Alexander Paul Neofytou1, Radhouene Neji1,2, James Wong3, Anastasia Fotaki1, Joana Ferreira1, Carl Evans1, Filippo Bosio1, Nabila Mughal1, Reza Razavi1, Kuberan Pushparajah1, and Sébastien Roujol1
1School of Biomedical Engineering and Imaging Sciences, Faculty of Life Sciences and Medicine, King's College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom, 3Department of Paediatric Cardiology, Evelina London Children’s Hospital, London, United Kingdom
Synopsis
A
novel reconstruction technique based on iterative rejection of segmented
k-space was developed for retrospective correction of respiratory motion in
multiple average, free-breathing cine images. In comparison to standard signal
averaging reconstruction, it provides higher sharpness, SNR, CNR, image
quality, and rate of diagnostic quality images.
Introduction
Cardiac cine imaging
is an essential technique in clinical cardiac magnetic resonance (CMR)
examinations and is used for the analysis of ventricular structures and function1. Cine images are
normally acquired within a breath-hold (1-2 slices/breathhold) using a
segmented bSSFP acquisition over 10-12 cardiac cycles. Unfortunately, images are severely degraded in patients unable to sustain stable breathholds such as patients with impaired lung function or patients less cooperative in following instructions (e.g. the paediatric population). A way to alleviate this
is to allow patients to breathe freely and acquire multiple signal averages
(typically three), which helps minimise motion artefacts seen in the reconstructed
images. However, free-breathing acquired images with multiple averages can sometimes suffer from important respiratory motion artefacts and be of non-diagnostic
quality. This study sought to develop a novel motion-robust reconstruction
technique for free-breathing, multiple average, cine acquisition.Methods
Proposed reconstruction:
The
proposed reconstruction is summarised in Figure 1 for a typical acquisition
with three signal averages. This technique relies on the iterative removal of
k-space segments that contribute the most to motion corruption/blurring. An image
sharpness or focus measure (image gradient energy2) was used as a surrogate for quantifying the level of motion
corruption/blurring in the cine images (i.e. the more blurry the image, the
lower the focus). At each iteration, the k-space segment across the three
corresponding k-space matrices that outputted the highest increase in focus, when removed, was removed. This process was repeated until convergence of the
image focus. A condition was set such that across the three signal averages,
one segment out of the respective set of three segments needed to remain to
ensure initial k-space under-sampling was preserved.
To
mitigate the potential loss of SNR/CNR associated with the removal of k-space segments,
this reconstruction was performed three times using different initial
conditions (removal of 2 out of 3 central segments) to generate three images. On
a slice-by-slice basis, the image series with the highest mean focus (measured
across cardiac phases) was labelled as the sharpest image series (Isharpest). The remaining two series were co-registered to Isharpest using non-rigid image registration3. An average image of these three
co-registered series was generated as the final reconstruction (Icombined3).
Evaluation:
Segmented
balanced steady-state free precession (bSSFP) cine images (short-axis stack)
were acquired free-breathing using retrospective ECG gating in 15 patients
undergoing routine CMR examination. Imaging parameters were as follows: TE/TR=~1.2/2.8ms,
Flip angle=52°, FOV=~225×300mm2, Acq.voxel size=~2.2×1.6mm2, slice
thickness=8mm, BW=930Hz/px, GRAPPA factor=2, NSA = 3, temporal resolution=~40ms, no. slices=15. Images were
acquired at 1.5T (MAGNETOM Aera, Siemens Healthineers, Erlangen, Germany). All
data were exported offline and three reconstructed datasets were generated in
Matlab for each scan: 1) standard approach/signal averaging (Istandard), 2) intermediate reconstruction (Isharpest) and 3) proposed
reconstruction (Icombined3).
Qualitative
image assessment (0 = non-diagnostic, 1 = diagnostic (sub-optimum image quality
- presence of motion artefacts/blurring), 2 = diagnostic (excellent image
quality - no motion artefacts/no blurring/sharp myocardial edges)) was
performed by consensus of two experienced CMR readers on the three
reconstruction techniques. The CMR readers were blinded from the clinical
details of the patients and the reconstruction method (presented in a
randomised order).
Quantitative
image analysis involving septal blood-myocardium sharpness4, signal-to-noise ratio (SNR) of blood
and myocardium and their respective contrast-to-noise ratio (CNR) were also
performed.Results
Example
images reconstructed from two patients using the three techniques is shown in Figure 2 and demonstrate the improved
sharpness and image quality achieved with both the “Sharpest” and “Combined3”
reconstructions. Over all patients, septal
blood-myocardium sharpness increased significantly (Figure 3a) in “Sharpest”
(0.79±0.09) and “Combined3” (0.79±0.1) in comparison to “Standard” (0.74±0.12,
p=0.004 & p=0.04, respectively). Blood SNR/myocardial SNR in “standard” (94±30/33±10)
was higher than in “Sharpest” (80±25, p=0.002/28±8, p=0.005) and tend to be lower
than in “Combined3” (105±33, p=0.02/36±12, p=0.06) (Figure 3b-c). Similarly, blood/myocardial
CNR in “Standard” (61±22) was higher than in “Sharpest” (53±19, p=0.003) and
lower than in “Combined3” (69±24, p=0.007). Image quality scores obtained with “Sharpest”
(1.8±0.2) and “Combined3” (1.9±0.2) were higher than in “Standard” (1.6±0.4,
p=0.02 & p=0.008, respectively). 94%
and 99% of slices were of diagnostic value in “Standard” and “Sharpest”,
respectively, while 100% of slices were of diagnostic value in “Combined3”
(Figure 4).Discussion
The
proposed reconstruction results in higher image quality and diagnostic rate,
which may prevent the need for sedation or repeat patient scanning. Evaluation
in a larger cohort of patients is now warranted. The motion correction
algorithm is currently computationally expensive (~1 hour to reconstruct one
short axis stack). Dedicated GPUs and parallel computing could be used to alleviate
this. The method was demonstrated using free-breathing CINE images acquired
with three signal averages, which is typically used for free-breathing CINE
imaging. However, this technique can in theory be generalised to any number of
signal averages. This technique may also be valuable for other sequences
acquired under free-breathing conditions with multiple averages, such as for
flow imaging.Conclusion
A
novel reconstruction technique based on iterative rejection of segmented
k-space was developed for retrospective correction of respiratory motion in
multiple average, free-breathing cine images. In comparison to standard signal
averaging reconstruction, it provides higher sharpness, SNR, CNR, image
quality, and rate of diagnostic quality images.Acknowledgements
This work was supported by the Engineering and Physical Sciences Research Council (EPSRC), part of the EPSRC Doctoral Training Partnership (DTP) grant (EP/R513064/1), the British Heart Foundation (BHF) (PG/19/11/34243), the Wellcome EPSRC Centre for Medical Engineering at King’s College London (WT 203148/Z/16/Z), the National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ National Health Service (NHS) Foundation Trust and King’s College London, and Siemens Healthineers. The views expressed are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health.References
- Kramer CM, Barkhausen J, Bucciarelli-Ducci C, et al. Standardized cardiovascular
magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22(17).
- Subbarao
M, Choi TS, Nikzad A. Focusing techniques. SPIE Proceedings. 1992;1823.
- Vercauteren T, Pennec X, Perchant A, et al. Diffeomorphic demons: Efficient
non-parametric image registration. NeuroImage. 2009;45(1):S61-72.
- McElroy
S, Ferrazzi G, Nazir MS, et al. Combined
simultaneous multislice bSSFP and compressed sensing for first-pass myocardial
perfusion at 1.5 T with high spatial resolution and coverage. Magn Reson Med. 2020;84(6):3103-3116.