Hanna Frantz1, Tobias Lobmeyer1, Patrick Metze1, Thomas Hüfken1, Kilian Stumpf1, Julien Rivoire2, Hizami Murad2, and Volker Rasche1
1Department of Internal Medicine II, Ulm University Medical Center, Ulm, Germany, 2RS2D, Mundolsheim, France
Synopsis
Zero echo
time (ZTE) imaging is an established approach for three-dimensional imaging of
tissues and materials with ultrashort T2* relaxation times. An intrinsic
disadvantage of this approach are missing data points in the center region of k-space
due to hardware limitations. These missing points are often retrospectively
interpolated or subsequently acquired using single-point imaging approaches or
additional ZTE acquisitions with decreased resolutions. The presented approach
relies on an interleaved combination of ZTE read-outs with different
resolutions in combination with Compressed Sensing, to acquire the k-space
center, thus enabling k-space based self-gating.
Introduction
Magnetic resonance imaging (MRI) of tissues or samples
with ultrashort T2* relaxation times such as e.g. lung1 or bone2
requires imaging approaches with short echo times (TE). Zero echo time (ZTE)
enables data acquisition with almost negligible TE by application of the
excitation pulse during the presence of a static read-out gradient. Since data
cannot be acquired during playing out the RF pulse, missing data points in the
center of k-space are one of the major limitations of ZTE, especially on
clinical scanners.
Several approaches have been introduced to acquire or
estimate missing data points3, including algebraic reconstruction of
the oversampled data, or additional acquisitions based on low-resolution ZTE or
Cartesian single-point imaging sequences. While algebraic reconstructions are
limited to dead time durations of a few Nyquist dwells4, additional
acquisitions lead to prolonged scan durations and therefore to an increased
patient discomfort.
We have introduced a concept for interfacing an
independent receive unit to a clinical scanner, thus minimizing front-end
switching times down to some µs5. However, the long
excitation pulses on clinical systems still limit the application of ZTE even
with T/R switch time in the µs range. The proposed method exploits the concept
of the ZTE sequence. But instead of performing additional acquisitions, every
n-th gradient is played out with a decreased amplitude (lower resolution), to
ensure coverage of the k-space center without any scan time penalties. Additionally,
every m-th excitation is performed without gradients, to measure the DC component (k0) in the
steady-state, which is used for gating purposes. Rising aliasing artifacts are
addressed by non-linear reconstruction techniques.Methods
All lung images were acquired during free-breathing
with a 3 T whole-body clinical imaging system (Ingenia 3.0T CX, Philips
Healthcare, Best, The Netherlands) with an independent spectrometer (MRI
Console Cameleon, RS2D, Mundolsheim, France), allowing for faster T/R
switching, as well as for higher readout bandwidths. A custom-built rectangular
single-loop coil with a size of 200 mm x 200 mm was used for signal reception.
The sequence is based on a kooshball trajectory6
but modified such that every n-th read-out gradient is executed with ¼
of its initial amplitude, presented in further detail in7. Additionally,
every 50-th excitation is performed without any read-out gradient, to measure the
center of k-space, in order to acquire a self-gating signal for multi-phase
lung imaging. To ensure sufficient respiration induced k0 signal variation,
the self-gating signal was acquired simultaneously via the Philips receive path
with an additional Flex-S (Philips Healthcare, Best, The Netherlands), placed
on the lung-liver-interface.
For image reconstruction, an in-house built
reconstruction framework, implemented in MATLAB (The MathWorks, Natick,
Massachusetts, USA) was used. The k-space density functions were calculated
using a Voronoi tessellation8. A Compressed Sensing reconstruction
was used in combination with TV sparsity transform to compensate artifacts
arising from undersampling in the outer regions of k-space and due to the
respiratory gating.
Power spectrum analysis was used to identify cut-off
frequencies used for bandpass filtering of the DC-amplitudes. The navigator
signal was then used to sort the acquired data into equidistant bins for 3
different respiratory stages.
The suggested approach was evaluated for lung imaging
on healthy volunteers after obtaining written informed consent. All relevant
scan parameters are listed in table 1.Results
Figure 1 shows a power spectrum obtained from k0 data,
which was acquired after every 50-th excitation (dt = 150 ms). The respiratory frequencies,
used for self-gating are marked in yellow and can be clearly located between
0.2 Hz and 0.3 Hz.
Figure 2 shows coronal (A-C), sagittal (D-F) and axial
(G-I) slices of the lung of a healthy volunteer. The data was acquired during
free-breathing and reconstructed with respiratory gating of 3 stages, namely
inspiration (A,D,G), expiration (C,F,I) and an intermediate state (B,E,H). As
common for ZTE sequences, the coil can be seen in the axial view, due to the
characteristic of imaging materials with ultrashort T2* relaxation times.
Furthermore, figure 2 shows the position of the lung-liver-interface in the coronal slices, marked by
the turquoise line for expiration and by the orange line for inspiration. The difference
in the respiratory amplitude was measured to be (2.89 ± 0.25) mm.Discussion and Conclusion
Replacing
ZTE read-outs with lower resolution gradients yields additional information,
missing in conventional ZTE acquisitions, within the center region of k-space
without any increase in total scan duration. Furthermore, the introduction of
gradient-free read-outs poses the possibility of obtaining k-space center data
that can be used for self-gating of lung data. The presented approach yields
qualitative ZTE images within the same acquisition duration as a standard ZTE
acquisition since no additional acquisitions or prospective triggering is
required.
Since the sensitivity of the single-loop coil is
clearly a limiting factor concerning image quality, the technique could be
further improved by choosing another coil setup. Adequate image quality would
enable quantitative analysis of functional values, such as fractional
ventilation or proton density.Acknowledgements
The authors thank the Ulm University Center for
Translational Imaging MoMAN for its support. This work was supported by DFG
RA1660/9-1. Technical support from Philips Healthcare is gratefully
acknowledged.References
-
Wild JM,
Marshall H, Bock M, et al. MRI of the lung (1/3): methods. Insights Imaging.
2012;3(4):345-53.
- Wehrli FW.
Magnetic resonance of calcified tissues. J Magn Reson. 2013;229:35-48.
- Froidevaux R, Weiger M, Brunner DO, et al. Filling the dead-time gap in zero
echo time MRI: Principles compared. Magn Reson Med. 2018;79(4):2036-2045.
- Weiger M, Brunner DO, Tabbert M, et al. Exploring the bandwidth limits of
ZTE imaging: Spatial response, out-of-band signals, and noise propagation. Magn
Reson Med. 2015;74(5):1236-47.
- Eder M,
Horneff A, Paul J, et al. A Signal Acquisition Setup for Ultrashort Echo Time
Imaging Operating in Parallel on Unmodified Clinical MRI Scanners Achieving an
Acquisition Delay of µs. IEEE Trans Med Imaging. 2020;39(1):218-225.
- Wong ST,
Roos MS. A strategy for sampling on a sphere applied to 3D selective RF pule
design. Magn Reson Med. 1994;32(6):776-84.
- Frantz H, Huefken T, Metze P, et al. Zero Echo Time Imaging Using Low
Resolution k-Space Interleaves. Poster presented at: 2021 ISMRM & SMRT
Annual Meeting & Exhibition; May 17, 2021; Vancouver, Canada.
- Rasche V, Proksa R, Sinkus R, et al. Resampling of data between arbitrary
grids using convolution interpolation. IEEE Trans Med Imaging.
1999;18(5):385-92