Teresa Nolte1, Shuo Zhang2, Nicola Pridöhl1, Malin Ciba1, Masami Yoneyama3, Christiane Kuhl1, and Sven Nebelung4
1Diagnostic and Interventional Radiology, University Hospital Aachen, Aachen, Germany, 2Philips GmbH Market DACH, Hamburg, Germany, 3Philips Japan, Tokyo, Japan, 4Clinic for Diagnostic and Interventional Radiology, Uniklinik RWTH Aachen University, Aachen, Germany
Synopsis
This work
aimed to propose a clinical sequence platform for simultaneous morphologic and
quantitative imaging of joints and identified TSE-based MIXTURE (Multi-Interleaved
X-prepared Turbo-Spin Echo with Intuitive Relaxometry) sequences in terms of
PDFS-T2 and T1-T1ρFS combinations for clinical implementation. Even
though capable of isotropic resolution, MIXTURE sequences were
acquired as pseudo-3D (thicker slices, higher in-plane resolution) to match standard-clinical
2D TSE sequences. After identification of clinical demands, MIXTURE sequences
were systematically optimized, while maintaining clinically feasible
acquisition times of 5:00 min (PDFS-T2 for T2-mapping) and 6:40 min (T1-T1ρFS for T1ρ-mapping), and evaluated in a cadaveric human
knee cartilage defect model.
Introduction
Quantitative
MRI techniques such as T2 and T1ρ mapping are diagnostically beneficial in
the comprehensive evaluation of various pathologies of the joint, yet require
long acquisition times which limits their clinical adoption. MIXTURE (Multi-Interleaved
X-prepared Turbo-Spin Echo with Intuitive Relaxometry) sequences have recently been
proposed for simultaneous morphologic and quantitative MRI of joints1.
Quantitative data are calculated based on two or more interleaved turbo-spin
echo (TSE) images with different preparation modules and hence adjustable
contrast. Standard 2D TSE sequences (proton density (PD)-weighted, T2-weighted
with or without fat-saturation (FS) and T1-weighted) are the diagnostic mainstay of
joint imaging in our department and beyond. Previously, isotropic MIXTURE sequences
were suggested, e.g., a PD-T2FS MIXTURE sequence with T2
preparation pulses of different lengths2, and a PD-T1ρFS MIXTURE
sequence with T1ρ preparation pulses of different spin lock times (TSL)3.
Pseudo-3D MIXTURE variants with higher in-plane resolution and thicker slices resemble
the clinical-standard 2D TSE sequences more closely and may hence be used for comparative
evaluation of the MIXTURE sequence platform. Hence, we present optimized
pseudo-3D MIXTURE sequences of human cadaveric knee joint specimens with standardized traumatic cartilage defects
and visually evaluate diagnostic aspects in comparison to clinical-standard 2D TSE sequences. Methods
Measurements
were performed on human cadaveric knee joint specimens following
approval by the local Ethical Committee on a clinical 3T system (Philips
Elition X, Best, The Netherlands with a 16-channel clinical T/R knee coil
(Philips). Clinical needs and optimization criteria for morphologic MIXTURE
images were discussed with a clinical radiologist (8 years of experience in musculoskeletal
imaging) and implemented accordingly. Guided by the our institution’s clinical
standard imaging protocol for the knee that consists of PDFS sequences (axial,
coronal, sagittal) and a T1-weighted sequence (sagittal), two sagittal MIXTURE
sequences, i.e., PDFS-T2 MIXTURE for T2 mapping and T1- T1ρFS MIXTURE
for T1ρ mapping were acquired in pseudo-3D configuration and
optimized to provide similar diagnostic quality as the 2D scans while optimizing
scan time (sequence parameters of reference sequences and optimized MIXTURE
protocols c.f. Figure 1). To evaluate the different sequences in
close-to-clinical conditions, the MIXTURE and clinical reference sequences were
acquired before (i.e., intact joint) and after creating standardized defects of
3, 5, and 8 mm in diameter at the weight-bearing area of the lateral femoral condyle
of n=3 knee joint specimens. Results
Figure
2 shows the optimization aspects with regard to clinical needs and relevant sequence
parameters. Image blurring in MIXTURE sequences associated with the underlying
3D technique and long TSE readouts were largely reduced by using shorter TSE
factors and adapting refocusing patterns4. Figure 3 shows the 2D PD-weighted
FS TSE (reference) sequence and the morphologic images of the optimized MIXTURE
PDFS-T2 sequence for the knee without and with standardized cartilage defects,
while Figure 4 indicates the corresponding 2D T1-weighted reference sequence
and the morphologic images of the optimized MIXTURE T1-T1ρFS sequence. After
optimization, the morphologic PDFS MIXTURE images allow clear visualization and
delineation of the cartilage tissue and their defects with comparable overall
image quality, contrast, and signal strength. The morphologic T1 MIXTURE image
depicts the same anatomic features, in particular regarding bone texture and
structure, as the clinical reference sequence. Figure 5 visualizes the MIXTURE-based
T2 and T1ρ maps of the respective sagittal slices, which demonstrate the layered
structure of cartilage as well as the defects.Discussion
T2
and T1ρ MIXTURE sequences, acquired in pseudo-3D configuration in human cadaveric
knee joint specimens, enabled a voxel-wise, motion-free comparison with the
clinical 2D protocols which is a step towards additional validation and
clinical implementation of the MIXTURE sequence platform. After optimization,
morphologic MIXTURE images are diagnostically sufficient in terms of image quality,
contrast, and signal strength. Thus, MIXTURE sequences acquired as pseudo-3D sequences
may be useful candidates for diagnostic morphologic joint imaging with adjunct
mapping of distinct biophysical properties that allow inferences on tissue
ultrastructure and composition beyond mere morphology. This study is the first to
propose pseudo-3D PDFS-T2 and T1-T1ρFS variants of MIXTURE in a human model of
cartilage pathology.Conclusion
MIXTURE
sequences are promising candidates for combined whole-joint morphologic and
quantitative assessment in the future at scan times of around 5-6 minutes.
Future research should validate the MIXTURE-based relaxation times against reference
T2 and T1ρ mapping sequences in dedicated phantoms and collect
MIXTURE data in volunteers and patients. Acknowledgements
No acknowledgement found.References
1Yoneyama
M, Sakai T, Zhang S, et al. MIXTURE: a novel sequence for
simultaenous morphological and quantitative imaging based on multi-interleaved
3D turbo-spin echo MRI. Proc. Intl. Soc. Mag. Reson. Med. 29 (2021), 4203.
2Sakai T, Yoneyama M, Watanabe A, et al. Simultaneous anatomical, pathological
and T2 quantitative knee imaging with 3D submillimeter isotropic resolution
using MIXTURE. Proc. Intl. Soc. Mag. Reson. Med. 29 (2021), 0845.
3Saruya S, Suzuki M, Yoneyama M, et al. 3D sub-millimeter isotropic knee
cartilage T1rho mapping using multi-interleaved fluid-attenuated TSE
acquisition (MIXTURE). Proc. Intl. Soc. Mag. Reson. Med. 29 (2021), 2977.
4Mugler JP. Optimized Three-Dimensional Fast-Spin-Echo MRI. J Magn. Reson.
Med. (2014) 39:745–767.