Alfredo Ordinola1, Walker Jackson2, and Evren Özarslan1
1Department of Biomedical Engineering, Linköping University, Linköping, Sweden, 2Department of Biomedical and Clinical Sciences, Linköping University, Linköping, Sweden
Synopsis
We demonstrate the experimental
determination of the diffusion propagator, indicating a conditional probability density function with two spatial
arguments. To this end, a recently introduced method was implemented on
a benchtop MR scanner and incorporated
into imaging sequences. The data involving two independent wavenumbers were
transformed from the measurement domain to the spatial domain, yielding an
apparent diffusion propagator. Experiments on freely diffusing water provides accurate determination
of the diffusion propagator while apparent propagators measured in
mouse spinal cord reveal significant differences between white and gray matter
regions.
Introduction & Theory
Recently, a novel technique for measuring the
diffusion propagator was introduced1. The employed diffusion encoding gradient waveform (Figure 1a), reduces to the
classic Stejskal-Tanner sequence2 for $$$\mathbf{q’ = -q}$$$ and to Laun et al.’s sequence3 for $$$\mathbf{q’}=0$$$. From
the signal, $$$ E_\Delta(\mathbf{q},\mathbf{q}')$$$, acquired using this
sequence, the propagator can be obtained through the expression1:
$$P(\mathbf{x}',\Delta|\mathbf{x})=\frac{(2\pi)^{-2d}\int\mathrm{d}\mathbf{q}\, e^{i\mathbf{q}\cdot\mathbf{x}}\int\mathrm{d}\mathbf{q}'\, e^{i\mathbf{q}'\cdot\mathbf{x}'}\, E_\Delta(\mathbf{q},\mathbf{q}')}{(2\pi)^{-d} \int\mathrm{d}\mathbf{q}\,e^{i\mathbf{q}\cdot\mathbf{x}}\, E_\Delta(\mathbf{q},\mathbf{0})} \qquad (1)$$
where $$$d$$$ is the number of measurement dimensions.
Compared to the ensemble average propagator available through the inverse
Fourier transform of the Stejskal-Tanner signal, the above quantity has an
extra spatial argument, thus could elucidate more information on the tissue
structure1.
Here, we demonstrate the first experimental
results on biological tissue obtained using this technique.Methods
The pulse sequence presented in Figure 1b was implemented on
a benchtop MRI scanner (Pure Devices GmbH, Germany). In our
implementation, each diffusion gradient of the effective waveform in Figure 1a is replaced with a
bipolar gradient pair to mitigate the effects of concomitant fields,
susceptibility variations, and eddy currents4.
Two samples were investigated: a water phantom
as a proof-of-concept study for free diffusion; and a sample of mouse spinal
cord (obtained following national and local ethical guidelines). This sample had approximately the following dimensions: $$$2\times 1.5\times2.5 mm^3$$$.
For the first sample, images were acquired with
a matrix size of $$$16\times16$$$ over an FOV of $$$10\times10 mm^2$$$. The
values of q and q’ applied in this study ranged from $$$-1.2
rad/\mu m$$$ to $$$+1.2 rad/\mu m$$$ with 21 linearly-spaced steps, yielding a
total of 441 different measurements.
The acquired images of the SC sample had a matrix size of $$$48\times 96$$$ and an in-plane voxel size of approximately $$$0.15\times0.15 mm^2$$$. The values
of q and q’ applied in this study ranged from $$$-0.8 rad/\mu m$$$
to $$$+0.8 rad/\mu m$$$ with 21 linearly-spaced steps, resulting in 441
different measurements.
Diffusion gradients in all experiments were
applied along a single direction, and for the SC sample this direction was set
perpendicular to the direction of the fibers in white matter. Furthermore, this
acquisition was repeated 16 times, averaged, and filtered with a simple
gaussian filter.
Data analysis for the water phantom involved
the reconstruction of the propagator using magnitude-valued data. The
reconstructed propagator was compared to the analytical expression of the
apparent diffusion propagator (accounting for the parameters of the waveform)
and the propagator reconstructed using simulated data obtained by the
analytical signal sampled consistently with the acquisition protocol.
For the second study, complex data acquired
from the scanner were used in the reconstruction pipeline. Furthermore, the real
component of the 2D Fourier transform of the signal was used to obtain a mask for
eliminating the noise-dominated voxels on the $$$xx’$$$ plane.
All computations were performed using MATLAB (Mathworks
Inc., Natick, MA, USA).Results
A $$$5\times5$$$ matrix of the normalized
images acquired from the water phantom is presented in Figure 2. The reconstructed propagator for the
center voxel of the acquired images, and the difference maps obtained with the
analytical apparent propagator and the one reconstructed from simulated data
are presented in Figure 3.
A $$$7\times7$$$ matrix of the normalized
images corresponding to the SC sample is presented in Figure 4. Furthermore, a proton density weighted
image of the spinal cord sample with a higher in-plane resolution is presented
in Figure 5a, and the
real parts of the numerator and denominator of Equation 1, as well as the reconstructed
propagator are shown in Figure
5b for a white matter (WM) and a gray matter (GM) voxel.Discussion & Conclusion
The reconstructed
propagator for free diffusion agrees with the apparent propagator obtained
from the analytical expression for the signal as well as
from simulated data. The peak in the center voxel of the measured
propagator is due to the magnitude-valued data having Rician
distribution.
Compared to free diffusion, the apparent propagator for the spinal cord
sample has limited extent, which is smaller in WM compared to GM. The
pulse sequence reveals the superposition of propagators for each “bouquet”
(containing particles having the same center-of-mass during the
application of the long pulse) translated to a common origin4. Thus, the
observed differences may be due to the presence of restricted
diffusion in WM in the probed orientation (perpendicular to fibers).
Furthermore, the elevated spread of propagator values in GM could be
indicative of its complex architecture.
With the current implementation of our technique, we managed to
successfully reconstruct apparent diffusion propagators for free
diffusion as well as in mouse spinal cord. Due to the limited breadth of
the propagators, the study could be performed with a smaller “FOV” in
the $$$xx’$$$ plane, i.e., with samples in $$$qq’$$$ space separated apart.
Nonetheless, the technique requires many measurements, thus
long acquisition times, even when the gradients are applied in the same
orientation. Therefore, a scheme for reducing the number of required
samples could be tackled in future studies. The feasibility of the
technique on a benchtop MRI scanner as illustrated here is
encouraging for future translation of the method to other platforms
including to clinical settings.Acknowledgements
We are grateful to Deneb Boito and Magnus Herberthson for their feedback on the manuscript, and to Anders Eklund for his support in the acquisition of the MR scanner.References
[1] E. Özarslan,
in Proc
Intl Soc Mag Reson Med, Vol. 29 (2021) p. 3637.
[2] E. O. Stejskal and J. E. Tanner, J Chem Phys 42, 288 (1965).
[3] F. B. Laun, T. A. Kuder, W. Semmler, and B.
Stieltjes, Phys Rev Lett 107,
048102 (2011).
[4] A. M. Ordinola Santisteban and E. Özarslan,
arXiv:2106.16181 (2021).