Ludovica Romanin1,2, Christopher W. Roy2, Milan Prsa3, Tobias Rutz4, Estelle Tenisch2, Matthias Stuber2,5, and Davide Piccini1,2
1Advanced Clinical Imaging Technology, Siemens Healthcare AG, Lausanne, Switzerland, 2Department of Diagnostic and Interventional Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 3Division of Pediatric Cardiology, Woman-Mother-Child Department, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 4Service of Cardiology, Heart and Vessel Department, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 5Center for Biomedical Imaging (CIBM), Lausanne, Switzerland
Synopsis
This work proposes an automated subject-specific individual pixels selection as an alternative to a fixed coil selection for the initialization of the input data to a similarity-driven multi-dimensional binning algorithm (SIMBA) for free-running motion-suppressed whole-heart acquisitions. By selecting timeseries with a high low-frequency energy content, we include only pixels with respiratory and cardiac information. Compared to the original method, this leads to a more accurate choice of end-expiration and diastolic phases for the reconstruction of sharp whole-heart and coronary images. Moving forward, the method needs to be refined, optimized and tested to further improve the image quality.
Introduction
SIMBA was proposed as a fast data-driven approach for the reconstruction of static motion-consistent 3D whole-heart images from untriggered and ungated free-running acquisitions1. This approach relies on the clustering of reference datapoints, i.e. superior-inferior (SI) projections, regularly acquired during the scan. The clustering is based on physical similarities in the signals, irrespective of their timepoint of acquisition. The original implementation (SIMBAc) uses the whole extent of the SI projections of the four central elements of the body array coil, located at the center of the chest, as an input to target the heart and liver2. However, in some instances SIMBAc fails at isolating precise motion states, leading to suboptimal image quality. We therefore propose a novel approach (SIMBAp) that consists of an automated subject-specific pixels selection to extract the signal components that can best promote motion consistency for SIMBA. This new algorithm was tested in a series of ferumoxytol-enhanced scans obtained from 15 patients with congenital heart disease.Methods
Pixels selection: First, all the SI projections were concatenated in time, for each coil, obtaining a 3D matrix $$$S$$$ of size ($$$N_{pixels} \text{x} N_{segments} \text{x} N_{coils}$$$). For each pixel $$$p$$$ for each coil $$$c$$$, the power spectral density $$$s(p,c)$$$ of each timeseries $$$t(p,c)$$$ is computed. $$$s(p,c)$$$ was then split into two parts, one encompassing only low-frequency components $$$s_{low}(p,c)$$$, i.e. containing relevant physiological information, and a high-frequency one $$$s_{high}(p,c)$$$, containing mainly components from trajectory imperfections. Considering only $$$s_{low}(p,c)$$$, the low-frequency energy of each pixel is calculated as $$$E(p,c) = \sum s_{low}(p,c,f)^2$$$. Finally, the pixel selection was set to include pixels with an energy value higher than an empirically selected 15% of the maximal energy at low frequencies (Figure 2). At this point, the final matrix $$$\hat{S}$$$ has a size of ($$$M \text{x} N_{segments}$$$), with $$$M$$$ being the total number of selected pixels. Finally, $$$\hat{S}$$$ was low-pass filtered to minimize the effect of non-physiological components.
SIMBA: The main steps of SIMBA were implemented as in the original publication1 as illustrated in Figure 1.
Data acquisition: Fifteen congenital heart disease patients (17$$$\pm$$$10 years; 56$$$\pm$$$26 kg; 13 males) were scanned on a 1.5T (MAGNETOM Sola, Siemens Healthcare, Erlangen, Germany), after injection of ferumoxytol (2mg/kg). All datasets were acquired using a prototype free-running GRE sequence1 with a 3D radial phyllotaxis trajectory3, and the ECG recorded.
Data analysis: The low-dimensional space (Figure 1) was characterized in terms of compactness of the largest cluster, inversely related to the mean distance of each point from the centroid, the number of segments in the largest cluster, and the trajectory uniformity1. Analysis of the cardiac data selection for SIMBAc and SIMBAp was performed by calculating the percentage of diastolic data, taking the QT interval from the ECG as the duration of systole4. The respiratory data selection was assessed using a self-gating respiratory signal extraction5.
Finally, the image quality for the two algorithms was compared in terms of blood-myocardium sharpness6, the visible length and sharpness of the right coronary artery (RCA), and those of the combined left main (LM) + left anterior descending coronary artery (LAD), determined using the Soap-Bubble software tool7.Results
Pixels selection: SIMBAp always resulted in a smaller input matrix, with an
average of 83$$$\pm$$$42 pixels vs 764 obtained with SIMBAc. In only 5 of 15
cases the selected pixels were all located in the four coil elements chosen by
SIMBAc.
Physiological data selection: For 7 of 15 cases, SIMBAp resulted in a
more accurate selection of end-expiration, and in 5 of these 7 cases in a more
precise selection of diastolic data: SIMBAp 86.8$$$\pm$$$13.8% vs SIMBAc
67.7$$$\pm$$$32.5% (Figure 3A). This greatly improved the image quality
(Figure 3B), resulting in a significantly higher blood-myocardium sharpness
(P<0.05). Moreover, these instances were the ones for which the
pixels selection differed for up to 75% from the standard body coil selection
of SIMBAc.
Data analysis: The compactness of the largest cluster was significantly
higher for SIMBAp compared to SIMBAc, while both trajectory uniformity and size
of the largest cluster were not different (Figure 4A). A comparison of the
mean blood-myocardium sharpness, visible length and sharpness of RCA and LM+LAD
for all 15 patients didn’t show any statistically significant differences
between the two methods (Figure 4B). Figure 5 shows an example reconstruction
with both methods. Discussion
In
this work we developed and tested an alternative approach to the original fixed
coil selection for the creation of the input data matrix for SIMBA. We
demonstrated the advantages of an automated subject-specific selection driven
by the frequency spectrum of single pixels from the SI projections of
individual coil elements. This new approach improved the physiological data
selection, resulting in sharper images where the original approach was
incapable of adequately capturing all physiological components. Using SIMBAp we
observed some instances where a good suppression of respiratory motion was not paralleled
by an equally good suppression of cardiac motion. The algorithm’s steps should
be optimized to improve the results in these cases as well. We plan
to further improve this technique, also considering different dimensionality
reduction and clustering methods. Moreover, we would like to test this pixel
selection for the self-gating signal extraction as part of the 5D free-running
framework. Acknowledgements
No acknowledgement found.References
- Heerfordt
J, Whitehead KK, Bastiaansen JAM, et al. Similarity‐driven multi‐dimensional binning algorithm
(SIMBA) for free‐running
motion‐suppressed whole‐heart MRA. Magn Reson Med
2021;86:213–229.
- Piccini D,
Littmann A, Nielles-Vallespin S, Zenge MO. Respiratory
self-navigation for whole-heart bright-blood coronary MRI: Methods for robust
isolation and automatic segmentation of the blood pool. Magn Reson Med.
2011;68:571–579.
- Piccini D, Littmann A, Nielles-Vallespin S, Zenge MO. Spiral
phyllotaxis: the natural way to construct a 3D radial trajectory in MRI. Magn
Reson Med. 2011;66:1049-1056.
- Roes SD,
Korosoglou G, Schär M, et al. Correction
for heart rate variability during 3D whole heart MR coronary angiography. J
Magn Reson Imaging. 2008;27:1046-1053.
- Di Sopra L, Piccini D,
Coppo S, Stuber M, Yerly J. An automated approach to fully self-gated
free-running cardiac and respiratory motion-resolved 5D whole-heart MRI. Magn
Reson Med. 2019;82:2118-2132.
- Ahmad,
R., Ding, Y., & Simonetti, O. P. (2015). Edge sharpness assessment by
parametric modeling: application to magnetic resonance imaging. Concepts
in Magnetic Resonance Part A, 44(3), 138-149.
- Etienne
A, Botnar RM, Van Muiswinkel AMC, Boesiger P, Manning WJ, Stuber M.
“Soap-Bubble” visualization and quantitative analysis of 3D coronary magnetic resonance
angiograms. Magn Reson Med. 2002;48:658-666.