Thomas Wilhelm Eigentler1,2, Bilguuun Nurzed1, Andre Kuehne3, and Thoralf Niendorf1,3,4
1Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrück Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany, 2Technische Universität Berlin, Chair of Medical Engineering, Berlin, Germany, 3MRI.TOOLS GmbH, Berlin, Germany, 4Experimental and Clinical Research Center (ECRC), a joint cooperation between the Charité Medical Faculty and the Max Delbrück Center for Molecular Medicine, Berlin, Germany
Synopsis
The
gain in signal-to-noise ratio is driving the ultrahigh field and extreme field
MRI. This work highlights the opportunities and electromagnetic constraints of realistic
radiofrequency array concepts tailored for cardiovascular MRI at 7.0T, 10.5T,
and 14.0T. For this purpose electromagnetic field simulations were performed
using 32, 48 and 64-channel RF transceiver array configurations at 7.0T, 10.5T,
and 14.0T. Our findings demonstrate, that a higher channel count increases the
degrees of freedom of B1+ shaping. Furthermore, the
preliminary results provide a strong mandate for static PTX or even dynamic PTX
tailored at the heart at 10.5T and 14.0T.
Introduction
The
gain in signal-to-noise ratio (SNR) is a major driving force for research into
ultrahigh field (UHF) MR.1,2 Progress in RF-coil design and
advancement of imaging techniques facilitated a growing portfolio of clinically
relevant UHF-MR applications. The success at 7.0 T MR provides convincing evidence for even
higher magnet field strengths. The field has already taken further steps into
the future to push the magnetic field strength boundaries into extreme field
(EF) MR.3 This envisions human MR at 14.0 T and at
20.0 T4,5, and is an important conceptual leap. Unlocking the potential of EF-MR requires
unravelling and leveraging the electrodynamics of the short wavelength regime.
Responding to the challenges and recognizing the opportunities, this study elucidates
the electrodynamic constraints at high spin excitation frequencies and explores
multi-element RF transceiver
array concepts tailored for human cardiovascular MR (CMR) at 10.5 T
and 14.0 T. To increase the degree of freedom for
transmission field shaping
the channel count is increased for RF arrays of Self-Grounded Bow-Tie (SGBT) antenna building
blocks. Electromagnetic field (EMF) simulations are performed at 10.5 T
and at 14.0 T and benchmarked against the 7.0 T reference.6Methods
The
transceiver SGBT antenna building block was scaled to the wavelength at 10.5 T,
and 14.0 T using the 7.0 T configuration as a starting point.6 To investigate the impact of the channel count 32-channel
and 48-channel configurations were examined at 10.5 T. At 14.0 T 32-,
48- and 64-channel configurations were designed and investigated (Figure 1). EMF
simulations were performed with CST Microwave Studio (CST Studio Suite 2020,
Dassault Systèmes, Vélizy-Villacoublay Cedex, France) using the human voxel
model Duke.7 Post processing was performed in Matlab 2019b (MathWorks,
Natick, MA) including tuning and matching, channel wise B1+
calculation, and SAR10g calculation. For B1+ shimming
a genetic algorithm (GA) as well as a multi-objective GA of the Matlab global
optimization toolbox was used. To benchmark the proposed array configurations
shim-independent measures were calculated including: intrinsic SNR (ISNR) and
intrinsic transmit efficiency (ITXE), as well as geometry-factor (g-factor) for sensitivity encoding
(SENSE) parallel imaging.Results
Figure
2 shows B1+ distribution deduced for GA optimization of min(B1+)
using phase shimming (CTX). Figure 3 shows the same optimization but with static
parallel transmit (PTX). PTX at 14.0 T yielded an increased min(B1+)
for the 32-channel and the 64-channel configurations. The other configurations
didn’t show a significant min(B1+) improvement, but mean(B1+)
was improved for all configurations. The multi-objective optimizer helped to reduce SAR
by 39% with only 7% reduction in min(B1+) for the 32-channel
7.0 T configuration (Figure 4). The 48-channel (10.5 T) and the 64-channel
(14.0 T) configurations yielded a SAR reduction of 64% and 84%
respectively with ~9% loss in min(B1+) (Figure
4). Figure 5a highlights the results obtained for ISNR; ITXE revealed
comparable results and are not shown in this abstract. The ISNR gain combined
with the signal gain of B0² in NMR relative to 7.0 T are shown
in Figure 5b.8 With a constant channel count of 32-channels the mean SNR gain
was found to be 2.07 for 10.5 T and 3.30 for 14.0 T for the heart.
Increasing the channel count to 64 for 14.0 T resulted in an SNR gain of
4.35 ± 1.09.
For the 64-channel configuration at 14.0 T the
max(g-factor) = 1.72 for R=6 was shown (R-L phase encoding). This
value is comparable to the 32-channel configuration at 7.0 T at R=4 with
max(g-factor) of 1.67.Discussion and Conclusion
Our EMF simulations show that the compact SGBT
building block permits high density and high channel count RF arrays that
provide RF penetration and transmission field uniformity suitable for CMRI at
14.0 T. Beyond the channel count our findings demonstrate that increasing the
degree of freedom by using static PTX instead of CTX benefits B1+
shimming. Our observations revealed that moving to EF increases SAR10g.
This SAR10g increase can be offset using multi objective
optimization GA for B1+ shimming. For ISNR and ITXE the
theoretically achievable values could be increased at 10.5 and 14.0 T with a
higher channel count facilitating a gain in SNR and TXE. The high channel count
enhances the parallel imaging performance. To facilitate high density RF arrays
a SGBT antenna building block was used in this work but the RF
array concepts can be adapted and expanded to RF array configurations using
other building blocks such as loops or hybrids of loops and dipole antenna.
To conclude, a higher channel count increases
the degrees of freedom of B1+ shaping tailored for CMR at
extreme fields. Our preliminary results underline the benefit of more
sophisticated transmission field shimming versus conventional CTX. This provides
a strong mandate for static PTX or even dynamic PTX at 10.5 T and 14.0 T. These findings are heartening and
provide the technical foundation for the development of RF array technology
dedicated to cardiac MRI at 14.0 T.Acknowledgements
This project has received funding from the
European Research Council (ERC) under the European Union's Horizon 2020
research and innovation program under grant agreement No 743077 (ThermalMR).References
1. Vaughan JT, Garwood M, Collins CM, et
al. 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head
images. Magn Reson Med. 2001;46(1):24-30. doi:10.1002/mrm.1156
2. Pfrommer A, Henning A. The ultimate
intrinsic signal‐to‐noise ratio of loop‐ and dipole‐like current patterns in a
realistic human head model. Magn Reson Med. 2018;80(5):2122-2138.
doi:10.1002/mrm.27169
3. Niendorf T, Barth M, Kober F, Trattnig
S. From ultrahigh to extreme field magnetic resonance: where physics, biology
and medicine meet. Magn Reson Mater Physics, Biol Med.
2016;29(3):309-311. doi:10.1007/s10334-016-0564-1
4. Winter L, Niendorf T. Electrodynamics
and radiofrequency antenna concepts for human magnetic resonance at 23.5 T (1
GHz) and beyond. Magn Reson Mater Physics, Biol Med. 2016;29(3):641-656.
doi:10.1007/s10334-016-0559-y
5. Budinger TF, Bird MD, Frydman L, et al.
Toward 20 T magnetic resonance for human brain studies: opportunities for
discovery and neuroscience rationale. Magn Reson Mater Physics, Biol Med.
2016;29(3):617-639. doi:10.1007/s10334-016-0561-4
6. Eigentler TW, Kuehne A, Boehmert L, et
al. 32‐Channel self‐grounded bow‐tie transceiver array for cardiac MR at 7.0T. Magn
Reson Med. June 2021:mrm.28885. doi:10.1002/mrm.28885
7. Christ A, Kainz W, Hahn EG, et al. The
Virtual Family—development of surface-based anatomical models of two adults and
two children for dosimetric simulations. Phys Med Biol.
2010;55(2):N23-N38. doi:10.1088/0031-9155/55/2/N01
8. Edelstein WA, Glover GH, Hardy CJ,
Redington RW. The intrinsic signal-to-noise ratio in NMR imaging. Magn Reson
Med. 1986;3(4):604-618. doi:10.1002/mrm.1910030413