Hyungseok Jang1, Yajun Ma1, Dina Moazamian1, Michael Carl2, Saeed Jerban1, Alecio F Lombardi1, Christine B Chung1,3, Eric Y Chang1,3, and Jiang Du1
1Radiology, University of California San Diego, San Diego, CA, United States, 2GE Healthcare, San Diego, CA, United States, 3Radiology Service, Veterans Affairs San Diego Healthcare System, San Diego, CA, United States
Synopsis
T1ρ has been
investigated as a quantitative biomarker sensitive to changes in macromolecules
such as proteoglycan and collagen in musculoskeletal systems. More recently,
adiabatic T1ρ (Adiab-T1ρ) has
emerged as an alternative to conventional continuous wave T1ρ to
reduce the magic angle effect, which is a major confounding factor in accurate
T1ρ estimation. In this study, we
investigated a new pulse sequence combining adiabatic T1ρ preparation and efficient 3D fast spin echo (FSE) for more robust adiab-T1ρ
mapping in the human knee. The efficacies of RF cycling and magnetization reset
were also demonstrated.
Introduction
Spin lattice relaxation in
the rotating frame (T1ρ) is a promising biomarker for early detection
of osteoarthritis (OA), as it is sensitive to changes in proteoglycan (PG)
and collagen in the musculoskeletal system 1–3. Conventional
continuous wave-based T1ρ (CW-T1ρ) imaging is disadvantaged by a strong magic
angle effect, with artificially increased CW-T1ρ values when fibers are reoriented from 0° to ~55° relative to the B0 direction
4, thus impairing reproducibility and diagnostic power. Recently, adiabatic
full passage (AFP) pulses were incorporated into T1ρ imaging to ameliorate the magic angle
effect for more robust detection of early cartilage degeneration 5-8.
In this study, we investigated the feasibility of 3D adiabatic T1ρ-prepared fast spin echo (3D Adiab-T1ρ-FSE) imaging. Methods
Figure 1A shows the modules of the 3D Adiab-T1ρ-FSE sequence. First, a magnetization
reset (MagReset) module is used (Figure 1B) to spoil longitudinal magnetization
and thus force the steady state signal to depend only on the fixed saturation
recovery time. This was shown to be effective in T1ρ imaging where variable spin-locking
times (TSLs) may cause different degrees of T1 recovery leading to
inconsistent signal 9. Figure 1C shows spin-locking preparation
utilizing an AFP pulse train (duration=6.048 ms, bandwidth=1.643 kHz, and maximum B1 amplitude=17
mT) followed by RF cycling (RC) using two adiabatic half pulses (AHPs). RC is
known to effectively remove T1 contamination and thus improve signal
fitting 9, which requires two acquisitions with positive and
negative RF phases. For imaging, a variable flip angle (VFA)-based 3D FSE
readout scheme was utilized 10.
The 3D Adiab-T1ρ-FSE sequence was implemented on a clinical
3T MR750 scanner (GE Healthcare). To evaluate the proposed sequence, a phantom experiment
comprised of two sets of three tubes with 2%, 3%, and 4% agarose, ex vivo imaging
with three cadaveric knee joints, and in vivo imaging with five healthy
volunteers were performed with an 8-channel transmit/receive knee coil using the
imaging parameters shown in Figure 1D.
All images were
reconstructed online and processed with Matlab. To yield Adiab-T1ρ parameters, non-linear least squares
curve fitting was performed based on a mono-exponential decay model with two
free parameters to fit (M0 and T1ρ) using either mean signal in a region
of interest (ROI) or pixelwise data. Results
Figure 2A shows T1ρ-weighted images with five different
TSLs from the agarose
phantom. For all tubes, monotonical decay was observed with increased TSLs.
Figures 2B and 2C show the average signal and the fitted curve with tubes 1-3
and 4-6, as well as linear correlation between the measured R1ρ (i.e., 1/T1ρ) and agarose concentrations. As
expected, tubes with higher agarose concentrations exhibited shorter T1ρ (i.e., higher R1ρ). The measured R1ρ showed high linearity with the agarose
concentrations (R2>0.99). The top sets of tubes (tubes 1-3)
showed slightly lower T1ρ than the bottom sets (tubes 4-6), presumably
due to the spatially varying B1 field causing different T1ρ dispersions.
Figure 3 shows images from a
cadaveric knee joint scanned with the 3D Adiab-T1ρ-FSE sequence with or without RC and MagReset.
Without RC, strong signal decay (blue arrows) and/or rebounded signal (red
arrows) were observed between images with TSL=0 and 24ms (Figure 3A), which were
ameliorated by RC (Figure 3B). No dramatic morphological difference was
observed between images with or without MagReset (Figure 3B or 3C).
The signal decay curves from an ROI within femoral
cartilage (Figure 4A) are shown in Figures 4B-D. As observed in Figure 3, the T1ρ signal dropped significantly between TSLs of 0 and
24ms, spoiling curve fitting (black dotted line) resulting in too short a T1ρ parameter. With RC, the signal decay clearly exhibited mono-exponential
decay (red dotted line) and the fitted T1ρ parameter was within the expected range. Figures 4C-D
show T1ρ signal curves obtained with three different TRs (1.7,
3, and 6sec) where the signal curve with TR of 6sec was considered as a
reference with negligible T1 effect. Without MagReset, the signal decay
depended on TR due to the unwanted variable T1 weighting (Figure 4C),
whereas the presence of MagReset dramatically removed the dependence on TR (Figure 4D).
Figure 5A shows T1ρ mapping of a representative healthy volunteer.
All four ROIs showed estimated T1ρ values within the expected range 5-7.
Figure 5B shows complete values for all subjects. Discussion and Conclusion
In this study, we demonstrated feasibility of 3D Adiab-T1ρ-FSE in the human knee. The combination
of 3D FSE and adiabatic-T1ρ preparation is expected to reduce the magic
angle effect and susceptibility, likely providing a novel effective quantitative
imaging tool in musculoskeletal applications. The spatially varying signal dropout
and rebound shown in Figures 3 and 4 are presumably due to FSE readout
contamination by the simulated echoes, more obvious in vivo due to stronger B1
inhomogeneity. We showed the efficacy of RC in removing such signal variation. We
will investigate further to clarify the source of this unwanted signal change. MagReset
was also effective in removing TR dependency, which is beneficial because it
removes the need for additional T1 mapping to compensate for different
T1 recovery times. In future studies, we will investigate these
effects in depth, optimize imaging protocols, and evaluate 3D Adiab-T1ρ-FSE in a group of patients. Acknowledgements
The authors
acknowledge grant support from the NIH (R01AR062581, R01AR068987, R01AR075825, R01AR078877,
and R21AR075851), Veterans Affairs (I01RX002604 and I01CX001388), and GE
Healthcare.References
1. Duvvuri U, Charagundla SR, Kudchodkar SB, et al. Human
Knee: In Vivo T1ρ -weighted MR Imaging at 1.5 T—Preliminary Experience.
Radiology 2001;220:822–826.
2. Regatte RR, Akella SVS, Wheaton AJ, et al.
3D-T1ρ-relaxation mapping of articular cartilage: In vivo assessment of early
degenerative changes in symptomatic osteoarthritic subjects. Acad. Radiol.
2004;11:741–749.
3. Akella S V, Regatte RR, Gougoutas a J, et al. Proteoglycan-induced changes in
T1rho-relaxation of articular cartilage at 4T. Magn. Reson. Med.
2001;46:419–423.
4. Shao H, Pauli C, Li S, Ma Y, Tadros AS, Kavanaugh A, et
al. Magic angle effect plays a major role in both T1rho and T2 relaxation in
articular cartilage. Osteoarthritis Cartilage 2017; 25:2022-2030.
5. Rautiainen J, Nissi MJ, Liimatainen T, Herzog W,
Korhonen RK, Nieminen MT. Adiabatic rotating frame relaxation of MRI reveals
early cartilage degeneration in a rabbit model of anterior cruciate ligament
transection. Osteoarthritis and Cartilage 2014; 22:1444-1452.
6. Rautiainen J, Nissi MJ, Salo E, Tiitu V, et al.
Multiparametric MRI assessment of human articular cartilage degeneration:
correlation with quantitative histology and mechanical properties. Magn Reason
Med 2015; 74:249-259.
7. Casula V, Nissi MJ, Podlipska J, et al. Elevated
adiabatic T1ρ and T2ρ in articular cartilage are associated with cartilage and
bone lesions in early osteoarthritis: a preliminary study. J Magn Reson Imaging
2017; 46:678-689.
8. Hänninen N, Rautiainen J, Rieppo L, Saarakkala S, Nissi
MJ. Orientation anisotropy of quantitative MRI relaxation parameters in ordered
tissue. Sci Rep 2017; 7:9606.
9. Li X, Han ET, Busse RF, Majumdar S. In vivo T1ρ mapping
in cartilage using 3D magnetization-prepared angle-modulated partitionedk-space
spoiled gradient echo snapshots (3D MAPSS). Magn. Reson. Med. 2008;59:298–307.
10. Busse RF, Brau ACS, Vu A, et al. Effects of refocusing
flip angle modulation and view ordering in 3D fast spin echo. Magn. Reson. Med.
2008;60:640–649.