sequence. SNR was evaluated for phantom and in-vivo images. Phantom analysis
focused on SNR per T1 and T2 ranges appropriate for known
gray and white matter relaxation times. Selected parameter sets were later used for in vivo scans for qualitative evaluation.
Introduction
Radial sampling offers several advantages. One is its
relative robustness to motion artifacts. A second advantage is the ability to
acquire images at arbitrary field of views (FOVs), i.e., meaning, field of views that
are located inside the imaged object, without being affected by aliasing
artifacts.1,2 This has the potential to
increase protocol flexibility, and reduce scan time, or alternatively increase
image resolution while maintaining original scan time. Obtaining high
spatial resolution images could benefit many applications, one of them is the
imaging of small CNS structures.3,4 To optimize the use of these
sequences, SNR characterization for T1 and T2 ranges relevant for brain tissues
is required.
In this work, we studied the effect of the interplay between
the number of excitations (Nexc) and the turbo factor (NTF
- number of echoes within each TR) on the SNR of images acquired using the
RAISE5 (RAdIal Spin Echo)
sequence - a spin-echo T2 weighted protocol. SNR was evaluated for phantom and
in vivo images. Phantom analysis focused on SNR per T1 and T2
ranges appropriate for known gray and white matter relaxation times. Parameter sets that produced the maximal evaluated SNR were later
used כםר in vivo scans for qualitative evaluation.Methods
All scans were performed on a 3T Siemens Prisma scanner at Tel-Aviv University.
Phantom Scans
Phatnom scans were performed using the ISMRM-NIST system phantom6,7. Data were acquired
using the RAISE sequence. Scan parameters were: slice
thickness=1 mm, matrix size=192x192, voxel size=1x1x1 mm3, TE/TR=10/2200
ms, Nexictations=128/64, Nturbo factor=8/6/4, acquisition BW=430 Hz/Px, Nrepeatitions = 16.
Brain Scans
In vivo scans were performed after obtaining informed consent and under the
approval of the local Helsinki and ethics committees. Brain imaging was done
for two healthy volunteers. Data were acquired using the same RAISE sequence.
Scan parameters for the first volunteer (full FOV) were: slice
thickness=1 mm, matrix size=192x192, voxel size=1x1x1 mm3, TE/TR=10/2200
ms, Nexc=128/64, NTF=8/6/4, acquisition BW=430 Hz/Px, Nrepeatitions
= 1.
For the second volunteer, three parameter sets were tested
on two different internal FOVs. Common parameters for all scans were: slice
thickness=1 mm, matrix size=80x80, voxel size=1x1x1 mm3, TE=10 ms, acquisition
BW=430 Hz/Px, Nrepeatitions = 1. Three parameters were changed
between scans: TR=4240/2200/2200 ms, Nexc=128/128/64, NTF=16/8/8.
TR for the first parameter set was increased because of SAR limitations.
SNR evaluation
SNR was
evaluated by performing repeated “identical” acquisitions of the phantom using
the gold standard method, according to which the SNR of a single voxel is equal
to the ratio between the mean value and the standard deviation of the signal
intensity time course.
Statistical analysis
Paired
t-test was used to compare the SNR values resulting from applying parameter
sets (Nexc=128,NTF=4) and (Nexc=64,NTF=6)
for each of the six selected ROIs.
Results
Fig. 1 shows images that were acquired from the first
out of 16 iterations performed for each parameter set. Parameter sets vary in
the number of excitations (Nexc) and in the Turbo factor (NTF) used. Total number of spokes is equal to (Nexc x NTF).
Fig. 2 shows the SNR maps that were evaluated per each
parameter set. Generally speaking, SNR decreases as the number of spokes decreases;
however, as can be seen when comparing (c) and (d), who were obtained using the
same total number of spokes, the balance between Nexc and NTF affects
the resulting SNR; the SNR map shown in (c), which was evaluated from images
that were acquired with more excitations and a lower turbo factor, generally
contains higher values than the SNR map shows in (d), although the number of
spokes used is equal in both.
Fig 3a shows marked ROIs for six selected
spheres (reference T1\T2 values: sphere 1 – 503.30/33.25
ms, sphere 2 - 703.51/49.26 ms, sphere 3 – 937.12/70.57 ms, sphere 4 –
1202.40/99.89 ms, sphere 5 – 1774.24/188.16 ms, sphere 6 - 1801.61/192.23 ms). Fig.
3b – 3g show SNR Measurements for the selected spheres, per each parameter
set.
In vivo demonstration
of selected parameter sets can be seen
in Fig. 4 and in Fig. 5. Fig. 4 shows a demonstration of a
full FOV in vivo brain scans using four parameter sets. Fig. 5 presents images from two partial FOV in vivo
brain scans using three tested parameter sets.
The six pairs of SNR value groups resulting from applying
parameter sets (Nexc=128,NTF=4) and (Nexc=64,NTF=6)
for each of the six selected ROIs were found to be significantly different (p-value
< 0.0005 for all pairs).Discussion
Several aspects should be taken into
consideration when determining scan parameters for RAISE. High SAR levels and long
scan times are limiting factors on the way to achieve high quality images. Increasing
Nexc increases SNR, but also linearly
increases scan time. Increasing NTF has the potential to increase SNR (but not always, as can be seen
in fig. 3b-c), and also to achieve better T2
contrast and produce heavily T2-weighted images thanks to the
participation of later echoes (as seen in fig. 5a, 5d). However, increasing the
turbo factor costs in SAR, which can be translated into again an increase in
scan time. Further work is required to test SNR dependence in
additional parameters, such as acquisition bandwidth and partition order.Acknowledgements
No acknowledgement found.References
1. Block KT, Chandarana H, Milla S, et al. Towards Routine
Clinical Use of Radial Stack-of-Stars 3D Gradient-Echo Sequences for Reducing
Motion Sensitivity. 2014;18(2):87-106.
2. Block KT. Advanced Methods for Radial Data Sampling in Magnetic
Resonance Imaging Dissertation. 2008.
3. Trampel R, Bazin PL, Pine K, Weiskopf N. In-vivo magnetic
resonance imaging (MRI) of laminae in the human cortex. Neuroimage.
2019;197(April 2017):707-715. doi:10.1016/j.neuroimage.2017.09.037
4. Li Z, Chen YA, Chow D, Talbott J, Glastonbury C, Shah V.
Practical applications of CISS MRI in spine imaging. Eur J Radiol Open. 2019;6(May):231-242.
doi:10.1016/j.ejro.2019.06.001
5. Ben-Eliezer N, Sodickson DK, Shepherd T, Wiggins GC, Block KT.
Accelerated and motion-robust in vivo T 2 mapping from radially undersampled
data using bloch-simulation-based iterative reconstruction. Magn Reson Med.
2016;75(3):1346-1354. doi:10.1002/mrm.25558
6. Keenan KE, Biller JR, Delfino JG, et al. Recommendations
towards standards for quantitative MRI (qMRI) and outstanding needs. J Magn
Reson Imaging. 2019;49(7):e26-e39. doi:10.1002/jmri.26598
7. Russek SE, Boss MA, Jackson EF, et al. Characterization of
NIST/ISMRM MRI System Phantom. in Proceedings of the 20th meeting of the
International Society of Magnetic Resonance in Medicine, Melbourne, Australia
2012