Lisa Y Hung1, Karl-Heinz Herrman1, and Jürgen R Reichenbach1
1Medical Physics Group, Institute of Diagnostic Radiology, Jena University Hospital, Jena, Germany
Synopsis
Some applications require T2-weighted mouse brain MRI
with isotropic high resolution (100µm3), but at 9.4T the short T2 of
mouse brain limits the echo train length of RARE sequences and exacerbates
T2-decay artifacts. Adapting a RARE sequence by using variable flip-angles
allows a reduction in scan time from 19min to 7-9min with similar image
quality. However, long RARE echo trains and strong imaging gradients can
unintentionally cause substantial diffusion weighting. To minimize the diffusion
effects crusher gradients were optimized and the variable flip angles were kept
above 40°.
Introduction
We aim to achieve artifact-free, isotropic high-resolution images of the mouse brain for neurological applications1. While 9.4T renders T1-based grey-white-matter contrast mostly useless, T2-weighted images achieve excellent anatomical contrast and can be efficiently acquired by 2D-RARE sequences. However, 2D acquisitions limit the achievable isotropic resolution due to minimal slice thickness and SAR. Using 3D acquisitions increases the acquisition time (TA), particularly since at 9.4T the short T2 of brain tissue limits echo train length with standard 180° refocusing pulses. The use of refocusing pulses with constant reduced flip angles was shown2 to prolong the echo train to facilitate anatomic mouse brain scans within 20-40min.
Our aim here was to reduce the acquisition time for T2w-3D-RARE by applying variable flip angles (VFAs) to shape the echoes' signal envelope3-4while preserving the image quality.Material & Methods
All data were acquired on a Bruker BioSpec 94/20USR with
660mT/m gradients using a helium-cooled 2-channel transceiver quadrature
surface coil (CryoProbe, Bruker Biopsin, Ettlingen). All mice were scanned in vivo under Isoflurane anesthesia.
The proposed varFlipRARE sequence uses three
initialization RF pulses4 to drive
the magnetization into a pseudo-steady-state matching the
target FA<< 90°. Subsequently, three acquisition phases can be adjusted
separately, each allowing for a linear transition from one FA to another within
a given number of echoes (Figure 3). Typically, the first phase comprised the
first 1/3 of echoes, repeating the target FA or slightly increasing it. In the
second phase, the FA increased to ~90° over 1/2 of the total echoes, followed
by the last phase, which ramped the FA up to 160-180°. Usually, k0 was acquired at the
center of the echo train.
The general scan parameters for all varFlipRARE
acquisitions were TA=7min 15s, TR=2200ms, isotropic resolution 100µm3, ΔTE=2.45ms,
TF=70, BW=150kHz. The excitation and refocusing pulses
were non-selective hard pulses with duration 128µs (exc) and 80µs (rfc)
corresponding to 10kHz bandwidth to minimize echo spacing. The phase encoding
gradients were 430µs long and the read and slice crusher gradients were played
out during the same time frame at 10% (approx. 66mT/m).
To explore further possibilities of applying VFAs, a faster
sequence with RARE factor TF=140 with TA=4min was used, setting the target FA
to 30-35°, following similar VFA increase patterns as stated above. The second
variation aimed to increase the resolution to 76µm3 with TF=86 and
TA=24min, ΔTE=3ms, TR=2700ms, and 2 averages.
For reference, a RARE acquisition2 with
slab-selective 90° excitation and 125° refocusing RF pulses was used with
TA=19min, TR=1600ms,
isotropic resolution 100µm3, ΔTE=5ms, TF=20, BW=150kHz.
All sequences used full Nyquist sampling of the
18x14x10mm3 FoV and were acquired in coronal orientation, i.e. read out
head-feet, phase left-right, and slice ventral-dorsal direction. All shown
images were reconstructed with T2-compensation and Gibbs-ringing-filter for
basic artifact reduction2. SNR and CNR values were
calculated from the mean signal in ROIs and the standard deviation in artifact-free
background areas. Results
Figure 1 presents the reference image (a) and images
acquired with the faster varFlipRARE sequence (b-i) for different FA schemes. With a low target FA=30° (h, i) a distinctly lower signal for
the ventricles(Vs) was observed. For a higher target, FA≈40° (f) or when k0 was shifted to 120ms (FA≈35°, e)
the T2- and ventricle-contrast was improved. The contrast could be further
enhanced by immediately ramping up to ~50° from the target FA≈40° (c,d) in the first phase. To further improve ventricle signal TR was
increased to 2700ms (TA=9min, b). The SNR grey matter (GM)=35, white matter(WM)=27,
V=51 and CNR CGM-WM=9, CV-GW=15, CV-WM=24 in
Fig 1b and c both match the reference image in GM and WM, the ventricle
contrast is 15-20% lower. The signal
envelope along the echo train is given below each MR image for the 2 channels
of the CryoProbe, providing more insights into the actual signal behavior.
To assess the effect of the crusher gradients in read and
slice direction Figure 2 shows an image comparison. In 2b) no crusher was used
at all, leading to some ripple artifacts in the frontal olfactory bulb.
Increasing the slice-crusher to 10% amplitude doesn't create a visual
difference.
Increasing the
read-crusher to 10% removes the ripple artifacts, but at 20% the frontal bulbous
and ventricles lose signal.
Figure 2e-f shows the resulting images using TF=140. The
low ventricle contrast in e) was improved by a faster increase of the target FA
(f). The higher resolution scans with 76µm are shown in Figure 2g-h.Discussion
The use of VFAs together with faster non-selective
acquisitions reduced the scan time from 19min to 7-9min achieving similar image
contrast and quality as the reference. Reducing the scan time further to
TA=4min leads to SNR below 10 due to unintentional diffusion weighting as has
been described before5. Simulations5 showed that the use of FAs
lower than 40° strongly enhances the effective b-value and 99% of the
b-contribution arises from the readout gradients,
leading to water-signal suppression. Increasing the read crushers to remove
artifacts will further increase the diffusion weighting and has to be carefully
chosen.Acknowledgements
We would like to thank Matthias Weigel for some
enlightening discussions on RARE sequences and VFAs. The small animal scanner
was provided by a large system grant from the DFG (INST 1757/18-1). This work
was supported by the German Research Foundation (DFG) within the Research Unit
Programme FOR5151 "QuaLiPerF (Quantifying Liver Perfusion–Function
Relationship in Complex Resection - A Systems Medicine Approach)" grant
number 436883643.References
1 Gaser
C, Schmidt S, Metzler M et al. Deformation- based brain morphometry in rats. Neuroimage.
2012, Oct 15;63(1):47-53.
2 Krämer
M, Herrmann K-H, Schmidt S et al.
Reduction of ringing artifacts for high-resolution 3D RARE imaging at high
magnetic field strength. ISMRM 2017, #5082.
3 Hennig
J, Weigel M, and Scheffler K. Multiecho sequences with variable refocusing flip
angles: Optimization of signal behavior using smooth transitions between pseudo
steady states (TRAPS). Magn. Reson. Med. 2003, 49: 527-535.
4 Hennig
J, Weigel M, and Scheffler K. Calculation of flip angles for echo trains with
predefined amplitudes with the extended phase graph (EPG)‐algorithm: Principles
and applications to hyperechoic and TRAPS sequences. Magn. Reson. Med. 2004, 51:
68-80.
5 Weigel
M and Hennig J. Diffusion sensitivity of turbo spin-echo sequences. Magn. Reson.
Med. 2012, 67: 1528-1537.