Hacene Serrai1
1Carle Foundation Hospital, Center for Clinical Imaging Research, Urbana, IL, United States
Synopsis
It
has been shown that RAPID-SI, implemented at 3 Tesla magnet, allows for
reduction of acquisition time while providing accurate metabolite information
as compared to CSI. To benefit from the higher sensitivity of the Ultra-High-Field
(UHF), RAPID-SI was implemented on a Siemens Terra 7T scanner, and phantom MRSI
data were acquired, with less averaging, and compared in terms of acquisition
time, signal-to-noise ratio (SNR), and data analysis to those collected using
CSI method. As expected, and in addition to the acquisition time reduction,
proportional to the speed factor R, RAPID-SI provides similar quantification
results with similar SNR values, afterward data reconstruction, as compared to
CSI.
Introduction
To
reduce the long acquisition time in MRSI, a chemical shift imaging technique
named Reduction of Acquisition time by Partition of the sIgnal Decay in
Spectroscopic Imaging (RAPID-SI), which uses blipped gradients inserted during
signal acquisition, has been proposed [1]. In addition to reducing acquisition
time in proportion to an acceleration factor R (set equal to the inserted
blipped gradients; Figure 1), RAPID-SI has been shown at 3 Tesla to provide
comparable quantification results to CSI. However, the partition of the signal
into sub-signals for spatial encoding purposes reduces the sensitivity of the
acquired data by √R, which is compensated during data reconstruction (Figure 1,
2) [1]. To benefit from the higher sensitivity offered at Ultra-High Field
(UHF), RAPID-SI was implemented on a 7 Tesla whole body magnet, and phantom
data were acquired and compared to CSI. The results obtained demonstrate that
in addition to the expected acquisition time reduction obtained at lower field,
RAPID-SI profits from the UHF to reduce signal averaging while providing similar
results in term of data analysis and sensitivity to CSI. Methods
The
sequence part of the RAPID-SI technique, displayed in Figure 1A, was
implemented on a Siemens Terra 7T scanner, and accelerated in the
anterior-posterior (AP) direction by inserting R blipped gradients during data
acquisition with classical phase encoding applied along the right-left (RL)
direction. The Flow diagram of the data pre-processing part of the technique is
presented in Figure 2. The data pre-processing was performed offline in MATLAB.
Using the multi-transmit capabilities (pTx) available at the 7 T scanner, 2D
(8x8, 16x16) and 3D (16x16x8) RAPID-SI and standard CSI data were acquired from
a spherical phantom containing main brain metabolites (e.g: NAA, Creatine,
Choline) using the following acquisition
parameters: TE/TR=30/1600 msec, BW=2000 Hz, N = 512, avg = 1 (set to 2 at 3
Tesla [2]) , phase encoding steps (PE) = 8 and 16, and number of collected
signals (FIDs) along blip (AP) direction = 1 (for PE=8) and 2 (for PE=16). The
RAPID acceleration factor was set to R = 8 (AP direction). The FOV dimensions
along RL and AP direction = 120 mm, and slice (slab) thickness was set to 10 mm
for the 2D and to 120 mm for the 3D.Results and Discussion
As
expected, RAPID-SI shortened the acquisition time by the acceleration factor R.
Compared to previous 3T studies, this time is further reduced at 7 Tesla by
signal averaging reduction. As an example, a 2D 16x16 data set acquired in
about 7 min with CSI, was reduced to less than 1 minute with the RAPID-SI with acceleration
factor R set to 8 (Table. 1). SNR in RAPID-SI was lower than CSI by
approximately the factor √R [1, 2]. However, during data pre-processing (Figs.
1, 2), the SNR recovers to a value closer to that for CSI, as shown in Table 1
and displayed in Figure 3. Data quantification presented here as metabolite
ratios (e.g: NAA/Cr and Cho/Cr) reported in Table 1, (absolute metabolite quantification
were demonstrated elsewhere [1]), were comparable between of RAPID-SI and CSI. Further
in-vivo testing are required to
validate this encouraging preliminary phantom results.Conclusion
RAPID-SI
reduces acquisition time, while preserving metabolite information, compared to
CSI at UHF. To further reduce acquisition time RAPID-SI will be combined with
other acceleration methods such as parallel imaging or compressed sensing. In
addition, the B1 field inhomogeneity effect manifested at 7T, will
be reduced by integrating RAPID-SI approach into a localized MRS sequence employing
adiabatic RF pulses (e.g: semi-LASER [3, 4]) with in-vivo validation. Acknowledgements
This
work was supported by the Stephens Family Clinical Research
Institute, Carle Foundation Hospital, Urbana, IL, USA.References
1. Bhaduri S, Clement P, Achten E, Serrai H.
Reduction of Acquisition time using Partition of the sIgnal Decay in
Spectroscopic Imaging technique (RAPID-SI). PLOS ONE. 2018; 13(11): e0207015.
2. Bhaduri
S., et al. Proc ISMRM-ESMRMB 2018, 1057.
3. Scheenen TW, Heerschap A, Klomp DW. Towards 1H-MRSI of the human
brain at 7T with slice-selective adiabatic refocusing pulses. MAGMA. 2008 Mar; 21(1-2):95-101.
4. Scheenen TW, Klomp DW, Wijnen JP, Heerschap A.
Short echo time 1H-MRSI of the human brain at 3T with minimal chemical shift
displacement errors using adiabatic refocusing pulses. Magn Reson Med. 2008
Jan; 59(1):1-6.