Fischer Johannes1 and Michael Bock1
1Dept. of Radiology, Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, Uni, Freiburg, Germany
Synopsis
Imaging sequences with sub-millisecond temporal resolution based on single point imaging (SPI) are very time inefficient. We designed and imaged a motion phantom capable of velocities up to 5.5m/s. A synchronization signal is acquired using a photoelectric barrier and combined with a sequence trigger signal. Reconstructed images show clear delineation of small structures with a spatial resolution of $$$\Delta x$$$=0.67mm and a temporal resolution of $$$\delta t$$$=246μs.
Introduction
Recently, MRI sequences have been developed
which can visualize rapid 2D motion with sub-millisecond temporal resolution,
e.g. of the vocal folds [1], heart valves [2], or electrical signals [3]. Our implementation, the SPIRE
sequence [1], uses single point imaging (SPI)
leading to long measurement times, so that volunteer measurements can become
challenging. To assess the precision of SPIRE, a motion phantom is required
that offers motion velocities up to 5m/s as can be found in the vocal folds [4]. Here, we present a custom-built, MR-compatible
rotation phantom to demonstrate the imaging capabilities of SPIRE.Methods and Materials
To create a repeatable angular motion, a rotation phantom was
constructed with small tubes of equal mass $$$m$$$ (V=0.5 mL, ∅o=7.5 mm) that
are placed at different radii $$$r_i$$$ and that can be filled with contrast
media solutions. The placement at different $$$r_i$$$ allows for simultaneous
measurements at different velocities and avoids signal ambiguities as the
samples move along different trajectories for each rotation. To reduce
vibrations at high angular frequencies, the mass distribution in the phantom was
balanced such that the center of mass coincides with the rotation axis:
$$\vec{CM}=\sum_i^N\frac{m\vec{r_i}}{M}=\frac{1}{N}\sum_i^N\vec{r_i}=\frac{1}{N}\sum_i^N
r_i\left(\begin{array}{c} \cos(\alpha_i) \\ \sin(\alpha_i)\end{array}\right)=\left(\begin{array}{c}
0 \\ 0\end{array}\right)$$
For $$$N$$$=3 tubes and a constant distance
$$$\Delta r$$$ between the radii this yields the conditions
$$\frac{r_1+r_3}{2}=r_2$$
and (with $$$\alpha_1$$$=0°)
$$\cos{\alpha_2} = \frac{3(r_1+r_3)}{4r_1}-2$$
$$\cos{\alpha_3} =
1-\frac{3(r_1+r_3)^2}{8r_1r_3}.$$
With $$$r_1$$$=22mm, $$$r_2$$$=31mm
and $$$r_3$$$=40 mm ($$$\Delta r$$$=9 mm) the radial separation between the samples is
1.5 mm and the angles are $$$\alpha_2$$$=83.5° and $$$\alpha_3$$$=129.6°. A schematic of the phantom can be seen in figure 1.
To
rotate the disk with the samples, compressed air is blown into the blades of
the phantom. As the SPIRE image reconstruction requires information about the
phantom’s rotation angle, 4 shutters are attached below the disk that
periodically close a photoelectric-barrier. The main body of the phantom (Fig.
2) was 3D-printed in polylactic acid (PLA) on a fused deposition modelling printer
(Prusa i3 Mk3S, Prusa Research) and is held in place in a frame made from plastic
sheets. The rotation axis is attached with ball bearings made of polyetheretherketon
(PEEK), as ceramic bearings can create image artefacts. The photoelectric
barrier is powered by a non-magnetic 9V battery inside the bore. The electric signal
from the barrier is transported via a coaxial cable which was equipped with baluns
to reduce RF coupling. It is recorded together with a synchronization signal
from the MR sequence that is created during the RF pulse with a stand-alone PC running
audio-recording software (Audacity, Audacity Team). In the reconstruction, time
delays between signal edges are detected to determine the phase of the rotation
in each TR, as well as the rotation frequency. The
material in the sample tubes comprises distilled water and contrast agent
(Prohance) to obtain $$$T_1\approx$$$100ms. Agar agar is added to solidify the
mixture in order to prevent artefacts from turbulence. Bacterial growth is
prohibited by the addition of ethanol (Figure 1b). To allow for the analysis of
higher spatial resolution, small inserts can be placed in the tubes which have
a thickness of 1.5 mm (Figure 1c). The inserts were printed with photoreactive
resin using masked stereolithography (Prusa SL1, Prusa Research).
As
the largest radius defines the necessary imaging FOV and, thus, the duration of
the measurement the outer sample tubes can be replaced by MR-invisible samples to
reduce measurement time. These samples were prepared with a negative contrast
agent (Resovist) which reduced $$${T_2}^*<$$$1 ms (Figure 1c). SPIRE images
with a maximum phase encoding duration of 800μs were acquired
of one visible high-resolution sample at $$$r_1$$$=22mm rotating at a target
frequency of 40 Hz ($$$v_{tangential}$$$=5.5 m/s). Other imaging parameters
were: FOV=60x60mm2, $$$\Delta x$$$=0.667mm, TE/TR=1.15/2.96ms, a=8°,
TA=40min). Using the
synchronization signals, the SPIRE data was sorted into 100 time frames (i.e.,
3.6° angular resolution) and reconstructed using a total variation constraint
along the temporal dimension [5].Results
In
Figure 3 exemplary signals from both the MR sequence and the optical are shown. In this experiment
the rotation frequency was very stable with $$$f$$$ = 40.7($$$\pm$$$0.3) Hz,
exhibiting only a small drift over the total measurement time which might be
attributed to small variations in the supplied air pressure. For 100
reconstructed frames this results in a temporal resolution of 246$$$\pm$$$2μs and the images clearly show the motion of the
phantom as well as internal structures. With a radius of 22mm the tangential
motion velocity was $$$v$$$=5.6 m/s.Discussion
During measurement
setup, the rotational center can be difficult to identify and may lead to
off-center FOV which has to be corrected in the reconstruction. This could be
improved with an MR visible filling of the axis. Although the longest phase
encoding duration in the SPIRE acquisition was more than 3x longer than the
reconstructed resolution, even small internal structures of the sample tubes
could be well resolved – this result shows that SPIRE can image very rapid
motion reconstructed at smaller temporal resolution than the duration of phase encoding
gradients.Acknowledgements
References
[1] J.
Fischer et al., “Sub-millisecond 2D MRI of the vocal fold oscillation
using single-point imaging with rapid encoding,” Magn Reson Mater Phy,
Sep. 2021, doi: 10.1007/s10334-021-00959-4.
[2] Z. Zhong et al., “Visualization
of Human Aortic Valve Dynamics Using Magnetic Resonance Imaging with
Sub-Millisecond Temporal Resolution,” Journal of Magnetic Resonance Imaging,
vol. 54, no. 4, pp. 1246–1254, 2021, doi: 10.1002/jmri.27603.
[3] C. J. G. Bakker et al., “Multiple
single-point imaging (mSPI) as a tool for capturing and characterizing MR
signals and repetitive signal disturbances with high temporal resolution: The
MRI scanner as a high-speed camera,” Magnetic Resonance Imaging, vol.
31, no. 7, pp. 1037–1043, Sep. 2013, doi: 10.1016/j.mri.2013.04.014.
[4] Y. W. Shau, C. L. Wang, F. J. Hsieh,
and T. Y. Hsiao, “Noninvasive assessment of vocal fold mucosal wave velocity
using color doppler imaging,” Ultrasound Med Biol, vol. 27, no. 11, Art.
no. 11, Nov. 2001, doi: 10.1016/s0301-5629(01)00453-7.
[5] M. Uecker et al., “Berkeley
Advanced Reconstruction Toolbox,” in In Proc. Intl. Soc. Mag. Reson. Med., Toronto,
2015, vol. 23, p. 2486.