Jérôme Yerly1,2, Christopher W Roy1, Bastien Milani1, Davide Piccini1,3, Aurélien Bustin1,4,5, Mariana B.L. Falcão1, Ruud B. van Heeswijk1, and Matthias Stuber1,2,4
1Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 2CIBM Center for Biomedical Imaging, Lausanne, Switzerland, 3Advanced Clinical Imaging Technology, Siemens Healthcare, Lausanne, Switzerland, 4Electrophysiology and Heart Modeling Institute, IHU LIRYC, Bordeaux, France, 5Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, Bordeaux, France
Synopsis
The free-running framework (FRF) was recently proposed to address the
limitations of current techniques to assess left ventricular (LV) ejection
fraction (LVEF). However, the accuracy of FRF to assess LVEF has yet to be
quantitatively examined. This work rigorously quantifies and optimizes the
effect of the regularization weights on LVEF and several image quality metrics using
a numerical phantom with well-controlled boundary conditions, and validates the
results in in-vivo 5D FRF data. The results demonstrated that the combination
of regularization weights that are optimal in terms of image quality do not
correspond to the optimal weights for LVEF assessment.
Introduction
Left ventricular (LV) ejection fraction (LVEF) is a strong
predictor of outcome in patients with heart failure.1 Cardiac magnetic resonance (CMR) is the gold-standard
for LVEF assessment and typically involves the acquisition of a stack of LV
short-axis 2D cine-images over multiple breath-holds. However, the low through-plane
spatial resolution and coverage may be suboptimal for accurate assessment of LVEF.2 Moreover, repeated breath-hold acquisitions may lead
to poor reproducibility of end-inspiration position, are often challenging for
patients, and may even cause Valsalva maneuvers resulting in biased LVEF measurements.
In addition, operator involvement remains considerable as several cardiac
localizers are still necessary to find the short-axis view. The free-running
framework (FRF)3 was recently proposed to remove these constraints and
to simplify workflow of LVEF assessment.4 However, the accuracy of FRF to assess LVEF has yet
to be quantitatively examined. To reconstruct the highly undersampled 3D+cardiac+respiratory
motion-resolved (5D) images, FRF requires careful optimization of the
regularization weights to find a compromise between residual aliasing and compression
of the underlying physiological motion. This work rigorously quantifies and
optimizes the effect of the regularization weights on LVEF and several image
quality metrics using a numerical phantom with well-controlled boundary
conditions, and validates the results in in-vivo 5D FRF data.Methods
Numerical
Phantom: Free-running 3D
golden-angle radial data were synthesized using a previously described numerical
simulation.5 The synthesized data simulated anatomical tissues with
realistic nonrigid cardiac and respiratory motion derived from the XCAT phantom6 and included heart-rate and respiratory motion variability. The
simulated LVEF was 63.2%. The n-th radial readout corresponding to the (n-1)*TR timepoint
was obtained by computing the 3D volume representing the desired cardiac and
respiratory phase using the XCAT software, converting the labelled volume to
CMR contrast using a bSSFP signal model with tissue relaxation properties from
the literature, simulating 3D coil sensitivities, and computing the inverse
NUFFT.5
Image
reconstruction: The
synthetic FRF radial data were sorted into a 5D (x-y-z-cardiac-respiratory)
matrix and reconstructed with the compressed sensing (CS) reconstruction
framework shown in Fig.1. The reconstruction implemented total variation
(TV) and local low-rank (LR) regularization terms along both the cardiac and
respiratory dimensions. The four corresponding regularization weights were the
independent variables that were optimized as described below.
Analysis: LVEF was automatically
computed for every reconstructed dataset using thresholding and the ground
truth position of the aortic and mitral valves. The image quality was
quantitatively assessed by the following metrics: mean squared error (MSE),
structural similarity index measure (SSIM), peak signal-to-noise ratio (PSNR),
and blood-myocardium interface sharpness. The four regularization terms were
first optimized individually by setting the other terms to zero. For each term,
the two regularization weights that yielded the highest image quality with an LVEF
error below 3% when compared to ground truth were selected for further
optimization. The 3% limit is well within the range of intra and interobserver
variability of LVEF reported in earlier studies.2 Next, the four regularization terms were optimized
by combining the weights previously selected. The optimal combination of
weights was determined by selecting the combination that resulted in the best
image quality given a maximum acceptable LVEF error of 3%. Finally, as a proof
of concept, the optimized combination was used to reconstruct an in-vivo 5D FRF
dataset4 and the LVEF was compared
to a reconstruction using weights that were optimized for image quality only
without considering the maximum acceptable error limit for LVEF. Results
Increasing the regularization weights generally resulted
in a decrease in LVEF and an improvement in image quality to a point beyond
which image quality degraded again (Fig.2). Regularization along the
respiratory dimension had a significantly lower impact on LVEF and image
quality than regularization along the cardiac dimension, i.e. lower value
spread (Fig.2). High respiratory regularization weights had little impact on LVEF,
but significantly compressed respiratory motion (almost no visible respiratory
motion for $$$\lambda_{TV_r}$$$=0.5 and $$$\lambda_{LR_r}$$$=10 in Fig.3). The optimization of each individual
regularization term resulted in the following weights: $$$\lambda_{TV_c}$$$={0.001,0.005}, $$$\lambda_{TV_r}$$$={0.005,0.01}, $$$\lambda_{LR_c}$$$={0.1,0.5}, and $$$\lambda_{LR_r}$$$={0.5,1.0}. All possible combinations
of the regularization terms with these weights resulted in an LVEF error of
less than 3% (Fig.4). Among all these combinations, the one that gave the best
image quality was ($$$\lambda_{TV_c}$$$,$$$\lambda_{TV_r}$$$,$$$\lambda_{LR_c}$$$,$$$\lambda_{LR_r}$$$)=(0.001,0.01,0.5,0.5). Reconstruction of the in-vivo
FRF dataset using this optimized combination of weights resulted in 50.04% LVEF. When using
the set of weights that maximizes image quality without considering LVEF error,
i.e. ($$$\lambda_{TV_c}$$$,$$$\lambda_{TV_r}$$$,$$$\lambda_{LR_c}$$$,$$$\lambda_{LR_r}$$$)=(0.05,0.01,1.0,1.0), the LVEF was 44.87% in this case.Discussion and Conclusion
This experimental study rigorously quantified and
optimized the effect of the regularization weights of a CS reconstruction on
the LVEF and several image quality metrics using a numerical phantom with
well-controlled boundaries. The results demonstrated that the combination of
regularization weights that are optimal in terms of image quality do not
correspond to the optimal weights for LVEF assessment. Increasing the
regularization weights generally resulted in better image quality, but also decreased
LVEF accuracy due to motion compression artifacts. A relatively wide range of
parameters provided an acceptable tradeoff between LVEF accuracy and image
quality. This study may help guide reconstruction parameters for CS
reconstruction of free-running 5D images and may help bridge a gap in
validating optimized FRF for LVEF assessment. Acknowledgements
No acknowledgement found.References
1. Curtis JP, Sokol SI, Wang Y,
Rathore SS, Ko DT, Jadbabaie F, Portnay EL, Marshalko SJ, Radford MJ, Krumholz
HM. The association of left ventricular ejection fraction, mortality, and cause
of death in stable outpatients with heart failure. J Am Coll Cardiol
2003;42:736–42.
2. Vincenti G, Monney P, Chaptinel
J, Rutz T, Coppo S, Zenge MO, Schmidt M, Nadar MS, Piccini D, Chèvre P, Stuber
M, Schwitter J. Compressed Sensing Single–Breath-Hold CMR for Fast
Quantification of LV Function, Volumes, and Mass. JACC Cardiovasc.
Imaging 2014;7:882–892 doi: 10.1016/j.jcmg.2014.04.016.
3. Sopra LD, Piccini D, Coppo S,
Stuber M, Yerly J. An automated approach to fully self-gated free-running
cardiac and respiratory motion-resolved 5D whole-heart MRI. Magn. Reson. Med.
2019;82:2118–2132 doi: 10.1002/mrm.27898.
4. Yerly J, Di Sopra L, Vincenti G,
Piccini D, Schwitter J, Stuber M. Fully Self-Gated Cardiac and Respiratory
Motion-Resolved 5D MRI for Rapid Assessment of Left Ventricular Function. In:
Montreal; 2019. p. 2106.
5. Roy CW, Heerfordt J, Piccini D,
Rossi G, Pavon AG, Schwitter J, Stuber M. Motion compensated whole-heart coronary
cardiovascular magnetic resonance angiography using focused navigation (fNAV).
J. Cardiovasc. Magn. Reson. 2021;23:33 doi: 10.1186/s12968-021-00717-4.
6. Segars WP, Sturgeon G, Mendonca S, Grimes J, Tsui BMW. 4D XCAT
phantom for multimodality imaging research. Med. Phys. 2010;37:4902–4915 doi:
10.1118/1.3480985.