Julian Rauch1,2, Frederik B. Laun3, Dominik Ludwig1, Maxim Zaitsev4,5, Mark E. Ladd1,2,6, Peter Bachert1,2, and Tristan A. Kuder1
1Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 3Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 4Medical Physics, Department of Radiology, Faculty of Medicine, Medical Center University of Freiburg, Freiburg, Germany, 5High Field Magnetic Resonance Center, Center for Medical Physics and Biomedical Engineering Medical University of Vienna, Vienna, Austria, 6Faculty of Medicine, Heidelberg University, Heidelberg, Germany
Synopsis
Intravoxel dephasing caused by Maxwell or concomitant
fields can lead to image artifacts as signal voids or falsify the quantitation.
In this study, a framework for compensation of concomitant field effects for
double diffusion encoding sequences with single pairs of bipolar gradients on
each axis is presented. Using pulse-width modulated oscillating gradients with
an optimized timing of the oscillating lobes allows for simultaneous diffusion weighting
and compensation of the self-squared terms of the concomitant fields as well as
for the cross terms. Simulations with realistic parameters reveal the
possibility of a signal gain with the proposed method without significant
drawbacks.
Introduction
Double
diffusion encoding (DDE) schemes with two pairs of bipolar gradients separated
by a refocusing pulse can be used to derive orientationally invariant metrics
such as microscopic anisotropy1,2. These sequences are prone to
severe artifacts induced by concomitant fields resulting from Maxwell’s
equations, which introduce an additional non-linear phase contribution on the
transverse magnetization3,4.
Recently, Maxwell field compensation in the context of DDE was demonstrated by
adding oscillating gradients5, which allows compensation of
self-squared concomitant fields.
In this study, we present simulations to investigate the possibility to
additionally compensate the cross terms occurring between the different
physical scanner axes. This is enabled by adding oscillations to all gradient
pulses for simultaneous diffusion weighting and concomitant field compensation
and applying pulse-width modulation (PWM) to control diffusion weighting ($$$\mathrm{q}$$$-
and $$$\mathrm{b}$$$-values).
Especially
local gradient coils with ultra-high amplitudes are expected to create
significant concomitant fields. On the other hand, oscillating gradients are
easier to implement for local gradient coils6,7 due to higher available
slew rates and less peripheral nerve stimulation (PNS) problems.Methods
Simulations
were performed using MATLAB (The Mathworks, Inc, MA). The presented framework
was evaluated on 60 pairs of diffusion wave vectors published by
Jespersen et al.2.
The original diffusion weightings were replaced by oscillating gradients with
pulse-width modulation for reduction of the concomitant phase (Figure 1), which
is given by $$\mathrm{\phi_C(x,y,z)=\gamma\int B_C(x,y,z,t)\,dt.\quad[1]}$$
Here, $$$\gamma$$$ is the gyromagnetic ratio and
$$$\mathrm{B_C(x,y,z,t)}$$$ are the concomitant fields, given to the lowest
order by $$\mathrm{B_C(x,y,z,t)}=\frac{1}{2\mathrm{B_0}}(\mathrm{G_x^2z^2+G_y^2z^2+G_z^2\frac{x^2+y^2}{4}-G_xG_zxz-G_yG_zyz}).\quad[2]$$
The gradients $$$\mathrm{G_x}$$$,$$$\,$$$$$$\mathrm{G_y}$$$ and $$$\mathrm{G_z}$$$
modify the static magnetic field $$$\mathrm{\vec{B}_0=B_0\hat{e}_z}$$$ along
the physical scanner axes $$$\mathrm{x}$$$, $$$\mathrm{y}$$$ and
$$$\mathrm{z}$$$, respectively.
The first three self-squared terms in [2] are compensated when the respective
gradients fulfill
$$\mathrm{\int_0^{TE}s_n(t)\,G_i(t)^2\,dt=0,{\quad}i\in\{x,y,z\}},\quad[3]$$
with $$$\mathrm{TE}$$$ being the echo time and
$$$\mathrm{s_n(t)=(-1)^{n_{rf}(t)}}$$$, where $$$\mathrm{n_{rf}(t)}$$$ is the
number of refocusing pulses within time interval $$$\mathrm{t}$$$ to
$$$\mathrm{TE}$$$.
The long and short duration of a modulated oscillation result from
$$\mathrm{\delta_{l,s}=\delta_{+,-}=\frac{\delta_0}{2N}\left(1\pm\frac{G_{orig}}{G_{PWM}}+\frac{\epsilon}{\delta_0}\right)-\epsilon},\quad[4]$$
which ensures conservation of the $$$\mathrm{q}$$$-value and the total duration
of the original gradient form.
For the given gradient rise time $$$\mathrm{\epsilon}$$$, the duration of the
original diffusion gradient $$$\mathrm{\delta_0}$$$ and the number of its
modulated oscillations $$$\mathrm{N}$$$, the ratio of the gradient amplitudes
$$$\mathrm{G_{orig}}$$$ and $$$\mathrm{G_{PWM}}$$$ of the original and
modulated gradient form, respectively, must fulfill
$$\mathrm{\frac{G_{orig}}{G_{PWM}}<\frac{(1-4N)\epsilon}{\delta_0}+1\quad[5]}$$
to guarantee non-negative flat top times for the oscillation lobes.
On a gradient axis, the gradient form with the highest amplitude undergoes a
gradient-based optimization to determine $$$\mathrm{G_{PWM}}$$$,
$$$\mathrm{\delta_l}$$$ and $$$\mathrm{\delta_s}$$$ with a minimal
$$$\mathrm{b}$$$-value deviation and the constraint in [5]. The optimum for the
same parameters for the lower amplitude gradients is then found with the
objective of concomitant field compensation based on [3].
The modulated lobes on the gradient axes $$$\mathrm{x}$$$ and $$$\mathrm{y}$$$
are cyclically shifted by one oscillation lobe independently until the smallest
impact of the resulting overlap integral on [1] of the last two cross terms in
[2] is reached.
In the simulation, every direction within 60 wave vector pairs2
applied a diffusion weighting of $$$\mathrm{b}=$$$$$$\,$$$500$$$\,$$$s/mm2
with $$$\mathrm{\delta_0=}$$$$$$\,$$$15.96$$$\,$$$ms at a rise time of
$$$\mathrm{\epsilon}=$$$$$$\,$$$0.4$$$\,$$$ms.
The effect of the concomitant phase on the signal was evaluated numerically at
$$$\mathrm{B_0=}$$$$$$\,$$$3$$$\,$$$T without modulation and for
$$$\mathrm{N}=$$$$$$\,$$$4 for a transversal slice with a FOV of 210$$$\,$$$mm
in both directions and a voxel size of
3$$$\,$$$$$$\times$$$$$$\,$$$3$$$\,$$$$$$\times$$$$$$\,$$$3$$$\,$$$mm3.
The slice was shifted from the isocenter by $$$\mathrm{{\Delta}z}=$$$$$$\,$$$5$$$\,$$$cm.Results
The
proposed modulation leads to a gain in the averaged signal of the slice for all
considered pairs of diffusion directions for the applied parameters (Figure 2).
The achieved absolute signal gain of around 6.5$$$\,$$$% for the first
direction pairs decreases due to growing influence of cross terms for the other
direction pairs, but still remains positive due to the shifting of the
modulated gradient lobes where necessary.
Figure 3 shows the effect of shifting the modulated lobes for an exemplary direction
pair. While the self-squared concomitant fields are highly reduced, the cross
term distribution leads to an overall signal loss (a). Shifting the modulated
oscillations on the $$$\mathrm{x}$$$-axis cyclically reduces the effect of the
cross terms and achieves a signal gain in the slice (b).Discussion and Conclusion
As
presented, using oscillating gradients with PWM and adequate temporal shifting enables
concomitant field compensation with simultaneous diffusion weighting for arbitrary
pairs of encoding wave vectors.
For the considered gradient direction pairs, modulation with four oscillations
per gradient is expected to achieve signal improvement on average without exceeding
a gradient amplitude of 80$$$\,$$$mT/m, which makes the method theoretically
feasible on commercially available gradient systems.
To maintain this amplitude limit and comply with [5], the duration of the
original diffusion gradients was increased by 4.37$$$\,$$$ms. The modulation
leads to a maximum deviation in $$$\mathrm{b}$$$-value of 70$$$\,$$$s/mm2,
which can be reduced by using more oscillations setting higher requirements on
the rise time and amplitude limit.
The effective diffusion time will not be substantially affected by the
modulation due to two different durations within one oscillation period.
On whole-body systems, PNS may be a major issue for oscillating gradients,
which will be less limiting for local gradient coils6,7.
In conclusion, the proposed PWM can lead to a signal gain by reducing the
influence of concomitant fields without the need for additional refocusing RF
pulses.Acknowledgements
No acknowledgement found.References
1. Lawrenz M, Finsterbusch J. Double-wave-vector diffusion-weighted imaging
reveals microscopic diffusion anisotropy in the living human brain. Magn Reson
Med. 2013;69(4):1072-82
2. Jespersen SN et al. Orientationally
invariant metrics of apparent compartment eccentricity from double pulsed field
gradient diffusion experiments. NMR Biomed. 2013;26(12):1647-62
3. Bernstein MA et al.
Concomitant gradient terms in phase contrast MR: analysis and correction. Magn
Reson Med. 1998;39(2):300-8
4. Zhou XJ, Tan SG, Bernstein MA. Artifacts Induced by
Concomitant Magnetic Field in Fast Spin Echo Imaging. Magn Reson Med. 1998;40(4):582-91
5. Rauch et al.
Concomitant field compensation using additional oscillating gradients in a
double diffusion encoding
imaging sequence. Ann. Meet. Int. Soc.
Magn. Reson. Med. 2021; In proc.,#0668
6. Weiger M, Overweg J, Rösler MB,
Froidevaux R, Hennel F, Wilm BJ, Penn A, Sturzenegger U, Schuth W, Mathlener M,
Borgo M, Börnert P, Leussler C, Luechinger R, Dietrich BE, Reber J, Brunner DO,
Schmid T, Vionnet L, Pruessmann KP. A high-performance gradient insert for
rapid and short-T2 imaging at full duty cycle. Magn Reson Med. 2018;79(6):3256-3266
7. Jia F, Littin S, Amrein P, Yu H, Magill AW, Kuder TA, Bickelhaupt S, Laun
F, Ladd ME, Zaitsev M. Design of a high-performance non-linear gradient coil
for diffusion weighted MRI of the breast, J. Magn. Reson. 2021;331:107052.