Harsh Kumar Agarwal1, Shaik Ahmed1, Rajagopalan Sundaresan1, Sudhanya Chatterjee1, Sajith Rajamani1, Ashok Kumar P Reddy1, Bhairav Mehta1, Dattesh Shanbhag1, and Ramesh Venkatesan1
1GE Healthcare, Bangalore, India
Synopsis
MR imaging is signal starve leading
to long acquisition time. Multiple averaging/excitation is most common way to
boost the SNR. However, it is associated with proportionately long scan time.
This abstract presents Variable Compress Sensed Excitation (vNEX) image acquisition
and image reconstruction technique which subsample each signal average by compressed
sensing fast MRI technique and robust image reconstruction is done to
reconstruct MRI images for individual signal averages. The proposed technique is
demonstrated for T2w brain MRI where SNR equivalence of 336sec scan time is
shown in a 200sec vNEX MRI. Three sampling patterns were proposed and
statistically compared.
PURPOSE
To improve the SNR of the MRI while reducing the
scan time associated with multiple excitations/averaging.INTRODUCTION
Signal to noise ratio (SNR) is an important
indicator of image quality of the MR. The most common way to improve SNR is to
increase the number of excitations/averaging (NEX) which leads to proportionate
increase in scan times with increased motion artifacts [1, 2]. In this
abstract, the increased scan time is mitigated by reducing the scan time of each
excitation with the use of compressed sensing [3]. The combined k-space
sampling across the averages has variable k-space averaging which is closer to
SNR optimal k-space averaging with variable k-space averaging mimicking the
power distribution of image in k-space [4]. The compressed sensing reconstruction
further improves the SNR of individual images prior to averaging, leading to an
improvement in the final image SNR [3]METHODS
The
proposed technique composed of novel data acquisition strategy and novel image
reconstruction workflow.
Data
Sampling: The proposed technique acquires different
k-space locations for different signal excitations. Three different sampling
schemes were tested in this abstract.
- Random vNEX (r-vNEX): each signal excitation
is independently and uniformly randomly subsampled.
- Complementary vNEX (c-vNEX): each
signal excitation is independently and uniformly subsampled such that each
k-space location is acquired at-least once across all the excitations.
- Variable density NEX (vd-vNEX): A sampling
density across the phase encode direction is defined, and each signal excitation
is independently and subsampled as per the prescribed sampling density.
Figure 1
shows the sampling of each of the sampling pattern for each of the sampling
pattern for NEX=4 and compressed sensing factor of 1.5. The stepped sampling density function is shown in Figure 1 for
vd-vNEX sampling pattern.
Image Reconstruction:
Compressed sensing reconstruction was done for each individual excitation [3].
The combined k-space averaged image across all the excitations was used as
initialization for compressed sensing algorithm to improve the robustness to artifacts
and poor SNR of a signal excitation data.
Volunteer
scanning: A Healthy subject was scanned using the IRB
approved protocol with informed consent on a commercially released 1.5T MRI
scanner. The commercial fast spin echo sequence with
multiple excitations and compressed sensing acceleration was modified for the
proposed technique. A fully sampled T2w MRI with NEX=2 and three prospectively
subsampled k-space as per the three sampling patterns, as shown in Figure 1 is
acquired with NEX=4 and compressed sensing factor of 1.5.
Atlas (ICBM 2009a Nonlinear Symmetric [5]) based
registration was done to create the mask for the brain in the MRI volume [6]. The brain signal, noise (from subtraction of first two signal averages) and
SSIM is computed over the masked brain to improve the clinical relevance of
these number. SSIM of vNEX images were computed with fully sampled T2w MRI for
each slice and Student’s paired t-test with most aggressive Bonferroni correction.
p-value <0.01 is used for statistical significance.
RESULTS
The MRI images from individual excitations for c-vNEX is shown in Figure 2.
Figure 3 shows the final reconstructed MR images from fully sampled and three
vNEX subsampling schemes. Table 1 shows the quantitative comparison of the three
vNEX sampling schemes with the fully sampled data acquisition. SSIM of three
vNEX subsampling is statistically different from each other with highest SSIM
for c-vNEX subsampling pattern.DISCUSSION
The proposed technique reduced the scan time of the NEX=4 MRI from 336sec to
200sec with compressed sensing (R-1.5) subsampling of individual image
excitations. The SNR provided by the proposed vNEX technique was equivalent to
the NEX=4 fully sampled MRI, as seen from the extension of the observed scan
time and SNR numbers of fully sampled (NEX=2) and vNEX subsampled MRI in Table
1.
The
usage of the combined k-space as initialization could have improved the robustness
of the low SNR individual NEX image reconstruction. Qualitatively, Individual excitation
MRI images were noisy, and their SNR improved by signal averaging. The three vNEX
sampling patterns accessed in this abstract had similar SNR, however c-vVEX had
statistically significant better SSIM over brain compared to other sampling
patterns. CONCLUSION
This abstract presents a novel method to boost the SNR of the MRI image with signal averaging without the time penalty associated with the multiple averaging. The reduction in scan time from 336sec (for NEX-=4) to 200sec for T2w Brian MRI is shown in a volunteer study. Complementary-vNEX sampling pattern have shown better performance compared to other sampling patterns. However, further study over larger patient population and multiple contrast types is warranted to assess the clinical significance of the proposed technique. Acknowledgements
No acknowledgement found.References
[1] Bernstein, M.A., King, K.F. and
Zhou, X.J., 2004. Handbook of MRI pulse sequences. Elsevier.
[2] Brown, R.W., Cheng, Y.C.N.,
Haacke, E.M., Thompson, M.R. and Venkatesan, R., 2014. Magnetic resonance
imaging: physical principles and sequence design. John Wiley & Sons.
[3] Lustig, M., Donoho, D.L.,
Santos, J.M. and Pauly, J.M., 2008. Compressed sensing MRI. IEEE signal
processing magazine, 25(2), pp.72-82.
[4] Zijlstra, F., Viergever, M.A.
and Seevinck, P.R., 2016. Evaluation of variable density and data-driven
k-space undersampling for compressed sensing magnetic resonance imaging.
Investigative radiology, 51(6), pp.410-419.
[5] Fonov, V., Evans, A.C.,
Botteron, K., Almli, C.R., McKinstry, R.C., Collins, D.L. and Brain Development
Cooperative Group, 2011. Unbiased average age-appropriate atlases for pediatric
studies. Neuroimage, 54(1), pp.313-327.
[6] Avants, B.B., Tustison,
N. and Song, G., 2009. Advanced normalization tools (ANTS). Insight j, 2(365),
pp.1-35