Elisa Marchetto1,2, Maxim Zaitsev3, and Daniel Gallichan1,2
1School of Engineering, Cardiff University, Cardiff, United Kingdom, 2Cardiff University Brain Research Centre (CUBRIC), Cardiff University, Cardiff, United Kingdom, 3High Field MR Center, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria
Synopsis
A radial GRE-sequence was modified
using Pulseq to reduce the acoustic noise amount produced by the rapid switching
of gradient currents without excessively increasing the TR-time. Prospective
motion-correction was performed using a markerless tracking device that sent
the motion estimates to the scanner via the libXPACE framework, incorporated
directly into Pulseq, to update gradients and RF pulses during the MR scan.
The quieter sequence resulted in a acoustic noise reduction of 6 dB(A) (i.e. a factor of 2) compared to the
standard one. Image quality was greatly improved using prospective
motion-correction in the case of deliberate motion during the scan.
Introduction
Rapid changes of gradient
currents are the source of magnetic resonance (MR) acoustic noise, because of
the Lorentz forces acting on the gradient coils1. Quiet sequences
can be adopted to reduce the scanning noise by increasing TE and TR2
or by using soft gradient waveforms3 to reduce abrupt changes of the
gradient level.
In this preliminary study, we
adjusted the gradient timings and shape of a radial Gradient Echo (GRE)4
sequence using Pulseq5. Prospective motion correction of voluntary
motion was carried out using a markerless device tracking communicating with
the scanner via the libXPACE (eXternal Prospective Acquisition
CorrEction)6 framework, that was linked directly to a
customized Pulseq interpreter. The aim moving forward is to obtain motion-robust
methods to acquire images with a low acoustic noise level, making them suitable
for a paediatric/infant age group where research scans are typically conducted
while the subject is sleeping. Methods
Pulseq is an open-source
framework for the development and execution of magnetic resonance (MR) pulse
sequences. We used the latest Matlab-based version (1.4.0) to reduce the
acoustic noise generated by a standard 2D radial GRE sequence by allowing the
gradients to ramp-up slower, without increasing the TR excessively. Spoiler
gradients were performed by extending the ramp-down gradient area, while the
slice-selection gradient was ramped more slowly minimising abrupt changes in
gradient amplitude. Care was also taken to align temporally all ramps that
would not affect the image encoding.
Standard and acoustically
optimized sequences were tested on a Prisma scanner (Siemens Healthcare,
Erlangen, Germany) with TR/TE at 18.2/8 ms, FA=20°, slice thickness=3mm and FOV=256mm.
Preliminary scans were performed with a phantom to perform acoustic-noise
measurements using Decibel X app (SkyPaw Co., Ltd) from a mobile phone device
positioned at the entrance of the scanner room. The sequences were tested in
vivo with the participant performing slow back-and-forth head rotation with and
without motion correction enabled. The real-time motion-tracking estimates were
sent to the scanner and coordinate systems updated using the libXPACE framework,
to allow prospective motion correction using a Tracoline (TCL) markerless
device (TracInnovations, Ballerup, Denmark). A separate image was also collected
without deliberate motion as reference.
Image reconstruction was
performed in MATLAB using the Pulseq reconstruction script for non-Cartesian
sampling.Results
A comparison between the standard
and the quieter pulse sequence is shown in Figure 1: spoiler gradient
amplitudes were considerably reduced as well as time-gaps between gradients,
optimized to allow slower gradient rising and falling.
The quiet and standard sequences
resulted in a LMax value of 62.2 db(A) and 68.5 db(A) respectively, with the acoustic
noise measurements shown in Figure 2 for comparison.
Figure 3 compares images acquired
with and without motion correction against the reference images for both
sequences: reasonable image contrast was obtained, and the prospective motion
correction worked well, reducing the motion artifacts occurring during the single
slice acquisition. The motion parameters measured by the tracking device during
the experiments with and without motion correction are shown in Figure 4.Discussion
Slower rising and falling of
gradients as well as optimization of spoiler gradient areas resulted in 6 db(A)
reduction of acoustic noise compared to the standard radial GRE sequence. The acoustic
noise measurement was performed at the entrance of the scanner room for safety
reasons. Future study will require the use of a calibrated MR safe equipment at
the entrance of the scanner bore where the strongest noise can be measured1.
Moreover, further acoustic noise reduction could be achieved by substituting
the traditional trapezoid shape with ‘soft’ gradients, such as sinusoids or
gaussians, which can be trivially achieved with Pulseq.
The motion parameters tracked by
the TCL device show that the motion performed by the subject has comparable
magnitude, although differing in the shape depending at what moment of the back-and-forth
head motion the scan took place. For this proof-of-concept study, the motion
correction was tested successfully for one single slice acquisition. Further
study will investigate the motion correction robustness in the case of
multi-slice acquisitions.
Even in the images where no
motion correction was applied, some fine anatomical features can still be
recognized. This is attributed to the inherent robustness of radial sampling to
motion, where motion artifacts tend to be less severe than for Cartesian
sampling. Nonetheless, prospectively corrected images show substantially higher
image sharpness and the level of detail. Conclusion
In this study, we successfully
reduced the acoustic noise of a standard 2D radial GRE sequence using Pulseq,
maintaining a relatively short TR time. Good image quality could be achieved
using prospective motion correction from the TCL markerless tracking device, integrated
with the scanner through the libXPACE framework. Further studies will
investigate the robustness of sequence and motion correction methods in the case
of multi-slice acquisition in infant subjects, taking advantage of the acoustic
noise reduction especially when no sedation or anaesthesia are used.Acknowledgements
This
work is partly funded by research support from TracInnovations (Ballerup,
Denmark).References
- Price D, De
Wilde J, Papadaki A, et al. Investigation of Acoustic Noise on 15 MRI Scanners
from 0.2 T to 3 T. Journal Of Magnetic Resonance Imaging 2001; 13:288–293.
-
Arthurs 0,
Edwards A, Austin T. The challenges of neonatal magnetic resonance imaging.
Pediatr Radiol 2012; 42:1183–1194.
- Hennel F, Giard
F, Loenneker T. Silent MRI with soft gradient pulses. Mag Res Med 1999;42:6-10.
-
Elster AD.
Gradient echo imaging: techniques and acronyms. Radiology 1993; 186:1-8.
- http://pulseq.github.io/
-
Zaitsev
M, Woletz M, Tik M. Real-time prospective motion correction of arbitrary MR
pulse sequences with XPACE-Pulseq. ISMRM 2021, Abstract #1378.