T2-weighted imaging is conventionally obtained using multi-slice 2D or 3D fast spin-echo (FSE) sequences. However, FSE suffers from severe specific absorption rate (SAR) issue and motion artifact particularly in the abdomen. A potential approach to address these challenges is using T2-prepared 3D SPGR sequence with a non-Cartesian trajectory such as cones. In this study, we proposed a T2-prepared 3D-cones sequence to achieve T2-weighted abdominal imaging with free breathing. The feasibility of the technique was tested on both healthy subjects and patients.
1. Takayama, Y. et al. Three-dimensional T2-weighted imaging for liver MRI: Clinical values of tissue-specific variable refocusing flip-angle turbo spin echo imaging: 3D T2-Weighted Imaging for Liver MRI. J. Magn. Reson. Imaging 41, 339–346 (2015).
2. Ream, J. M. and Rosenkrantz, A. B. Advances in T1-Weighted and T2-Weighted Imaging in the Abdomen and Pelvis. Radiol. Clin. North Am. 53, 583–598 (2015).
3. Pipe, J. G. Motion correction with PROPELLER MRI: application to head motion and free-breathing cardiac imaging. Magn. Reson. Med. 42, 963–969 (1999).
4. Zhu, J. et al. T2-prepared segmented 3D-gradient-echo for fast T2-weighted high-resolution three-dimensional imaging of the carotid artery wall at 3T: a feasibility study. Biomed. Eng. OnLine 15, 165 (2016).
5. Klupp, E. et al. B1-insensitive T2 mapping of healthy thigh muscles using a T2-prepared 3D TSE sequence. PLOS ONE 12, e0171337 (2017).
6. Brittain, J. H. et al. Coronary Angiography with Magnetization-PreparedT2 Contrast. Magn. Reson. Med. 33, 689–696 (1995).
7. Gurney, P. T., Hargreaves, B. A. & Nishimura, D. G. Design and analysis of a practical 3D cones trajectory. Magn. Reson. Med. 55, 575–582 (2006).
8. Chen, B. et al. Three-dimensional ultrashort echo time cones (3D UTE-Cones) magnetic resonance imaging of entheses and tendons. Magn. Reson. Imaging 49, 4–9 (2018).
9. Norbeck, O. et al. Optimizing 3D EPI for rapid T 1 ‐weighted imaging. Magn. Reson. Med. 84, 1441–1455 (2020).
10. Uecker, Martin et al. mrirecon/bart: version 0.7.00. (Zenodo, 2021). doi:10.5281/ZENODO.592960.
Figure 1: A sequence diagram used in this study with T2 preparation module in a 3D-cones sequence. Four MLVT pulses were designed to be more robust to B1 inhomogeneity. A large spoiling gradient is played after the tipup pulse to destroy the residual transverse magnetization. Followed by the T2 preparation module is a set of 3D-cones acquisition blocks (N = 20 in this study). For each TR, a wait time is also required to allow for the longitudinal magnetization to recover. A 2.3 ms hard RF pulse is used to achieve fat suppression.
Figure 2: Illustration of the banding artifact resulting from B1 inhomogeneity (A, indicated by the red arrows). The artifacts can be reduced by adding a pair of crushers in the T2 preparation pulses (B).
Figure 3: Illustration of the blurred image due to off-resonance using longer cones readout (8ms, A). Decreasing the readout can provide better image quality (6.5ms, B), however, at the expense of doubling the acquisition time.
Figure 4: Liver images acquired from a healthy subject (male, 29 yo) using gridding reconstruction (A) and neural network off-resonance correction (B), respectively. A focal lesion can be clearly identified from both images as indicated by the red arrow. Compared with the gridding reconstruction alone, neural network off-resonance correction provides enhanced image quality.
Figure 5: Liver images acquired from a pediatric patient (two years-old) using T2-weighted 2D FSE (A), T2-prepared 3D-cones (B) and contrast-enhanced T1-weighted 3D-cones (C), respectively. The 2D acquisition suffers motion artifacts (indicated by the red arrows), whereas 3D-cones sequence is robust to motion. The T2 contrast provided by T2-prepared 3D-cones makes the vessels bright, which is comparable to the contrast-enhanced images.